- ICH GCP
- Registro degli studi clinici negli Stati Uniti
- Sperimentazione clinica NCT00920452
The 4-SITE Field Following Study (4-SITE FF) (4-SITE FF)
RELIANCE QUADRIPOLAR DEFIBRILLATION LEADS (4-SITE)FIELD FOLLOWING The 4-SITE Field Following Study (4-SITE FF)
The purpose of this study is to evaluate and document appropriate clinical performance of the new 4-SITE Header / Lead interface in the TELIGEN 100 HE 4-SITE (VR and DR) implantable cardioverter defibrillator (ICD) PGs (Models F103; F111), the COGNIS 100 HE 4-SITE cardiac resynchronization therapy ICD (CRT-D) PG (Model P108) when connected to any of the RELIANCE Quadripolar (4-SITE) defibrillation leads.
Appropriate clinical performance of system components of the 4-SITE systems other than the new 4-SITE header / lead interface has been evaluated and documented already in the previously conducted COGNIS and TELIGEN 100 HE FIELD FOLLOWING STUDY, The COGENT-4 Field Following Study.
Data collected may be used to support international regulatory submissions including the FDA and the Japanese Ministry of Health, Labour and Welfare.
Panoramica dello studio
Stato
Condizioni
Descrizione dettagliata
The following devices will be used (all are CE-marked):
TELIGEN 100 HE 4-SITE (Single and Dual Chamber ICD)
- Model Number: F103 (Quadripolar, VR);
- Model Number: F111 (Quadripolar, IS-1, DR)
COGNIS 100 HE 4-SITE (CRT-D)
o Model Number: P108 (Quadripolar, IS-1, IS-1)
- RELIANCE 4-SITE leads (12 different model numbers in total as shown in the table below)
Tipo di studio
Iscrizione (Effettivo)
Contatti e Sedi
Luoghi di studio
-
-
Nordrhein-Westfalen
-
Bad Oeynhausen, Nordrhein-Westfalen, Germania, D - 32545
- Bad Oeynhausen Herz- und Diabeteszentrum Nordrhein-Westfalen
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Metodo di campionamento
Popolazione di studio
Descrizione
Inclusion Criteria:
Study Specific:
- ICD / CRT-D Indication according to normal clinical practice
- Patients receiving:
- a single or dual chamber 4-SITE compatible ICD
- or a 4-SITE compatible CRT-D
- one of the RELIANCE 4-SITE defibrillation leads
- Patients currently implanted with a pacemaker
- upgraded to a 4-SITE compatible ICD or CRT-D (4-SITE header)
- one of the RELIANCE 4-SITE defibrillation leads
General:
- Willing and capable of providing informed consent for
- undergoing a 4-SITE system implant,
- participating in all testing associated with this clinical investigation at an approved clinical investigational centre and at the intervals defined by this protocol
- Geographically stable patients who are available for follow-up at a study centre
- Age 18 or above, or of legal age to give informed consent specific to national law
Exclusion Criteria:
Study Specific:
- ICD and CRT-D Patients scheduled for a device replacement
- CRM Patients who have or who would need an lead adaptor
- All patients who have an active or non-active defibrillation lead other than 4-SITE
General:
- Not willing and not capable of providing informed consent, undergoing a device implant, participating in all testing associated with this clinical investigation (including VT/VF shock conversion) at an approved clinical investigational centre and at the intervals defined by this protocol
- Patients who were in NYHA Class IV during the last 3 month
- Patients with pre-existing diseases, which may confound study results
- Patients currently requiring dialysis,
- Cancer patients
- Patients with drug and/or alcohol abuse history
- Life expectancy < 12 months (or expected heart transplant within 12 months)
- Patients on a Heart Transplant List
- Women who are pregnant or plan to become pregnant. Method of assessment per physician discretion.
- Enrolled in any other concurrent study
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Prospettive temporali: Prospettiva
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Appropriate Detection of induced VT
Lasso di tempo: At Implant or during VT/VF testing up to one month after implant
|
Induced mean VT/VF detection time < 4.5 seconds
|
At Implant or during VT/VF testing up to one month after implant
|
|
Appropriate induced VT/VF Shock Conversion
Lasso di tempo: At Implant or during testing within one month after implant
|
Successful VT/VF conversions (with 41 J, no external shock) will be 93% or higher
|
At Implant or during testing within one month after implant
|
|
Appropriate pacing thresholds at follow-up with the 4-SITE defibrillation leads
Lasso di tempo: entire duration of the study
|
Mean pacing threshold post implant ≤ 1.5V
|
entire duration of the study
|
|
Appropriate Lead Impedances as a measure of lead integrity over 12 month time
Lasso di tempo: entire study duration
|
|
entire study duration
|
|
Appropriate sensing and absence of artefacts
Lasso di tempo: entire duration of study
|
Appropriate sensing and absence of artefacts during standardized pocket manipulations and standardized provocative maneuvers during FU
|
entire duration of study
|
Collaboratori e investigatori
Sponsor
Studiare le date dei record
Studia le date principali
Inizio studio
Completamento primario (Effettivo)
Completamento dello studio (Effettivo)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Stima)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Stima)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Altri numeri di identificazione dello studio
- 4-SITE FF 0209
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .
Prove cliniche su ICD and CRT-D Indicated Patients
-
Biotronik SE & Co. KGCompletato
-
Abbott Medical DevicesCompletatoIl paziente soddisfa le linee guida ACC/AHA/ESC per il defibrillatore cardioverter impiantabile (ICD) o il dispositivo per la terapia di risincronizzazione cardiaca (CRT-D)Regno Unito, Germania