Peripheral intravenous cannulation with support of infrared laser vein viewing system in a pre-operation setting in pediatric patients

Andreas Rothbart, Peng Yu, Lutz Müller-Lobeck, Claudia D Spies, Klaus-Dieter Wernecke, Irit Nachtigall, Andreas Rothbart, Peng Yu, Lutz Müller-Lobeck, Claudia D Spies, Klaus-Dieter Wernecke, Irit Nachtigall

Abstract

Background: Venous access, a prerequisite for anesthesiological and surgical intervention in pediatric patients, is often difficult to establish and potentially painful. AV300 uses near infrared laser light to improve visibility of peripheral veins and could help cannulating them. The aim of this study was to examine if use of Accuvein(®) AV300 vein viewer could facilitate venous cannulation in children.

Methods: From January to March 2011, 238 consecutive pediatric patients (0-17 years) preceding surgical interventions were included. All participants including newborns, infants and children were allocated to groups [control group (124 patients) and intervention group (114 patients)] in a non-random way. Randomization was not feasible because data was acquired retrospectively from a clinical quality management project. In control group, peripheral IV cannulation was performed without supporting device, in intervention group with support of AV300. Time and number of attempts until successful venous cannulation were defined as primary end points.

Results: Median time until successful cannulation was 2 min (range 0.1-20, quartiles: 25 %: 1; 75 %: 5) in the intervention group and 1 min (range 0.1-18, quartiles: 25 %: 0.2; 75 %: 2) in the control group (p < 0.01). Median number of attempts was higher in the intervention group (2; range 1-6, quartiles: 25 %: 1; 75 %: 3) than in the control group (1; range 1-6, quartiles: 25 %: 1; 75 %: 2, p < 0.01). Rate of cannulations successful at first attempt was 0.45 (51 of 114, 95 % CI 0.35-0.54) in the intervention group and 0.73 (90 of 124, 95 % CI 0.65-0.81) in the control group (p < 0.01).

Conclusions: In our study we were not able to reduce neither time nor number of attempts until a successful venous cannulation in children using the vein viewer. Given certain limitations of our study as the lack of randomization and no control for inter-operator variability, the conclusions drawn from it are also limited, but by our results laser-supported cannulation cannot be recommended for standard procedures.

Trial registration: ClinicalTrials.gov NCT01434537. Registered 29 July 2011.

Figures

Fig. 1
Fig. 1
Time until successful venous cannulation (CON control; EXP experimental group)
Fig. 2
Fig. 2
Number of attempts until succesful venous cannulation (CON control; EXP experimental group)
Fig. 3
Fig. 3
Time until successful venous cannulation in subgroup up to 72 months (CON control; EXP experimental group)
Fig. 4
Fig. 4
Number of attempts until succesful venous cannulation in subgroup up to 72 months (CON control; EXP experimental group)

References

    1. Koh JL, Harrison D, Myers R, Dembinski R, Turner H, McGraw T. A randomized, double-blind comparison study of EMLA and ELA-Max for topical anesthesia in children undergoing intravenous insertion. Paediatr Anaesth. 2004;14(12):977–982. doi: 10.1111/j.1460-9592.2004.01381.x.
    1. Yen K, Riegert A, Gorelick MH. Derivation of the DIVA score: a clinical prediction rule for the identification of children with difficult intravenous access. Pediatr Emerg Care. 2008;24(3):143–147. doi: 10.1097/PEC.0b013e3181666f32.
    1. Shykoff BE, Hawari FI, Izzo JL., Jr Diameter, pressure and compliance relationships in dorsal hand veins. Vasc Med. 2001;6(2):97–102. doi: 10.1177/1358836X0100600205.
    1. Crandall CG, Johnson JM, Kosiba WA, Kellogg DL., Jr Baroreceptor control of the cutaneous active vasodilator system. J Appl Physiol. 1996;81(5):2192–2198.
    1. Nee PA, Picton AJ, Ralston DR, Perks AG. Facilitation of peripheral intravenous access: an evaluation of two methods to augment venous filling. Ann Emerg Med. 1994;24(5):944–946. doi: 10.1016/S0196-0644(94)70211-X.
    1. Doniger SJ, Ishimine P, Fox JC, Kanegaye JT. Randomized controlled trial of ultrasound-guided peripheral intravenous catheter placement versus traditional techniques in difficult-access pediatric patients. Pediatr Emerg Care. 2009;25(3):154–159. doi: 10.1097/PEC.0b013e31819a8946.
    1. Hosokawa K, Kato H, Kishi C, Kato Y, Shime N. Transillumination by light-emitting diode facilitates peripheral venous cannulations in infants and small children. Acta Anaesthesiol Scand. 2010;54(8):957–961. doi: 10.1111/j.1399-6576.2010.02270.x.
    1. Simhi E, Kachko L, Bruckheimer E, Katz J. A vein entry indicator device for facilitating peripheral intravenous cannulation in children: a prospective, randomized, controlled trial. Anesth Analg. 2008;107(5):1531–1535. doi: 10.1213/ane.0b013e318185cdab.
    1. Zeman HD, Lovhoiden G, Vrancken C. The clinical evaluation of vein contrast enhancement. Conf Proc IEEE Eng Med Biol Soc. 2004;2:1203–1206.
    1. Zharov VP, Ferguson S, Eidt JF, Howard PC, Fink LM, Waner M. Infrared imaging of subcutaneous veins. Lasers Surg Med. 2004;34(1):56–61. doi: 10.1002/lsm.10248.
    1. Morgan NY, English S, Chen W, Chernomordik V, Russo A, Smith PD, Gandjbakhche A. Real time in vivo non-invasive optical imaging using near-infrared fluorescent quantum dots. Acad Radiol. 2005;12(3):313–323. doi: 10.1016/j.acra.2004.04.023.
    1. Cuper NJ, Verdaasdonk RM, de Roode R, de Vooght KM, Viergever MA, Kalkman CJ, de Graaff JC. Visualizing veins with near-infrared light to facilitate blood withdrawal in children. Clin Pediatr (Phila) 2011;50(6):508–512. doi: 10.1177/0009922810395932.
    1. AV300 A. AV300 User Manual, AV300 Operating Guide: AccuVein®, LLC; 2009.
    1. Perry AM, Caviness AC, Hsu DC. Efficacy of a near-infrared light device in pediatric intravenous cannulation: a randomized controlled trial. Pediatr Emerg Care. 2011;27(1):5–10. doi: 10.1097/PEC.0b013e3182037caf.
    1. Chapman LL, Sullivan B, Pacheco AL, Draleau CP, Becker BM. VeinViewer-assisted Intravenous catheter placement in a pediatric emergency department. Acad Emerg Med: Off J Soc Acad Emerg Med. 2011;18(9):966–971. doi: 10.1111/j.1553-2712.2011.01155.x.
    1. Kim MJ, Park JM, Rhee N, Je SM, Hong SH, Lee YM, Chung SP, Kim SH. Efficacy of VeinViewer in pediatric peripheral intravenous access: a randomized controlled trial. Eur J Pediatr. 2012;171(7):1121–1125. doi: 10.1007/s00431-012-1713-9.
    1. Teillol-Foo WL, Kassab JY. Topical glyceryl trinitrate and eutectic mixture of local anaesthetics in children. A randomised controlled trial on choice of site and ease of venous cannulation. Anaesthesia. 1991;46(10):881–884. doi: 10.1111/j.1365-2044.1991.tb09610.x.
    1. Evers H, von Dardel O, Juhlin L, Ohlsen L, Vinnars E. Dermal effects of compositions based on the eutectic mixture of lignocaine and prilocaine (EMLA). Studies in volunteers. Br J Anaesth. 1985;57(10):997–1005. doi: 10.1093/bja/57.10.997.
    1. Egekvist H, Bjerring P. Effect of EMLA cream on skin thickness and subcutaneous venous diameter. A randomized, placebo-controlled study in children. Acta Derm Venereol. 2000;80(5):340–343. doi: 10.1080/000155500459268.
    1. Strehle EM. Making the invisible visible: near-infrared spectroscopy and phlebotomy in children. Telemed J E Health. 2010;16(8):889–893. doi: 10.1089/tmj.2010.0061.
    1. Miyake RK, Zeman HD, Duarte FH, Kikuchi R, Ramacciotti E, Lovhoiden G, Vrancken C. Vein imaging: a new method of near infrared imaging, where a processed image is projected onto the skin for the enhancement of vein treatment. Dermatol Surg. 2006;32(8):1031–1038.
    1. Szmuk P, Steiner J, Pop RB, Farrow-Gillespie A, Mascha EJ, Sessler DI. The VeinViewer vascular imaging system worsens first-attempt cannulation rate for experienced nurses in infants and children with anticipated difficult intravenous access. Anesth Analg. 2013;116(5):1087–1092. doi: 10.1213/ANE.0b013e31828a739e.
    1. de Graaff JC, Cuper NJ, Mungra RA, Vlaardingerbroek K, Numan SC, Kalkman CJ. Near-infrared light to aid peripheral intravenous cannulation in children: a cluster randomised clinical trial of three devices. Anaesthesia. 2013;68(8):835–845. doi: 10.1111/anae.12294.

Source: PubMed