Biplanar Handheld Ultrasound for Transradial Artery Puncture
Tuoren Biplanar Combined Handheld Ultrasound Versus Conventional Ultrasound for Transradial Artery Puncture A Prospective Randomized Parallel Single-Center Superiority Clinical Study
Radial artery puncture is commonly used to obtain vascular access for cerebral angiography and neurointerventional procedures. Conventional ultrasound guidance usually provides either a short-axis or a long-axis view of the radial artery. A biplanar handheld ultrasound system can provide simultaneous long-axis and short-axis views, which may help the operator visualize the puncture needle and radial artery more clearly.
This prospective, single-center, randomized, parallel-group superiority trial will compare the TUOREN biplanar handheld ultrasound system with a conventional high-frequency linear-array ultrasound system for ultrasound-guided radial artery puncture. A total of 152 adults scheduled to undergo cerebral angiography or neurointerventional treatment through a transradial approach will be randomly assigned in a 1:1 ratio to one of the two ultrasound-guidance groups.
The main objective is to determine whether biplanar handheld ultrasound guidance improves the percentage of participants with successful radial artery cannulation on the first puncture attempt. Other assessments include the total number of puncture attempts, cumulative active puncture time, fixation time after successful cannulation, and the occurrence of posterior wall puncture.
Panoramica dello studio
Stato
Condizioni
Descrizione dettagliata
Ultrasound guidance is used to improve visualization of the radial artery during vascular access. Conventional high-frequency linear-array ultrasound systems provide either a short-axis or a long-axis image, depending on probe orientation and puncture technique. During puncture, the operator may need to adjust the probe repeatedly to maintain visualization of the needle tip and its position relative to the radial artery.
The TUOREN biplanar handheld ultrasound system uses a T-shaped dual-probe configuration to display simultaneous long-axis and short-axis views of the radial artery and puncture needle. The long-axis view is intended to show the needle trajectory and advancement, while the short-axis view is intended to show the lateral position of the needle relative to the vessel.
This is a prospective, randomized, parallel-group, single-center superiority trial conducted at Shanghai Fourth People's Hospital. Eligible participants scheduled for cerebral angiography or neurointerventional treatment through a transradial approach will be allocated in a 1:1 ratio using block randomization with allocation concealment.
Participants in the experimental group will undergo radial artery puncture guided by the TUOREN biplanar handheld ultrasound system. Participants in the active comparator group will undergo radial artery puncture guided by a conventional high-frequency linear-array ultrasound system using either a short-axis out-of-plane technique or a long-axis in-plane technique. Other components of the radial artery cannulation procedure will be standardized between the two groups.
The study is designed to evaluate whether simultaneous biplanar ultrasound visualization improves first-attempt radial artery cannulation success and procedural efficiency compared with conventional ultrasound guidance. Puncture attempts, procedure-related time intervals, and ultrasound findings will be recorded according to prespecified study definitions.
Tipo di studio
Iscrizione (Stimato)
Fase
- Non applicabile
Contatti e Sedi
Luoghi di studio
-
-
Shanghai Municipality
-
Shanghai, Shanghai Municipality, Cina, 200435
- Shanghai Fourth People's Hospital
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
- Adulto
- Adulto più anziano
Accetta volontari sani
Descrizione
Inclusion Criteria:
- Adults aged 18 years or older, regardless of sex, who are scheduled to undergo cerebral angiography or neurointerventional treatment through a transradial approach.
- Adequate collateral circulation between the radial and ulnar arteries, as demonstrated by the modified Allen test or vascular ultrasound.
- Radial artery diameter of at least 1.5 mm on preprocedural ultrasound examination.
- Ability to understand the study procedures, willingness to participate voluntarily, and provision of written informed consent.
Exclusion Criteria:
- Infection or skin damage at the intended puncture site; known radial artery occlusion or severe stenosis; or inadequate collateral circulation between the radial and ulnar arteries.
- Radial artery diameter of less than 1.5 mm on preprocedural ultrasound examination, or severe radial artery tortuosity or anatomical variation that may interfere with the puncture procedure.
- Previous puncture, surgery, or trauma involving the radial artery on the intended puncture side that may have altered the local anatomy.
- Hemodynamic instability requiring urgent hemodynamic support, including cardiogenic shock or severe hypotension.
- Impaired consciousness, psychiatric illness, or any other condition that prevents the participant from cooperating with the study procedures or completing the required follow-up.
- Severe coagulation abnormality, defined as an international normalized ratio greater than 1.8 or a platelet count below 80 × 10^9/L.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Altro
- Assegnazione: Randomizzato
- Modello interventistico: Assegnazione parallela
- Mascheramento: Nessuno (etichetta aperta)
Armi e interventi
Gruppo di partecipanti / Arm |
Intervento / Trattamento |
|---|---|
|
Sperimentale: TUOREN Biplanar Handheld Ultrasound Guidance
Participants assigned to this arm will undergo transradial artery puncture guided by TUOREN biplanar handheld ultrasound which provides simultaneous long-axis and short-axis views.
|
A biplanar handheld ultrasound system used to guide radial artery puncture.
Its T-shaped dual-probe configuration provides simultaneous long-axis and short-axis views of the radial artery and puncture needle.
|
|
Comparatore attivo: Conventional Ultrasound Guidance
Participants assigned to this arm will undergo transradial artery puncture guided by a conventional high-frequency linear array ultrasound probe using short-axis out-of-plane or long-axis in-plane techniques.
|
A conventional high-frequency linear-array ultrasound system used to guide radial artery puncture using either a short-axis out-of-plane technique or a long-axis in-plane technique.
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
First-attempt success rate
Lasso di tempo: During the puncture procedure
|
Proportion of participants with successful radial artery cannulation on the first continuous puncture attempt.
Success is confirmed by sustained blood return guidewire advancement at least 5 to 10 cm without resistance and an arterial pressure waveform after catheter connection.
|
During the puncture procedure
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Mean total number of puncture attempts required for successful radial artery cannulation
Lasso di tempo: From initial needle-to-skin contact until confirmation of successful radial artery cannulation during the index procedure
|
The total number of puncture attempts required to achieve successful radial artery cannulation will be recorded for each participant.
An attempt begins when the needle first contacts the skin and ends when successful cannulation is confirmed or when the needle is completely withdrawn from the skin.
Each reinsertion after complete withdrawal of the needle from the skin is counted as a new attempt.
The outcome will be reported as the mean number of puncture attempts per participant.
|
From initial needle-to-skin contact until confirmation of successful radial artery cannulation during the index procedure
|
|
Mean cumulative active puncture time to successful radial artery cannulation
Lasso di tempo: From initial needle-to-skin contact until confirmation of successful radial artery cannulation during the index procedure
|
Cumulative active puncture time is defined as the total active puncture time, measured in seconds, from the first needle-to-skin contact until successful radial artery cannulation is confirmed by an arterial pressure waveform.
Timing will be paused whenever the needle is completely withdrawn from the skin and resumed when the needle contacts the skin for the next attempt.
Time used for skin preparation, draping, ultrasound probe preparation, and intervals between puncture attempts will not be included.
The outcome will be reported as the mean cumulative active puncture time per participant.
|
From initial needle-to-skin contact until confirmation of successful radial artery cannulation during the index procedure
|
|
Mean time required for catheter fixation after successful radial artery cannulation
Lasso di tempo: From confirmation of successful radial artery cannulation until completion of catheter fixation and dressing during the index procedure
|
Catheter fixation time is defined as the time, measured in seconds, from confirmation of successful radial artery cannulation by an arterial pressure waveform until the catheter is securely fixed and the dressing is completed.
The outcome will be reported as the mean catheter fixation time per participant.
|
From confirmation of successful radial artery cannulation until completion of catheter fixation and dressing during the index procedure
|
|
Percentage of participants with posterior wall puncture
Lasso di tempo: From the first puncture attempt until confirmation of successful radial artery cannulation during the index procedure
|
Posterior wall puncture is defined as visualization of the needle tip penetrating the posterior wall of the radial artery during any puncture attempt.
A participant will be classified as having posterior wall puncture if at least one such event occurs during the procedure.
The event will be identified by the operator during the procedure or confirmed through review of the recorded ultrasound images or video.
The outcome will be reported as the percentage of participants with at least one posterior wall puncture.
|
From the first puncture attempt until confirmation of successful radial artery cannulation during the index procedure
|
|
Mean systolic blood pressure before the first puncture attempt
Lasso di tempo: Within 1 minute before the first puncture attempt of the index procedure
|
Systolic blood pressure will be measured in millimeters of mercury within 1 minute before the first puncture attempt.
The outcome will be reported as the mean systolic blood pressure per participant.
|
Within 1 minute before the first puncture attempt of the index procedure
|
|
Mean diastolic blood pressure before the first puncture attempt
Lasso di tempo: Within 1 minute before the first puncture attempt of the index procedure
|
Diastolic blood pressure will be measured in millimeters of mercury within 1 minute before the first puncture attempt.
The outcome will be reported as the mean diastolic blood pressure per participant.
|
Within 1 minute before the first puncture attempt of the index procedure
|
|
Mean heart rate before the first puncture attempt
Lasso di tempo: Within 1 minute before the first puncture attempt of the index procedure
|
Heart rate will be measured in beats per minute within 1 minute before the first puncture attempt.
The outcome will be reported as the mean heart rate per participant.
|
Within 1 minute before the first puncture attempt of the index procedure
|
Collaboratori e investigatori
Studiare le date dei record
Studia le date principali
Inizio studio (Effettivo)
Completamento primario (Stimato)
Completamento dello studio (Stimato)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Altri numeri di identificazione dello studio
- 2026027-001
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .
Prove cliniche su Incannulazione dell'arteria radiale
-
I.R.C.C.S Ospedale Galeazzi-Sant'AmbrogioReclutamentoCoronary Artery DiseaseItalia
-
University of MichiganAgency for Healthcare Research and Quality (AHRQ)CompletatoPolmonite | Sepsi | Malattia cardiovascolare | Toracotomia | Mediastinite | Malattie infettive associate all'assistenza sanitaria | Infezione della ferita superficiale sternale | Infezione sternale profonda | Conduit Harvest o Cannulation SiteStati Uniti
Prove cliniche su Biplanar handheld ultrasound system
-
IRCCS San RaffaeleReclutamentoComplicanza della procedura chirurgica | Complicazione, Postoperatorio | Esofago; Fistola | Fistola; RettaleItalia
-
ReCor Medical, Inc.Reclutamento