Biplanar Handheld Ultrasound for Transradial Artery Puncture
Tuoren Biplanar Combined Handheld Ultrasound Versus Conventional Ultrasound for Transradial Artery Puncture A Prospective Randomized Parallel Single-Center Superiority Clinical Study
Radial artery puncture is commonly used to obtain vascular access for cerebral angiography and neurointerventional procedures. Conventional ultrasound guidance usually provides either a short-axis or a long-axis view of the radial artery. A biplanar handheld ultrasound system can provide simultaneous long-axis and short-axis views, which may help the operator visualize the puncture needle and radial artery more clearly.
This prospective, single-center, randomized, parallel-group superiority trial will compare the TUOREN biplanar handheld ultrasound system with a conventional high-frequency linear-array ultrasound system for ultrasound-guided radial artery puncture. A total of 152 adults scheduled to undergo cerebral angiography or neurointerventional treatment through a transradial approach will be randomly assigned in a 1:1 ratio to one of the two ultrasound-guidance groups.
The main objective is to determine whether biplanar handheld ultrasound guidance improves the percentage of participants with successful radial artery cannulation on the first puncture attempt. Other assessments include the total number of puncture attempts, cumulative active puncture time, fixation time after successful cannulation, and the occurrence of posterior wall puncture.
Studieoversigt
Status
Betingelser
Intervention / Behandling
Detaljeret beskrivelse
Ultrasound guidance is used to improve visualization of the radial artery during vascular access. Conventional high-frequency linear-array ultrasound systems provide either a short-axis or a long-axis image, depending on probe orientation and puncture technique. During puncture, the operator may need to adjust the probe repeatedly to maintain visualization of the needle tip and its position relative to the radial artery.
The TUOREN biplanar handheld ultrasound system uses a T-shaped dual-probe configuration to display simultaneous long-axis and short-axis views of the radial artery and puncture needle. The long-axis view is intended to show the needle trajectory and advancement, while the short-axis view is intended to show the lateral position of the needle relative to the vessel.
This is a prospective, randomized, parallel-group, single-center superiority trial conducted at Shanghai Fourth People's Hospital. Eligible participants scheduled for cerebral angiography or neurointerventional treatment through a transradial approach will be allocated in a 1:1 ratio using block randomization with allocation concealment.
Participants in the experimental group will undergo radial artery puncture guided by the TUOREN biplanar handheld ultrasound system. Participants in the active comparator group will undergo radial artery puncture guided by a conventional high-frequency linear-array ultrasound system using either a short-axis out-of-plane technique or a long-axis in-plane technique. Other components of the radial artery cannulation procedure will be standardized between the two groups.
The study is designed to evaluate whether simultaneous biplanar ultrasound visualization improves first-attempt radial artery cannulation success and procedural efficiency compared with conventional ultrasound guidance. Puncture attempts, procedure-related time intervals, and ultrasound findings will be recorded according to prespecified study definitions.
Undersøgelsestype
Tilmelding (Anslået)
Fase
- Ikke anvendelig
Kontakter og lokationer
Studiesteder
-
-
Shanghai Municipality
-
Shanghai, Shanghai Municipality, Kina, 200435
- Shanghai Fourth People's Hospital
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
- Voksen
- Ældre voksen
Tager imod sunde frivillige
Beskrivelse
Inclusion Criteria:
- Adults aged 18 years or older, regardless of sex, who are scheduled to undergo cerebral angiography or neurointerventional treatment through a transradial approach.
- Adequate collateral circulation between the radial and ulnar arteries, as demonstrated by the modified Allen test or vascular ultrasound.
- Radial artery diameter of at least 1.5 mm on preprocedural ultrasound examination.
- Ability to understand the study procedures, willingness to participate voluntarily, and provision of written informed consent.
Exclusion Criteria:
- Infection or skin damage at the intended puncture site; known radial artery occlusion or severe stenosis; or inadequate collateral circulation between the radial and ulnar arteries.
- Radial artery diameter of less than 1.5 mm on preprocedural ultrasound examination, or severe radial artery tortuosity or anatomical variation that may interfere with the puncture procedure.
- Previous puncture, surgery, or trauma involving the radial artery on the intended puncture side that may have altered the local anatomy.
- Hemodynamic instability requiring urgent hemodynamic support, including cardiogenic shock or severe hypotension.
- Impaired consciousness, psychiatric illness, or any other condition that prevents the participant from cooperating with the study procedures or completing the required follow-up.
- Severe coagulation abnormality, defined as an international normalized ratio greater than 1.8 or a platelet count below 80 × 10^9/L.
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Andet
- Tildeling: Randomiseret
- Interventionel model: Parallel tildeling
- Maskning: Ingen (Åben etiket)
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
|---|---|
|
Eksperimentel: TUOREN Biplanar Handheld Ultrasound Guidance
Participants assigned to this arm will undergo transradial artery puncture guided by TUOREN biplanar handheld ultrasound which provides simultaneous long-axis and short-axis views.
|
A biplanar handheld ultrasound system used to guide radial artery puncture.
Its T-shaped dual-probe configuration provides simultaneous long-axis and short-axis views of the radial artery and puncture needle.
|
|
Aktiv komparator: Conventional Ultrasound Guidance
Participants assigned to this arm will undergo transradial artery puncture guided by a conventional high-frequency linear array ultrasound probe using short-axis out-of-plane or long-axis in-plane techniques.
|
A conventional high-frequency linear-array ultrasound system used to guide radial artery puncture using either a short-axis out-of-plane technique or a long-axis in-plane technique.
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
First-attempt success rate
Tidsramme: During the puncture procedure
|
Proportion of participants with successful radial artery cannulation on the first continuous puncture attempt.
Success is confirmed by sustained blood return guidewire advancement at least 5 to 10 cm without resistance and an arterial pressure waveform after catheter connection.
|
During the puncture procedure
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Mean total number of puncture attempts required for successful radial artery cannulation
Tidsramme: From initial needle-to-skin contact until confirmation of successful radial artery cannulation during the index procedure
|
The total number of puncture attempts required to achieve successful radial artery cannulation will be recorded for each participant.
An attempt begins when the needle first contacts the skin and ends when successful cannulation is confirmed or when the needle is completely withdrawn from the skin.
Each reinsertion after complete withdrawal of the needle from the skin is counted as a new attempt.
The outcome will be reported as the mean number of puncture attempts per participant.
|
From initial needle-to-skin contact until confirmation of successful radial artery cannulation during the index procedure
|
|
Mean cumulative active puncture time to successful radial artery cannulation
Tidsramme: From initial needle-to-skin contact until confirmation of successful radial artery cannulation during the index procedure
|
Cumulative active puncture time is defined as the total active puncture time, measured in seconds, from the first needle-to-skin contact until successful radial artery cannulation is confirmed by an arterial pressure waveform.
Timing will be paused whenever the needle is completely withdrawn from the skin and resumed when the needle contacts the skin for the next attempt.
Time used for skin preparation, draping, ultrasound probe preparation, and intervals between puncture attempts will not be included.
The outcome will be reported as the mean cumulative active puncture time per participant.
|
From initial needle-to-skin contact until confirmation of successful radial artery cannulation during the index procedure
|
|
Mean time required for catheter fixation after successful radial artery cannulation
Tidsramme: From confirmation of successful radial artery cannulation until completion of catheter fixation and dressing during the index procedure
|
Catheter fixation time is defined as the time, measured in seconds, from confirmation of successful radial artery cannulation by an arterial pressure waveform until the catheter is securely fixed and the dressing is completed.
The outcome will be reported as the mean catheter fixation time per participant.
|
From confirmation of successful radial artery cannulation until completion of catheter fixation and dressing during the index procedure
|
|
Percentage of participants with posterior wall puncture
Tidsramme: From the first puncture attempt until confirmation of successful radial artery cannulation during the index procedure
|
Posterior wall puncture is defined as visualization of the needle tip penetrating the posterior wall of the radial artery during any puncture attempt.
A participant will be classified as having posterior wall puncture if at least one such event occurs during the procedure.
The event will be identified by the operator during the procedure or confirmed through review of the recorded ultrasound images or video.
The outcome will be reported as the percentage of participants with at least one posterior wall puncture.
|
From the first puncture attempt until confirmation of successful radial artery cannulation during the index procedure
|
|
Mean systolic blood pressure before the first puncture attempt
Tidsramme: Within 1 minute before the first puncture attempt of the index procedure
|
Systolic blood pressure will be measured in millimeters of mercury within 1 minute before the first puncture attempt.
The outcome will be reported as the mean systolic blood pressure per participant.
|
Within 1 minute before the first puncture attempt of the index procedure
|
|
Mean diastolic blood pressure before the first puncture attempt
Tidsramme: Within 1 minute before the first puncture attempt of the index procedure
|
Diastolic blood pressure will be measured in millimeters of mercury within 1 minute before the first puncture attempt.
The outcome will be reported as the mean diastolic blood pressure per participant.
|
Within 1 minute before the first puncture attempt of the index procedure
|
|
Mean heart rate before the first puncture attempt
Tidsramme: Within 1 minute before the first puncture attempt of the index procedure
|
Heart rate will be measured in beats per minute within 1 minute before the first puncture attempt.
The outcome will be reported as the mean heart rate per participant.
|
Within 1 minute before the first puncture attempt of the index procedure
|
Samarbejdspartnere og efterforskere
Datoer for undersøgelser
Studer store datoer
Studiestart (Faktiske)
Primær færdiggørelse (Anslået)
Studieafslutning (Anslået)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Nøgleord
Andre undersøgelses-id-numre
- 2026027-001
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Radial arteriekanylering
-
University Hospital of PatrasRekrutteringDistal Radial Artery Access (dTRA) | Adgang til radial arterieGrækenland
-
Inova Health Care ServicesBoston Scientific CorporationAfsluttetKoronar angiografi | Transradial adgang | Radial arterie Intimal Medial Tykkelse | Distal Radial Artery Access (dTRA)Forenede Stater
-
University Hospital OstravaRekrutteringIn-Stent Carotis Artery RestenosisTjekkiet
-
Zhejiang Cancer HospitalRekrutteringHepatic Artery Infusion | Levermetastase fra BrystkræftKina
-
IRCCS Policlinico S. DonatoRekrutteringAnomalous aorta origin of the coronary artery (AAOCA)Italien
-
Peking University Third HospitalRekrutteringRadial arterie | Quantitative Flow Reserve | Minimally Invasive Coronary SurgeryKina
-
King's College LondonRoyal Free Hospital NHS Foundation Trust; Chelsea and Westminster NHS Foundation... og andre samarbejdspartnereRekruttering
-
Ting XuTilmelding efter invitationRadial arterieskadeKina
-
Cairo UniversityAfsluttetRadial arteriekanulationEgypten
-
Oregon Health and Science UniversityAfsluttet
Kliniske forsøg med Biplanar handheld ultrasound system
-
ReCor Medical, Inc.RekrutteringForhøjet blodtrykØstrig, Schweiz, Tyskland, Belgien, Holland, Det Forenede Kongerige, Frankrig, Monaco, Spanien
-
Otsuka Medical Devices Co., Ltd. JapanAfsluttetKronisk hjertesvigtJapan
-
Sunnybrook Health Sciences CentreTerry Fox Research InstituteRekruttering
-
JIMRO Co., Ltd.AfsluttetKarsygdomme | Forhøjet blodtrykKorea, Republikken, Japan
-
University of CoimbraFundação para a Ciência e a TecnologiaIkke rekrutterer endnuGrå stær | Ultralydsterapi; KomplikationerPortugal
-
Otsuka Medical Devices Co., Ltd. JapanAktiv, ikke rekrutterendeHjerte-kar-sygdomme | Karsygdomme | Forhøjet blodtrykJapan
-
McMaster UniversityCanadian Institutes of Health Research (CIHR); Bioventus LLCAfsluttet
-
ICE Intelligent Healthcare Co., LtdIkke rekrutterer endnuAtrieflimren | Arytmier, hjerte | Atriel septaldefekt | Patent Foramen Ovale | Ventilsygdom, hjerte | Supraventrikulær takykardi | Strukturelle hjertesygdommeKina
-
Innovative MedicalAfsluttet
-
ReCor Medical, Inc.Aktiv, ikke rekrutterendeKarsygdomme | Forhøjet blodtrykForenede Stater, Holland, Det Forenede Kongerige, Tyskland, Frankrig, Polen, Belgien