Effect of High-Flow Humidified Oxygen on Ventilatory Work During the Spontaneous Breathing Trial (OTRAVAIL)

20 luglio 2026 aggiornato da: University Hospital, Grenoble

Weaning from mechanical ventilation (MV) accounts for a significant portion of the total MV duration, and failure to extubate or delayed extubation increases complications (particularly Ventilator-Associated Pneumonia (VAP)) and mortality. Standard tests (T-piece, Spontaneous Ventilation with Inspiratory Assistance (SV-IA)) have not demonstrated clear superiority, and the T-piece is still widely used despite its limitations. High-flow oxygen (HFO) may reduce ventilatory work and facilitate secretion clearance; however, there is a lack of physiological data in humans with an endotracheal tube in place. This study aims to address this gap prior to a subsequent clinical evaluation.

An experimental, prospective, single-center, 3-period crossover study with randomized sequence order and single-blind design (primary endpoint assessment and statistical analysis were conducted in a blinded manner).

Panoramica dello studio

Tipo di studio

Interventistico

Iscrizione (Stimato)

30

Fase

  • Non applicabile

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

  • Adulto
  • Adulto più anziano

Accetta volontari sani

No

Descrizione

Inclusion Criteria:

  • Adult patient, of legal age
  • On invasive mechanical ventilation for more than 48 hours
  • With at least one criterion indicating a high risk of extubation failure:

    • Age > 65 years
    • Chronic respiratory disease
    • Chronic cardiac disease
  • Capable of undergoing a mechanical ventilation weaning test ("weaning criterion"):

    • Respiratory rate < 35 bpm
    • Adequate oxygenation:

      • SpO₂ ≥ 90% with FiO₂ ≤ 40% and PEEP ≤ 8 cm H₂O
      • Or PaO₂/FiO₂ ≥ 150 with PEEP ≤ 8 cm H₂O
    • Effective cough
    • Patient is alert (RASS score -2 to +1), without sedation
    • Hemodynamically stable or on the minimum effective dose
  • Patient has an invasive arterial pressure catheter for arterial blood sampling
  • Informed consent signed by a family member
  • Patient is enrolled in or eligible for a social security program

Exclusion Criteria:

  • Resuscitation-associated neuromyopathy
  • History of neuromuscular disease
  • Admission for a neurological condition
  • Subjects currently excluded from another study,
  • Persons referred to in Articles L1121-5 through L1121-8 of the CSP (corresponding to all protected persons: pregnant women, women in labor, breastfeeding mothers, persons deprived of liberty by judicial or administrative decision, persons receiving psychiatric care pursuant to Articles L. 3212-1 and L. 3213-1 who are not covered by the provisions of Article L. 1121-8, persons admitted to a health or social care facility for purposes other than research, minors, and persons subject to a legal protection order or unable to give informed consent).
  • Staff members who report to the principal investigator

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Scopo principale: Trattamento
  • Assegnazione: Randomizzato
  • Modello interventistico: Assegnazione incrociata
  • Mascheramento: Doppio

Armi e interventi

Gruppo di partecipanti / Arm
Intervento / Trattamento
Altro: TP/HFO/PSV
T piece follow by HFO follow by PSV
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph. The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph. The OHD flow rate will be set to 50 L/min. A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
Altro: TP/PSV/HFO
T piece follow by PSV follow by HFO
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph. The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph. The OHD flow rate will be set to 50 L/min. A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
Altro: PSV/HFO/TP
PSV follow by HFO follow by T piece
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph. The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph. The OHD flow rate will be set to 50 L/min. A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
Altro: PSV/TP/HFO
PSV follow by T piece follow by HFO
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph. The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph. The OHD flow rate will be set to 50 L/min. A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
Altro: HFO/TP/PSV
HFO follow by T piece follow by PSV
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph. The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph. The OHD flow rate will be set to 50 L/min. A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
Altro: HFO/PSV/TP
HFO follow by PSV follow by TP
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph. The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph. The OHD flow rate will be set to 50 L/min. A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Lasso di tempo
Respiratory work during the weaning test, defined as the energy expended to inhale and exhale a respiratory gas, measured in joules per liter per minute.
Lasso di tempo: 190 min
190 min

Misure di risultato secondarie

Misura del risultato
Lasso di tempo
Tidal Volume
Lasso di tempo: 190 min
190 min
Total Positive End-Expiratory Pressure
Lasso di tempo: 190 min
190 min
Respiratory Rate
Lasso di tempo: 190 min
190 min
minute ventilation
Lasso di tempo: 190 min
190 min
Ratio Respiratory rate/ Tidal volume
Lasso di tempo: 190 min
190 min
product pressure time
Lasso di tempo: 190 min
190 min
Occlusion pressure (or P 0.1)
Lasso di tempo: 190 min
190 min
Percentage of success on the weaning test (predefined criteria)
Lasso di tempo: 190 min
190 min

Altre misure di risultato

Misura del risultato
Lasso di tempo
Percentage of patients extubated following the series of weaning trials
Lasso di tempo: 190 min
190 min
Percentage of reintubations within 48 hours after extubation
Lasso di tempo: 52 hours
52 hours

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Stimato)

1 gennaio 2027

Completamento primario (Stimato)

1 gennaio 2027

Completamento dello studio (Stimato)

1 gennaio 2027

Date di iscrizione allo studio

Primo inviato

20 luglio 2026

Primo inviato che soddisfa i criteri di controllo qualità

20 luglio 2026

Primo Inserito (Effettivo)

24 luglio 2026

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

24 luglio 2026

Ultimo aggiornamento inviato che soddisfa i criteri QC

20 luglio 2026

Ultimo verificato

1 luglio 2026

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • 38RC25.0307

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

No

prodotto fabbricato ed esportato dagli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .