Effect of High-Flow Humidified Oxygen on Ventilatory Work During the Spontaneous Breathing Trial (OTRAVAIL)
Weaning from mechanical ventilation (MV) accounts for a significant portion of the total MV duration, and failure to extubate or delayed extubation increases complications (particularly Ventilator-Associated Pneumonia (VAP)) and mortality. Standard tests (T-piece, Spontaneous Ventilation with Inspiratory Assistance (SV-IA)) have not demonstrated clear superiority, and the T-piece is still widely used despite its limitations. High-flow oxygen (HFO) may reduce ventilatory work and facilitate secretion clearance; however, there is a lack of physiological data in humans with an endotracheal tube in place. This study aims to address this gap prior to a subsequent clinical evaluation.
An experimental, prospective, single-center, 3-period crossover study with randomized sequence order and single-blind design (primary endpoint assessment and statistical analysis were conducted in a blinded manner).
Visão geral do estudo
Status
Condições
Intervenção / Tratamento
Tipo de estudo
Inscrição (Estimado)
Estágio
- Não aplicável
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
- Adulto
- Adulto mais velho
Aceita Voluntários Saudáveis
Descrição
Inclusion Criteria:
- Adult patient, of legal age
- On invasive mechanical ventilation for more than 48 hours
With at least one criterion indicating a high risk of extubation failure:
- Age > 65 years
- Chronic respiratory disease
- Chronic cardiac disease
Capable of undergoing a mechanical ventilation weaning test ("weaning criterion"):
- Respiratory rate < 35 bpm
Adequate oxygenation:
- SpO₂ ≥ 90% with FiO₂ ≤ 40% and PEEP ≤ 8 cm H₂O
- Or PaO₂/FiO₂ ≥ 150 with PEEP ≤ 8 cm H₂O
- Effective cough
- Patient is alert (RASS score -2 to +1), without sedation
- Hemodynamically stable or on the minimum effective dose
- Patient has an invasive arterial pressure catheter for arterial blood sampling
- Informed consent signed by a family member
- Patient is enrolled in or eligible for a social security program
Exclusion Criteria:
- Resuscitation-associated neuromyopathy
- History of neuromuscular disease
- Admission for a neurological condition
- Subjects currently excluded from another study,
- Persons referred to in Articles L1121-5 through L1121-8 of the CSP (corresponding to all protected persons: pregnant women, women in labor, breastfeeding mothers, persons deprived of liberty by judicial or administrative decision, persons receiving psychiatric care pursuant to Articles L. 3212-1 and L. 3213-1 who are not covered by the provisions of Article L. 1121-8, persons admitted to a health or social care facility for purposes other than research, minors, and persons subject to a legal protection order or unable to give informed consent).
- Staff members who report to the principal investigator
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Tratamento
- Alocação: Randomizado
- Modelo Intervencional: Atribuição cruzada
- Mascaramento: Dobro
Armas e Intervenções
Grupo de Participantes / Braço |
Intervenção / Tratamento |
|---|---|
|
Outro: TP/HFO/PSV
T piece follow by HFO follow by PSV
|
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph.
The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph.
The OHD flow rate will be set to 50 L/min.
A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
|
|
Outro: TP/PSV/HFO
T piece follow by PSV follow by HFO
|
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph.
The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph.
The OHD flow rate will be set to 50 L/min.
A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
|
|
Outro: PSV/HFO/TP
PSV follow by HFO follow by T piece
|
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph.
The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph.
The OHD flow rate will be set to 50 L/min.
A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
|
|
Outro: PSV/TP/HFO
PSV follow by T piece follow by HFO
|
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph.
The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph.
The OHD flow rate will be set to 50 L/min.
A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
|
|
Outro: HFO/TP/PSV
HFO follow by T piece follow by PSV
|
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph.
The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph.
The OHD flow rate will be set to 50 L/min.
A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
|
|
Outro: HFO/PSV/TP
HFO follow by PSV follow by TP
|
The endotracheal tube will be disconnected from the ventilator, and a T-piece (Teleflex, Trach-Vent +®) with oxygen delivery
The endotracheal tube will be connected to the pneumotachograph.
The ventilator settings will be adjusted as follows: VS mode, inspiratory support at 7 cmH₂O, PEEP at 0 cmH₂O.
The endotracheal tube will be disconnected from the ventilator, and an OHD delivery device normally used for tracheostomies (Fisher & Paykel, Optiflow+®) will be connected to the pneumotachograph.
The OHD flow rate will be set to 50 L/min.
A Monnal T60 ventilator (Air Liquide Healthcare) will be used for OHD administration.
|
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Prazo |
|---|---|
|
Respiratory work during the weaning test, defined as the energy expended to inhale and exhale a respiratory gas, measured in joules per liter per minute.
Prazo: 190 min
|
190 min
|
Medidas de resultados secundários
Medida de resultado |
Prazo |
|---|---|
|
Tidal Volume
Prazo: 190 min
|
190 min
|
|
Total Positive End-Expiratory Pressure
Prazo: 190 min
|
190 min
|
|
Respiratory Rate
Prazo: 190 min
|
190 min
|
|
minute ventilation
Prazo: 190 min
|
190 min
|
|
Ratio Respiratory rate/ Tidal volume
Prazo: 190 min
|
190 min
|
|
product pressure time
Prazo: 190 min
|
190 min
|
|
Occlusion pressure (or P 0.1)
Prazo: 190 min
|
190 min
|
|
Percentage of success on the weaning test (predefined criteria)
Prazo: 190 min
|
190 min
|
Outras medidas de resultado
Medida de resultado |
Prazo |
|---|---|
|
Percentage of patients extubated following the series of weaning trials
Prazo: 190 min
|
190 min
|
|
Percentage of reintubations within 48 hours after extubation
Prazo: 52 hours
|
52 hours
|
Colaboradores e Investigadores
Patrocinador
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Estimado)
Conclusão Primária (Estimado)
Conclusão do estudo (Estimado)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Outros números de identificação do estudo
- 38RC25.0307
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
produto fabricado e exportado dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .