Surface Refractive Index Shift of Kalifilcon A Compared With Dailies Total1 and Precision1 Lenses

2026年8月3日 更新者:Bausch & Lomb Incorporated

Evaluation of Surface Refractive Index Shift and Lens Absorption / Adsorption of Tear Components of Commercially Available Kalifilcon A Lenses Compared to Dailies Total1 and Precision1

This was a single-site, bilateral, repeated-measures insertion study in habitual soft contact lens wearers. Ten participants were enrolled. All participants attended a Screening/Dispensing Visit at which informed consent was obtained and eligibility was assessed. Eligible participants then wore each of three commercially available daily disposable soft contact lens types - kalifilcon A (Bausch + Lomb), Dailies Total1 (delefilcon A, Alcon) and Precision1 (verofilcon A, Alcon), all -3.00 D - bilaterally for wear times of 0.5, 1, 3, 5 and 10 minutes over a total of 6 study visits. A new lens pair was inserted for each wear time, with a 3-5 minute washout period between lens pairs. Immediately following removal, the right-eye lens was measured for surface refractive index on a Metricon M-2010 Prism Coupler.

調査の概要

状態

完了

条件

介入・治療

詳細な説明

Rationale: earlier work (ROC2-20-012) showed a change in surface refractive index for Dailies Total1 (delefilcon A) and Precision1 (verofilcon A) lenses after 5 minutes of wear, attributed to a change in surface water content. ROC2-21-007 was designed to determine whether that change occurs in less than 5 minutes. Three study lenses (one test, two controls) were each worn bilaterally. Wear times of 0.5, 1 and 3 minutes were completed at one visit and 5 and 10 minutes at another visit for each lens type, giving 6 visits of approximately 30 minutes each, each on a separate day; the clinical investigation took approximately 2 weeks. Lens type order followed the study randomization schedule prepared before enrolment by an unmasked statistician not otherwise involved in the study; unmasked study personnel dispensed and collected study materials so that the participant and Investigator remained masked. Endpoint: surface refractive index of the worn right-eye lens, measured on a Metricon M-2010 Prism Coupler, compared with unworn (baseline) values for the same lens types. Continuous data were summarized with n, mean, standard deviation, median, minimum and maximum, and paired t-tests were used to compare unworn with worn values, with p < 0.05 considered statistically significant.

研究の種類

介入

入学 (実際)

10

段階

  • 適用できない

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究場所

    • New York
      • Rochester、New York、アメリカ、14609
        • Bausch & Lomb Incorporated

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

18年歳以上 (大人、高齢者)

健康ボランティアの受け入れ

はい

説明

Inclusion Criteria:

  1. Be 18 years or older and have full legal capacity to volunteer.
  2. Have physiologically normal anterior segments not exhibiting clinically significant biomicroscopy findings.
  3. Have no active ocular disease or allergic conjunctivitis.
  4. Not be using any topical ocular medications.
  5. Be willing and able to follow instructions.
  6. Have signed a statement of informed consent.

Exclusion Criteria:

  1. Participating in a conflicting study in the opinion of the Investigator.
  2. Considered by the Investigator to not be a suitable candidate for participation.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

  • 主な目的:基礎科学
  • 割り当て:なし
  • 介入モデル:単一グループの割り当て
  • マスキング:なし(オープンラベル)

武器と介入

参加者グループ / アーム
介入・治療
他の:All Participants: kalifilcon A, Then Dailies Total1, Then Precision1
Each participant wore all three study lens types bilaterally, one lens type per pair of visits, in the order specified by the study randomization schedule (kalifilcon A, then Dailies Total1, then Precision1). For each lens type a new lens pair was inserted for each wear time of 0.5, 1, 3, 5 and 10 minutes, with a 3-5 minute washout period between lens pairs. Wear times of 0.5, 1 and 3 minutes were completed at one visit and 5 and 10 minutes at another visit, for a total of 6 visits. Immediately after removal the right-eye lens was measured for surface refractive index on a Metricon M-2010 Prism Coupler.
Commercially available Infuse (kalifilcon A) daily disposable spherical soft contact lens, -3.00 D, worn bilaterally for 0.5, 1, 3, 5 and 10 minutes.
他の名前:
  • 注入する
Commercially available Dailies Total1 (delefilcon A) daily disposable spherical soft contact lens (Alcon), -3.00 D, worn bilaterally for 0.5, 1, 3, 5 and 10 minutes.
Commercially available Precision1 (verofilcon A) daily disposable spherical soft contact lens (Alcon), -3.00 D, worn bilaterally for 0.5, 1, 3, 5 and 10 minutes.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Surface Refractive Index of Worn Lenses
時間枠:Unworn lens (baseline) and immediately after 0.5, 1, 3, 5, and 10 minutes of lens wear
Immediately following lens removal, the right-eye worn lens for each participant was measured for surface refractive index on a Metricon M-2010 Prism Coupler. A new lens pair was worn for each wear time. Surface refractive index is a dimensionless ratio and has no units.
Unworn lens (baseline) and immediately after 0.5, 1, 3, 5, and 10 minutes of lens wear

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

捜査官

  • 主任研究者:Jeffery Schafer, OD, MS、Bausch & Lomb Incorporated

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2021年4月7日

一次修了 (実際)

2021年4月15日

研究の完了 (実際)

2021年4月15日

試験登録日

最初に提出

2021年4月7日

QC基準を満たした最初の提出物

2021年4月8日

最初の投稿 (実際)

2021年4月12日

学習記録の更新

投稿された最後の更新 (実際)

2026年8月5日

QC基準を満たした最後の更新が送信されました

2026年8月3日

最終確認日

2026年8月1日

詳しくは

本研究に関する用語

キーワード

その他の研究ID番号

  • ROC2-21-007

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

IPD プランの説明

Individual participant data will not be shared. This was an internal Bausch + Lomb device performance study; source data are retained in the Trial Master File.

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

はい

米国で製造され、米国から輸出された製品。

いいえ

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。