Operator-Assisted Oscilla Audiometry as a Basis for Hearing Aid Fitting (The Oscilla Validation Study)
Validation of Operator-Assisted Oscilla Audiometry as a Diagnostic Tool and Basis for Hearing Aid Fitting in Adults With Uncomplicated Hearing Loss: A Blinded Non-Inferiority Randomised Controlled Trial
The purpose of this study is to find out whether hearing tests performed with the Oscilla A60 audiometer, where the patient largely tests themselves with minimal staff assistance, are accurate enough to be used for diagnosing hearing loss and as the sole basis for fitting hearing aids. Adults referred for their first hearing aids at Aalborg University Hospital will have their hearing tested both with the Oscilla audiometer and by a trained audiology assistant at the same visit. Participants are then randomly assigned to receive hearing aids fitted based on either the Oscilla hearing test or the audiology assistant's hearing test. The main questions are: (1) Do the Oscilla results agree with the audiology assistant's results for both air-conduction and bone-conduction hearing thresholds? (2) Do participants fitted based on the Oscilla test report a hearing benefit that is not worse than that of participants fitted based on the standard test, two months after fitting?
調査の概要
状態
状態
条件
条件
介入・治療
介入・治療
詳細な説明
Demand for audiometric testing in public hearing clinics is increasing due to population ageing. Operator-assisted automated audiometry may expand capacity, but its use as the sole basis for hearing aid prescription, and its accuracy for bone-conduction (BC) thresholds, has not been established. This single-site study at the Department of Audiology, Aalborg University Hospital (locations in Aalborg and Hobro) enrols 268 first-time hearing aid users with uncomplicated, symmetric sensorineural hearing loss.
At Visit 1, all participants undergo both Oscilla A60 audiometry (air conduction [AC] 250-8000 Hz; BC 250-4000 Hz) and standard manual audiometry by an audiology assistant (reference standard), in randomised order. Agreement between methods is assessed with Bland-Altman analysis against pre-specified acceptance criteria (95% CI of bias within ±5 dB; 95% limits of agreement within ±15 dB for AC and ±20 dB for BC). After confirmation of eligibility, participants are randomised 1:1, stratified by hearing loss severity (PTA-4 ≤45 vs >45 dB HL) and location, to hearing aid fitting based on the Oscilla audiogram or the audiology assistant audiogram. All participants are fitted bilaterally with BTE/RIC hearing aids to the NAL-NL2 target, verified with automated real-ear measurements (within ±5 dB). Volume control and app connectivity are disabled until the two-month follow-up. The primary patient-reported outcome is the change in SSQ12 total score from baseline to two months post-fitting, with a non-inferiority margin of -0.81 points. Secondary outcomes are IOI-HA, daily hearing aid use (data logging) and SSQ12 subscales. The study is a clinical investigation under Article 82 of Regulation (EU) 2017/745.
研究の種類
研究の種類
入学 (推定)
入学
段階
段階
- 適用できない
連絡先と場所
研究連絡先
研究連絡先
- 名前:Lene D Siggaard, MD, PhD
- 電話番号:+45 20633492
- メール:lene.dahl@rn.dk
研究連絡先のバックアップ
- 名前:Camilla L Sørensen, Cand. med.
- メール:c.lorents@rn.dk
研究場所
-
-
-
Aalborg、デンマーク、9000
- Department of Audiology, Aalborg University Hospital
-
コンタクト:
- Camilla L Sørensen, Cand. med.
- メール:c.lorents@rn.dk
-
コンタクト:
- Lene D Siggaard, MD, PhD
- 電話番号:+4520633492
- メール:lene.dahl@rn.dk
-
Hobro、デンマーク、9500
- Hearing Clinic, Aalborg University Hospital
-
コンタクト:
- Camilla L Sørensen, Cand. med.
- メール:c.lorents@rn.dk
-
コンタクト:
- Lene D Siggaard, MD, PhD
- 電話番号:+4520633492
- メール:lene.dahl@rn.dk
-
-
参加基準
適格基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Adults (≥18 years) referred for first-time hearing aid fitting at the Department of Audiology, Aalborg University Hospital
- Uncomplicated sensorineural hearing loss as defined by the Danish national clinical guidelines
- Hearing thresholds exceeding 20 dB HL at two or more frequencies in the range 0.5-4 kHz
- Pure-tone average (PTA-4; 0.5, 1, 2, 4 kHz) ≥30 dB HL in the better ear
- Symmetric hearing loss (PTA-4 difference between ears ≤15 dB)
- Word discrimination score ≥25% (DANTALE I at PTA + 40 dB HL)
- Danish native speaker
- No prior hearing aid experience
- Able to complete questionnaires online or on paper
- Written informed consent
Exclusion Criteria:
- Air-conduction thresholds exceeding 80 dB HL at two or more frequencies
- Air-bone gap >15 dB at one or more frequencies below 1 kHz
- Fluctuating hearing loss (e.g., Ménière's disease) or ongoing ototoxic medication
- Ear surgery (ear canal, middle ear or inner ear) within the past 12 months
- Hearing loss classified as complicated according to national clinical guidelines
- Visual or motor impairment precluding independent use of the Oscilla system
- Candidates for in-the-ear hearing aids only (all participants are fitted with BTE or RIC hearing aids)
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:診断
- 割り当て:ランダム化
- 介入モデル:並列代入
- マスキング:独身
アーム数
武器と介入
参加者グループ / アーム参加者グループ / アーム |
介入・治療介入・治療 |
|---|---|
|
実験的:Oscilla-based fitting
Hearing aids fitted bilaterally to NAL-NL2 targets based on the Oscilla A60 audiogram, verified with automated real-ear measurement. Volume control and app connectivity disabled for two months (linked interventions: Paired audiometry and Oscilla-based hearing aid fitting). |
At Visit 1, all participants undergo both operator-assisted Oscilla A60 audiometry (AudioConsole v4.6.0; AC and BC thresholds; operator places the BC headband only) in a quiet room (<50 dB SPL) and manual pure-tone audiometry (AC and BC, ISO 8253-1) by a qualified audiology assistant in a sound-attenuating booth, in randomised order. Bilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the Oscilla A60 audiogram as the sole audiometric basis, verified with auto-REM (within ±5 dB of target). Other names: Oscilla A60 Diagnostic Audiometer. |
|
アクティブコンパレータ:Manual audiometry-based fitting
Hearing aids fitted bilaterally to NAL-NL2 targets based on the audiogram obtained by a qualified audiology assistant, verified with automated real-ear measurement. Volume control and app connectivity disabled for two months (linked interventions: paired audiometry and manual audiometry-based hearing aid fitting). |
At Visit 1, all participants undergo both operator-assisted Oscilla A60 audiometry (AudioConsole v4.6.0; AC and BC thresholds; operator places the BC headband only) in a quiet room (<50 dB SPL) and manual pure-tone audiometry (AC and BC, ISO 8253-1) by a qualified audiology assistant in a sound-attenuating booth, in randomised order. Bilateral BTE/RIC hearing aid fitting to NAL-NL2 targets using the audiology assistant audiogram, verified with auto-REM (within ±5 dB of target). Standard clinical practice. |
この研究は何を測定していますか?
主要な結果の測定
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Agreement between Oscilla A60 and manual audiometry for air-conduction thresholds
時間枠:Visit 1 (Week 0)
|
Bland-Altman mean difference (Oscilla minus manual) and 95% limits of agreement in dB, plus mean absolute difference, at 0.5, 1, 2 and 4 kHz separately. Agreement is acceptable if the 95% CI of the bias is within ±5 dB and the 95% limits of agreement are within ±15 dB. Both ears included, clustering handled with a mixed-effects model. |
Visit 1 (Week 0)
|
|
Agreement between Oscilla A60 and manual audiometry for bone-conduction thresholds
時間枠:Visit 1 (Week 0)
|
As above for bone-conduction thresholds at 0.5, 1, 2 and 4 kHz. Agreement is acceptable if the 95% CI of the bias is within ±5 dB and the 95% limits of agreement are within ±20 dB. |
Visit 1 (Week 0)
|
|
Change in SSQ12 total score
時間枠:Baseline (Week 0) and 2 months after hearing aid fitting
|
Speech, Spatial and Qualities of Hearing Scale, 12-item version; total score = mean of 12 items, range 0-10, higher = better. Non-inferiority if the lower bound of the two-sided 95% CI for the between-group difference in change (Oscilla minus control) is above -0.81 points. |
Baseline (Week 0) and 2 months after hearing aid fitting
|
二次結果の測定
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
IOI-HA total score
時間枠:2 months after hearing aid fitting
|
International Outcome Inventory for Hearing Aids, Danish version; 7 items scored 1-5, total range 7-35, higher = better outcome. Compared between groups |
2 months after hearing aid fitting
|
|
Daily hearing aid use
時間枠:From hearing aid fitting to 2 months after fitting
|
Mean daily hearing aid usage time in hours/day from the hearing aid data log, compared between groups |
From hearing aid fitting to 2 months after fitting
|
|
Change in SSQ12 subscale scores
時間枠:Baseline (Week 0) and 2 months after hearing aid fitting
|
Change in the speech, spatial and qualities subscale scores (each range 0-10, higher = better) |
Baseline (Week 0) and 2 months after hearing aid fitting
|
協力者と研究者
捜査官
捜査官
- 主任研究者:Lene D Siggaard, MD, PhD、Department of Otolaryngology, Head- and Neck Surgery and Audiology, Aalborg University Hospital
研究記録日
主要日程の研究
研究開始 (推定)
研究開始
一次修了 (推定)
一次修了
研究の完了 (推定)
研究の完了
試験登録日
最初に提出
最初に提出
QC基準を満たした最初の提出物
QC基準を満たした最初の提出物
最初の投稿 (実際)
最初の投稿
学習記録の更新
投稿された最後の更新 (実際)
投稿された最後の更新
QC基準を満たした最後の更新が送信されました
QC基準を満たした最後の更新が送信されました
最終確認日
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
その他の研究ID番号
- F2026-115
- Regional Innovation Fund (その他の助成金/資金番号:North Denmark Region)
- Danish Medicines Agency (レジストリ識別子:16-0302-225)
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
IPD プランの説明
Pseudonymised individual participant data may be shared upon reasonable request after publication, subject to approval under the GDPR and Danish data protection law and a data processing agreement.
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。