Peds Proton Therapy
Operational and Clinical Evaluation of the Treatment Planning Software and Oncology Information System for Pediatric Proton Therapy
This study is being done to evaluate the clinical utilization and safety of RayStation 2025 and RayCare 2025 treatment planning software upgrades when used to support standard-of-care proton therapy delivery in patients deemed clinically appropriate for such therapy by their treating radiation oncologist.
調査の概要
状態
状態
条件
条件
介入・治療
介入・治療
詳細な説明
In this study, the radiation treatment itself, including target definitions, prescription dose, treatment delivery, verification imaging, and adaptive treatment planning, is standard of care.
Current standard radiation therapy planning is based on a treatment plan generated from a planning computed tomography (CT) scan obtained before treatment begins. This plan is generally used for the entire course of therapy, with verification imaging performed during treatment primarily used to verify patient positioning rather than to recalculate delivered dose. On occasion, it is standard practice for a physician to determine, based on his or her interpretation of verification imaging, that an adaptive replanning of the treatment is needed to account for anatomic changes that the physician observes on verification imaging which mandate a change to the plan. Comprehensive software-based evaluation of the cumulative dose delivered over multiple fractions is not routinely performed in most clinical settings. Performing such evaluations typically requires manual steps including image registration, dose recalculation on additional imaging datasets, deformable dose mapping, and manual accumulation of dose across fractions. With currently FDA-approved tools for proton therapy treatment plan evaluation, these processes are time-consuming and require significant physics and physician resources. As a result, detailed adaptive dose evaluation is usually performed only in selected cases or as a one-time assessment during a treatment course, rather than as a routine component of daily clinical care.
The investigational component involves the prospective evaluation of the RayAdaptive and RayTracker software tools to assess their performance, reliability, usability, and clinical feasibility. These software features are not currently available for routine clinical use, being not yet FDA-cleared, in the configuration evaluated in this protocol and are being assessed in a defined prospective clinical research setting.
This is a nonrandomized, single arm, open-label, prospective interventional trial to determine the rate of adaptive replanning for proton therapy when RayTracker and RayAdaptive are used as decision-support tools in the analysis of every verification image obtained during each participant's proton therapy treatment course. This study will evaluate, under a defined prospective clinical research framework, the performance, reliability, usability, and clinical feasibility of implementing RayTracker and RayAdaptive - software components that have not yet been 510(k) cleared by the U.S. Food and Drug Administration (FDA).
Primary Objective
• To determine the frequency of adaptive replanning during proton therapy when RayTracker and RayAdaptive are incorporated into routine verification image review as clinical decision-support tools.
Secondary Objectives
- To evaluate the treating physician-reported rate at which RayTracker and RayAdaptive influenced the decision to pursue adaptive replanning during proton therapy.
- To explore concordance between RayAdaptive-predicted inadequate target coverage on verification imaging and findings on repeat CT simulation when performed.
- To explore concordance between RayAdaptive-predicted organ-at-risk dose constraint violations and findings on repeat CT simulation when performed.
- To report the incidence of adverse events (AEs) and device deficiencies related to the investigational RayTracker/RayAdaptive software.
研究の種類
研究の種類
入学 (推定)
入学
段階
段階
- 適用できない
連絡先と場所
研究連絡先
研究連絡先
- 名前:Research Team
- 電話番号:608-915-1201
- メール:clinicaltrials@cancer.wisc.edu
研究場所
-
-
Wisconsin
-
Madison、Wisconsin、アメリカ、53792
- 募集
- University of Wisconsin
-
-
参加基準
適格基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Age less than or equal to 17 years at the time of consent/assent.
- Patient has a clinical indication for radiation therapy, and the treating radiation oncologist has independently determined that proton therapy is clinically appropriate and has recommended proton therapy as part of the patient's routine clinical care, independent of participation in this research study. Proton therapy may be delivered with or without anesthesia when clinically indicated.
- Written informed consent will be obtained from adult subjects or the parent/legal guardian of pediatric subjects. Assent will be obtained from pediatric participants when appropriate according to institutional policy.
Exclusion Criteria:
- Patient has contraindications to proton therapy (per standard clinical guidelines).
- Patient is unable to comply with routine treatment workflow or follow-up.
- Patients who will be treated with a radiation plan of 5 fractions or fewer.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:処理
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
アーム数
武器と介入
参加者グループ / アーム参加者グループ / アーム |
介入・治療介入・治療 |
|---|---|
|
実験的:Standard of Care Proton Therapy
|
Part of a not yet FDA-cleared software, the RaySearch treatment planning system and RayCare oncology information system. RayTracker enables longitudinal evaluation of treatment by recalculating dose on daily images and accumulating dose across fractions using deformable image registration, thereby providing an estimate of the cumulative delivered dose to targets and organs at risk. RayAdaptive uses daily treatment imaging (such as cone-beam CT (CBCT)) together with the planning CT to generate an updated representation of patient anatomy and recalculation of dose on the current patient anatomy, allowing clinicians to evaluate whether anatomical changes could meaningfully alter the planned dose distribution. These features are designed to be used in tandem. Their combined capabilities are intended to support clinical decision-making related to treatment monitoring and potential adaptation. |
この研究は何を測定していますか?
主要な結果の測定
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Proportion of verification image assessments for each subject resulting in a clinical adaptive replanning decision
時間枠:up to 6 months
|
Defined as ordering repeat CT simulation and/or treatment plan reoptimization. |
up to 6 months
|
二次結果の測定
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Treating physician-reported influence of RayTracker/RayAdaptive on adaptive replanning decisions
時間枠:up to 6 months
|
Can I define this outcome to be: Did the RayTracker / RayAdaptive intervention influence your clinical decision? [Yes / No] This will be reported by physician count. |
up to 6 months
|
|
Concordance between RayAdaptive predictions of inadequate target coverage and findings on repeat CT simulation
時間枠:up to 6 months
|
up to 6 months
|
|
|
Concordance between RayAdaptive predictions of organ-at-risk dose constraint violations and findings on repeat CT simulation
時間枠:up to 6 months
|
up to 6 months
|
|
|
Incidence of AEs and device deficiencies
時間枠:up to 6 months
|
Incidence of AEs and device deficiencies at least possibly related to use of the investigational RayTracker/RayAdaptive software or its incorporation into the clinical workflow as assessed throughout the course of proton therapy and for 90 days following completion of proton treatment. Device deficiencies may include software malfunction, calculation error, deformable registration failure, software defect, or software limitations affecting treatment plan evaluation, image review, or data analysis. |
up to 6 months
|
協力者と研究者
捜査官
捜査官
- 主任研究者:Brett Morris, MD, PhD、UW School of Medicine and Public Health
研究記録日
主要日程の研究
研究開始 (実際)
研究開始
一次修了 (推定)
一次修了
研究の完了 (推定)
研究の完了
試験登録日
最初に提出
最初に提出
QC基準を満たした最初の提出物
QC基準を満たした最初の提出物
最初の投稿 (実際)
最初の投稿
学習記録の更新
投稿された最後の更新 (実際)
投稿された最後の更新
QC基準を満たした最後の更新が送信されました
QC基準を満たした最後の更新が送信されました
最終確認日
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
その他の研究ID番号
- 2026-1312
- Protocol Version 7/26/26 (その他の識別子:UW Madison)
- SMPH | Radiation Medicine (その他の識別子:UW Madison)
- UW26111 (その他の識別子:OnCore ID)
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
IPD 共有サポート情報タイプ
- STUDY_PROTOCOL
- SAP
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
米国で製造され、米国から輸出された製品。
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。