Real-world Effectiveness and Safety of Omnipod 5 in Adults With Type 1 Diabetes: a 12-month Observational Study.
연구 개요
상태
상태
정황
정황
상세 설명
Automated insulin delivery systems have demonstrated significant improvements in glycemic outcomes in randomized clinical trials. However, evidence from routine clinical practice remains essential to evaluate effectiveness across broader patient populations and routine care settings.
This retrospective observational cohort study will include consecutive adults with type 1 diabetes who initiated treatment with the Omnipod® 5 Automated Insulin Delivery System during routine clinical practice at ASST Papa Giovanni XXIII Hospital, Bergamo, Italy.
Clinical and continuous glucose monitoring (CGM) data routinely collected during clinical follow-up will be extracted from medical records. Glycemic outcomes obtained before initiation of Omnipod 5 will be compared with those recorded after approximately six, twelve and eighteen months of system use.
The primary objective is to evaluate changes in Time in Range (70-180 mg/dL). Secondary objectives include assessment of HbA1c, Time Above Range, Time Below Range, glycemic variability, the proportion of participants achieving recommended international CGM targets; the prevalence of acute diabetes complications (severe hypoglycemia, diabetic ketoacidosis); and rates and causes of system dropout.
연구 유형
연구 유형
등록 (실제)
등록
연락처 및 위치
연구 장소
-
-
Italy
-
Bergamo, Italy, 이탈리아, 20126
- ASST Papa Giovanni XXIII Bergamo
-
-
참여기준
자격 기준
자격 기준
공부할 수 있는 나이
- 성인
- 고령자
건강한 자원 봉사자를 받아들입니다
샘플링 방법
연구 인구
설명
Inclusion Criteria:
- Adults aged ≥18 years
- Diagnosis of type 1 diabetes
- Initiation of Omnipod® 5 during routine clinical care
- Availability of baseline CGM data
- Availability of at least six-month follow-up CGM data
Exclusion Criteria
- Missing baseline CGM data
- Missing follow-up CGM data
- Discontinuation of Omnipod® 5 before outcome assessment
- Other conditions judged by investigators to preclude reliable evaluation
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
연구는 무엇을 측정합니까?
주요 결과 측정
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Change in Time in Range (70-180 mg/dL)
기간: Baseline, approximately 6 months. approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
Change from baseline in Time in Range (70-180 mg/dL), defined as the percentage of time spent with glucose values within the target glucose range (70-180 mg/dL), as measured by continuous glucose monitoring (CGM), at approximately 6, 12, and 18 months after initiation of therapy with the Omnipod 5 system.
|
Baseline, approximately 6 months. approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
2차 결과 측정
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Change in HbA1c (mmol/mol or %)
기간: Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
Change in glycated hemoglobin (HbA1c) from baseline to approximately 6, 12, and 18 months after starting therapy with the Omnipod 5 system.
|
Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
|
Change in Time Below Range Level 1 (54-69 mg/dL)
기간: Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
Change from baseline in Time Below Range Level 1 (54-69 mg/dL), defined as the percentage of time spent in the glucose range 54-69 mg/dL, as measured by continuous glucose monitoring (CGM), at approximately 6, 12, and 18 months after initiation of therapy with the Omnipod 5 system.
|
Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
|
Change in Time Below Range Level 2 (<54 mg/dL)
기간: Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
Change from baseline in Time Below Range Level 2 (<54 mg/dL), defined as the percentage of time spent with glucose values <54 mg/dL, as measured by continuous glucose monitoring (CGM), at approximately 6, 12, and 18 months after initiation of therapy with the Omnipod 5 system.
|
Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
|
Change in Time Above Range Level 1 (>180 mg/dL)
기간: Baseline, approximately 6 months, approximately 12 months, and approximately 18 months Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
Change from baseline in Time Above Range Level 1 (181-250 mg/dL), defined as the percentage of time spent in the glucose range181-250 mg/dL, as measured by continuous glucose monitoring (CGM), at approximately 6, 12, and 18 months after initiation of the therapy with the Omnipod 5 system.
|
Baseline, approximately 6 months, approximately 12 months, and approximately 18 months Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
|
Change in Time Above Range Level 2 (>250 mg/dL)
기간: Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
Change from baseline in Time Above Range Level 2 (<250 mg/dL), defined as the percentage of time spent with glucose values >250 mg/dL, as measured by continuous glucose monitoring (CGM), at approximately 6, 12, and 18 months after initiation of therapy with the Omnipod 5 system.
|
Baseline, approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
|
Change from baseline in the number of participants experiencing at least one episode of severe hypoglycemia
기간: Baseline (12 months before initiation of the Omnipod 5 system), approximately 6 months, approximately 12 months, and approximately 18 months after starting of therapy with the Omnipod 5 system.
|
Change from baseline in the number of participants experiencing at least one episode of severe hypoglycemia during each follow-up period.
Severe hypoglycemia is defined according to the American Diabetes Association as an event requiring the assistance of another person to actively administer carbohydrate, glucagon, or other resuscitative actions.
|
Baseline (12 months before initiation of the Omnipod 5 system), approximately 6 months, approximately 12 months, and approximately 18 months after starting of therapy with the Omnipod 5 system.
|
|
Change from baseline in the total number of severe hypoglycemia episodes
기간: Baseline (12 months before initiation of the Omnipod 5 system), approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
Change from baseline in the total number of severe hypoglycemic episodes occurring during each follow-up period.
Severe hypoglycemia is defined according to the American Diabetes Association as an event requiring the assistance of another person to actively administer carbohydrate, glucagon, or other resuscitative actions.
|
Baseline (12 months before initiation of the Omnipod 5 system), approximately 6 months, approximately 12 months, and approximately 18 months after starting therapy with the Omnipod 5 system.
|
|
Change from baseline in the number of participants experiencing at least one episode of diabetic ketoacidosis
기간: Baseline (12 months before initiation of the Omnipod 5 system), approximately 6 months, approximately 12 months, and approximately 18 months after starting of therapy with the Omnipod 5 system.
|
Change from baseline in the number of participants experiencing at least one episode of diabetic ketoacidosis during each follow-up period.
Diabetic ketoacidosis will be identified according to the diagnosis documented in the medical record and standard clinical diagnostic criteria.
|
Baseline (12 months before initiation of the Omnipod 5 system), approximately 6 months, approximately 12 months, and approximately 18 months after starting of therapy with the Omnipod 5 system.
|
|
Change from baseline in the total number of diabetic ketoacidosis episodes
기간: Baseline (12 months before initiation of the Omnipod 5 system), approximately 6 months, approximately 12 months, and approximately 18 months after starting of therapy with the Omnipod 5 system.
|
Change from baseline in the total number of diabetic ketoacidosis episodes occurring during each follow-up period.
Diabetic ketoacidosis will be identified according to the diagnosis documented in the medical record and standard clinical diagnostic criteria.
|
Baseline (12 months before initiation of the Omnipod 5 system), approximately 6 months, approximately 12 months, and approximately 18 months after starting of therapy with the Omnipod 5 system.
|
공동 작업자 및 조사자
수사관
수사관
- 연구 책임자: Nicolò Borella, MD, Endocrine Disease and Diabetes Unit, ASST Papa Giovanni XXIII Bergamo, Italy
간행물 및 유용한 링크
일반 간행물
- Forlenza GP, DeSalvo DJ, Aleppo G, Wilmot EG, Berget C, Huyett LM, Hadjiyianni I, Mendez JJ, Conroy LR, Ly TT, Sherr JL. Real-World Evidence of Omnipod(R) 5 Automated Insulin Delivery System Use in 69,902 People with Type 1 Diabetes. Diabetes Technol Ther. 2024 Aug;26(8):514-525. doi: 10.1089/dia.2023.0578. Epub 2024 Feb 16.
- Forlenza GP, Buckingham BA, Brown SA, Bode BW, Levy CJ, Criego AB, Wadwa RP, Cobry EC, Slover RJ, Messer LH, Berget C, McCoy S, Ekhlaspour L, Kingman RS, Voelmle MK, Boyd J, O'Malley G, Grieme A, Kivilaid K, Kleve K, Dumais B, Vienneau T, Huyett LM, Lee JB, O'Connor J, Benjamin E, Ly TT. First Outpatient Evaluation of a Tubeless Automated Insulin Delivery System with Customizable Glucose Targets in Children and Adults with Type 1 Diabetes. Diabetes Technol Ther. 2021 Jun;23(6):410-424. doi: 10.1089/dia.2020.0546. Epub 2021 Jan 18.
- Brown SA, Forlenza GP, Bode BW, Pinsker JE, Levy CJ, Criego AB, Hansen DW, Hirsch IB, Carlson AL, Bergenstal RM, Sherr JL, Mehta SN, Laffel LM, Shah VN, Bhargava A, Weinstock RS, MacLeish SA, DeSalvo DJ, Jones TC, Aleppo G, Buckingham BA, Ly TT; Omnipod 5 Research Group. Multicenter Trial of a Tubeless, On-Body Automated Insulin Delivery System With Customizable Glycemic Targets in Pediatric and Adult Participants With Type 1 Diabetes. Diabetes Care. 2021 Jul;44(7):1630-1640. doi: 10.2337/dc21-0172. Epub 2021 Jun 7.
연구 기록 날짜
연구 주요 날짜
연구 시작 (실제)
연구 시작
기본 완료 (실제)
기본 완료
연구 완료 (실제)
연구 완료
연구 등록 날짜
최초 제출
최초 제출
QC 기준을 충족하는 최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
처음 게시됨
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
마지막 업데이트 게시됨
QC 기준을 충족하는 마지막 업데이트 제출
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
기타 연구 ID 번호
기타 연구 ID 번호
- Tubeless AID systems
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
IPD 계획 설명
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .