Computed Tomography Dose Reduction Using Sequential or Fast Pitch Sprial Technique (CTDOSE)
Computed Tomography Dose Reduction Using Sequential or Fast Pitch Sprial Technique Employed in Cardiothoracic Imaging
Studie Overzicht
Toestand
Gedetailleerde beschrijving
The development of a novel CT scanner capable of sequential or fast-pitch spiral techniques in most patients (Siemens Definition Flash) represents a potentially crucial technical innovation at a particularly critical juncture. National healthcare systems are under pressure to deliver accurate and cost-effective diagnosis of an increasing number of patients and CT angiography in various settings is a proven, highly accurate technique that has been shown to improve diagnostic efficiency and reduce the cost of care.
However, there has been increasing concern about the lifetime attributable risk of cancer from radiation related to diagnostic procedures. Such concern extends to other radiation-based procedures that contribute to the cumulative lifetime radiation exposure of patients. The Flash scanner holds the promise of substantially reducing exposure resulting from a variety of examinations.
The CT DOSE multicenter trial is therefore designed to validate the extent of dose reduction attendant to the use of the Flash scanner in cardiothoracic scanning, and to determine whether image quality is preserved in spite of significant dose reduction.
Studietype
Inschrijving (Werkelijk)
Contacten en locaties
Studie Locaties
-
-
Kingdom of Saudi Arabia
-
Riyadh, Kingdom of Saudi Arabia, Saoedi-Arabië, 11426
- King Abdul-Aziz Cardiac Center, national Guard Health Affairs
-
-
-
-
California
-
San Diego, California, Verenigde Staten, 92134
- Naval Medical Center
-
-
Massachusetts
-
Boston, Massachusetts, Verenigde Staten, 02114
- Massachusetts General Hospital
-
-
Michigan
-
Royal Oak, Michigan, Verenigde Staten, 48073
- William Beaumont Hospital-Royal Oak
-
Troy, Michigan, Verenigde Staten, 48085
- William Beaumont Hospital-Troy
-
-
Minnesota
-
Minneapolis, Minnesota, Verenigde Staten, 55407
- Minneapolis Heart Institute/ Abbott Northwestern
-
-
New York
-
New York, New York, Verenigde Staten, 10006
- NYU Langone Medical Center
-
-
Ohio
-
Cleveland, Ohio, Verenigde Staten, 44195
- Cleveland Clinic Foundation
-
-
South Carolina
-
Charleston, South Carolina, Verenigde Staten, 29425
- Medical University of South Carolina
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Bemonsteringsmethode
Studie Bevolking
Beschrijving
Inclusion Criteria:
- Scans from patients undergoing clinically indicated pulmonary angiography, aortography and/or CCTA.
- Age greater than or equal to 18 years.
Exclusion Criteria:
- No exclusions
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
Cohorten en interventies
Groep / Cohort |
|---|
|
Control Group
Data from patients in the control group will be obtained retrospectively from patient medical records and stored image data
|
|
Prospective Patient Group
Data from patients in the Flash group will be obtained prospectively from scanner consoles and medical records.
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Radiation dose
Tijdsspanne: Time of CT scan
|
The primary endpoint is radiation dose, measured as dose-length-product (DLP, in mGy∙cm) and effective radiation dose, calculated as DLP x 0.014 (in mSv).
|
Time of CT scan
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Assessment of Image Quality
Tijdsspanne: Time of CT Scan
|
Signal noise (standard deviation of attenuation in Hounsfield units) will be determined in regions of interest in: A) air anterior to the chest wall, B) mid-LV septum and C) mid thoracic aorta or main PA.
This will be done in a sequential sample of every 5th case (approximately 600 cases total).
|
Time of CT Scan
|
Medewerkers en onderzoekers
Sponsor
Medewerkers
Onderzoekers
- Hoofdonderzoeker: Gilbert Raff, MD, William Beaumont Hospitals
Studie record data
Bestudeer belangrijke data
Studie start
Primaire voltooiing (Werkelijk)
Studie voltooiing (Werkelijk)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Schatting)
Updates van studierecords
Laatste update geplaatst (Schatting)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- 2010-209
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .