Validation of Fore-Sight Cerebral & Tissue Oximeter Saturation Measurements on Neonates ≤ 5 kg (FORE-SIGHT)
6 oktober 2017 bijgewerkt door: Anup Katheria, M.D., Sharp HealthCare
Validation of FORE-SIGHT Cerebral & Tissue Oximeter to Measure Cerebral & Somatic Tissue Oxygen Saturation Measurements at Different Body Locations on Neonates ≤ 5 kg
CASMED, INC. wishes to validate the next generation FORE-SIGHT-II (FS-II) NIRS tissue oximeter monitor for neonatal brain and somatic tissue oxygen saturation (StO2) measurements.
This monitor was recently FDA cleared for use on adult subjects.
Studie Overzicht
Toestand
Voltooid
Conditie
Gedetailleerde beschrijving
- Validate FS-II abdominal StO2 measurements from the intestines, liver, and flank to umbilical venous and arterial blood oxygen saturation.
- Validate FS-II cerebral StO2 measurements against the predicate FDA cleared FORE-SIGHT-I (FS-I) monitor.
- Collect data from FS-II sensors on the forehead and abdomen (intestine) for 24 hours to compare dynamic changes vs the pulse oximeter. For larger neonates, add a predicate FS-I sensor to the opposite side forehead for dynamic comparison to FS-II measurements.
- Obtain abdominal Near Infrared Spectroscopy (NIRS) data simultaneous with accepted measures of bowel motility (a. frequency and composition of stools, and b. Ultrasound assessment for the frequency and magnitude of bowel peristalsis, if available).
Studietype
Observationeel
Inschrijving (Werkelijk)
19
Contacten en locaties
In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.
Studie Locaties
-
-
California
-
San Diego, California, Verenigde Staten, 92123
- Sharp Mary Birch Hospital for Women and Newborns
-
-
Deelname Criteria
Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Niet ouder dan 1 week (Kind)
Accepteert gezonde vrijwilligers
Nee
Geslachten die in aanmerking komen voor studie
Allemaal
Bemonsteringsmethode
Niet-waarschijnlijkheidssteekproef
Studie Bevolking
Neonatal subjects (32 to 36 and 6/7 weeks), with umbilical venous catheterization will be enrolled from the NICU.
Demographic information of the subject will be recorded, including race, gender, weight, gestational age (GA), study age, and diagnosis.
Beschrijving
Inclusion Criteria:
Neonatal subjects (32 to 36 and 6/7 weeks), with umbilical venous catheterization will be enrolled from the Neonatal Intensive Care Unit.
Exclusion Criteria:
Enrollment in other studies during study period. Inability to tolerate ultrasounds or NIRS sensors
Studie plan
Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.
Hoe is de studie opgezet?
Ontwerpdetails
- Observatiemodellen: Cohort
- Tijdsperspectieven: Prospectief
Cohorten en interventies
Groep / Cohort |
|---|
|
Preterm Infants
Preterm infants 32 to 36 weeks will be eligible for this study
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
StO2 measurement from intestines, liver and flank at time of venous and arterial blood oxygen saturation.
Tijdsspanne: 1 hour
|
At or immediately following an umbilical venous blood draw for CO-oximetry, the FS-II NIRS Small sensor will be briefly placed on the abdominal wall for spot check measurements over the 1) left & 2) right flank, 3) liver and on three intestine locations: just inferior and lateral to the umbilicus in the 4) Right Lower Quadrant (RLQ) and 5) Left Lower Quadrant (LLQ) positions, as well as 6) Mid Lower Quadrant (MLQ).
Then a reference FS-I measurement of the 7) flank for 2 minutes will be done.
|
1 hour
|
|
Cerebral StO2 measurements from Fore-Sight FS-II monitor and Fore-Sight-I (FS-I) monitor.
Tijdsspanne: 1 hour
|
Following abdominal measurements, the FS-II Small sensor will be placed on the head with the following positions in the following sequence 1) forehead, over the frontal cortex for two minutes; 2) the FS-II sensor is lifted off the head and placed back in the same forehead position to make a repeat measurement for two minutes; and 3) on the side of the head above the ear (frontal and parietal cortex) for two minutes.
For reference cerebral measurements, a FDA cleared FS-I Small sensor will be placed on the 4) forehead for two minutes; 5) the FS-I sensor will be removed and replaced on the same forehead position to make a repeat measurement for two minutes.
The sensors will take a measurement of 2 minutes in each position and 10 minutes total.
|
1 hour
|
|
Pulse and Tissue oxygen saturation and infant activity
Tijdsspanne: 24 hours
|
Then two FS-II Small sensors will be attached longer term, one sensor on the forehead and one sensor on the abdomen (RLQ), with neonatal adhesives to allow monitoring for up to 24 hours.
Dynamic changes of StO2 at these locations will be observed, along with pulse oximetry SpO2 to capture any acute events such as apnea and bradycardia, which will help in modeling and verifying dynamic StO2 change behavior within the FS-II algorithm.
FS-II, and pulse oximetry data will be collected simultaneously on a laptop computer.
|
24 hours
|
|
Stool Interference Index over 24 hrs.
Tijdsspanne: 24 hours
|
To calculate abdominal StO2 over the intestine regions, FORE-SIGHT also measures a stool interference index, which is used to compensate the algorithm to calculate StO2 correctly.
This stool interference index is hypothesized to be also useful as a bowel motility index based on its variability over time.
To test this hypothesis, the following two references for bowel motility will be tested: a) the study site will minimally record the number of stools for a period up to 24 hours.
The event field in the laptop computer will be used to note each observation.
b) if the study site has the capacity, abdominal ultrasound will be used to observe for the frequency and magnitude of bowel peristalsis.
These impressions will be added to an event field in the laptop computer.
|
24 hours
|
Medewerkers en onderzoekers
Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.
Sponsor
Onderzoekers
- Hoofdonderzoeker: Anuup Katheria, MD, Sharp Mary Birch Hospital for Women & Newborns
Studie record data
Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.
Bestudeer belangrijke data
Studie start
1 december 2013
Primaire voltooiing (Werkelijk)
1 december 2014
Studie voltooiing (Werkelijk)
1 december 2014
Studieregistratiedata
Eerst ingediend
11 december 2013
Eerst ingediend dat voldeed aan de QC-criteria
16 december 2013
Eerst geplaatst (Schatting)
23 december 2013
Updates van studierecords
Laatste update geplaatst (Werkelijk)
9 oktober 2017
Laatste update ingediend die voldeed aan QC-criteria
6 oktober 2017
Laatst geverifieerd
1 oktober 2017
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Andere studie-ID-nummers
- FORE-SIGHT
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .
Klinische onderzoeken op Bowel Motility
-
Central Hospital, Nancy, FranceBeaujon Hospital; Société Francophone Nutrition Clinique et MétabolismeOnbekendSBS - Short Bowel SyndroomFrankrijk
-
GlyPharma TherapeuticsVectivBio AGVoltooidSBS - Short Bowel SyndroomDenemarken