Denne siden ble automatisk oversatt og nøyaktigheten av oversettelsen er ikke garantert. Vennligst referer til engelsk versjon for en kildetekst.

Validation of Fore-Sight Cerebral & Tissue Oximeter Saturation Measurements on Neonates ≤ 5 kg (FORE-SIGHT)

6. oktober 2017 oppdatert av: Anup Katheria, M.D., Sharp HealthCare

Validation of FORE-SIGHT Cerebral & Tissue Oximeter to Measure Cerebral & Somatic Tissue Oxygen Saturation Measurements at Different Body Locations on Neonates ≤ 5 kg

CASMED, INC. wishes to validate the next generation FORE-SIGHT-II (FS-II) NIRS tissue oximeter monitor for neonatal brain and somatic tissue oxygen saturation (StO2) measurements. This monitor was recently FDA cleared for use on adult subjects.

Studieoversikt

Status

Fullført

Forhold

Detaljert beskrivelse

  1. Validate FS-II abdominal StO2 measurements from the intestines, liver, and flank to umbilical venous and arterial blood oxygen saturation.
  2. Validate FS-II cerebral StO2 measurements against the predicate FDA cleared FORE-SIGHT-I (FS-I) monitor.
  3. Collect data from FS-II sensors on the forehead and abdomen (intestine) for 24 hours to compare dynamic changes vs the pulse oximeter. For larger neonates, add a predicate FS-I sensor to the opposite side forehead for dynamic comparison to FS-II measurements.
  4. Obtain abdominal Near Infrared Spectroscopy (NIRS) data simultaneous with accepted measures of bowel motility (a. frequency and composition of stools, and b. Ultrasound assessment for the frequency and magnitude of bowel peristalsis, if available).

Studietype

Observasjonsmessig

Registrering (Faktiske)

19

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiesteder

    • California
      • San Diego, California, Forente stater, 92123
        • Sharp Mary Birch Hospital for Women and Newborns

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

Ikke eldre enn 1 uke (Barn)

Tar imot friske frivillige

Nei

Kjønn som er kvalifisert for studier

Alle

Prøvetakingsmetode

Ikke-sannsynlighetsprøve

Studiepopulasjon

Neonatal subjects (32 to 36 and 6/7 weeks), with umbilical venous catheterization will be enrolled from the NICU. Demographic information of the subject will be recorded, including race, gender, weight, gestational age (GA), study age, and diagnosis.

Beskrivelse

Inclusion Criteria:

Neonatal subjects (32 to 36 and 6/7 weeks), with umbilical venous catheterization will be enrolled from the Neonatal Intensive Care Unit.

Exclusion Criteria:

Enrollment in other studies during study period. Inability to tolerate ultrasounds or NIRS sensors

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

  • Observasjonsmodeller: Kohort
  • Tidsperspektiver: Potensielle

Kohorter og intervensjoner

Gruppe / Kohort
Preterm Infants
Preterm infants 32 to 36 weeks will be eligible for this study

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
StO2 measurement from intestines, liver and flank at time of venous and arterial blood oxygen saturation.
Tidsramme: 1 hour
At or immediately following an umbilical venous blood draw for CO-oximetry, the FS-II NIRS Small sensor will be briefly placed on the abdominal wall for spot check measurements over the 1) left & 2) right flank, 3) liver and on three intestine locations: just inferior and lateral to the umbilicus in the 4) Right Lower Quadrant (RLQ) and 5) Left Lower Quadrant (LLQ) positions, as well as 6) Mid Lower Quadrant (MLQ). Then a reference FS-I measurement of the 7) flank for 2 minutes will be done.
1 hour
Cerebral StO2 measurements from Fore-Sight FS-II monitor and Fore-Sight-I (FS-I) monitor.
Tidsramme: 1 hour
Following abdominal measurements, the FS-II Small sensor will be placed on the head with the following positions in the following sequence 1) forehead, over the frontal cortex for two minutes; 2) the FS-II sensor is lifted off the head and placed back in the same forehead position to make a repeat measurement for two minutes; and 3) on the side of the head above the ear (frontal and parietal cortex) for two minutes. For reference cerebral measurements, a FDA cleared FS-I Small sensor will be placed on the 4) forehead for two minutes; 5) the FS-I sensor will be removed and replaced on the same forehead position to make a repeat measurement for two minutes. The sensors will take a measurement of 2 minutes in each position and 10 minutes total.
1 hour
Pulse and Tissue oxygen saturation and infant activity
Tidsramme: 24 hours
Then two FS-II Small sensors will be attached longer term, one sensor on the forehead and one sensor on the abdomen (RLQ), with neonatal adhesives to allow monitoring for up to 24 hours. Dynamic changes of StO2 at these locations will be observed, along with pulse oximetry SpO2 to capture any acute events such as apnea and bradycardia, which will help in modeling and verifying dynamic StO2 change behavior within the FS-II algorithm. FS-II, and pulse oximetry data will be collected simultaneously on a laptop computer.
24 hours
Stool Interference Index over 24 hrs.
Tidsramme: 24 hours
To calculate abdominal StO2 over the intestine regions, FORE-SIGHT also measures a stool interference index, which is used to compensate the algorithm to calculate StO2 correctly. This stool interference index is hypothesized to be also useful as a bowel motility index based on its variability over time. To test this hypothesis, the following two references for bowel motility will be tested: a) the study site will minimally record the number of stools for a period up to 24 hours. The event field in the laptop computer will be used to note each observation. b) if the study site has the capacity, abdominal ultrasound will be used to observe for the frequency and magnitude of bowel peristalsis. These impressions will be added to an event field in the laptop computer.
24 hours

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Etterforskere

  • Hovedetterforsker: Anuup Katheria, MD, Sharp Mary Birch Hospital for Women & Newborns

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart

1. desember 2013

Primær fullføring (Faktiske)

1. desember 2014

Studiet fullført (Faktiske)

1. desember 2014

Datoer for studieregistrering

Først innsendt

11. desember 2013

Først innsendt som oppfylte QC-kriteriene

16. desember 2013

Først lagt ut (Anslag)

23. desember 2013

Oppdateringer av studieposter

Sist oppdatering lagt ut (Faktiske)

9. oktober 2017

Siste oppdatering sendt inn som oppfylte QC-kriteriene

6. oktober 2017

Sist bekreftet

1. oktober 2017

Mer informasjon

Begreper knyttet til denne studien

Andre studie-ID-numre

  • FORE-SIGHT

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .

Kliniske studier på Bowel Motility

3
Abonnere