Optical verification and in-vitro characterization of two commercially available acoustic bubble counters for cardiopulmonary bypass systems
Tim Segers, Marco C Stehouwer, Filip M J J de Somer, Bastian A de Mol, Michel Versluis, Tim Segers, Marco C Stehouwer, Filip M J J de Somer, Bastian A de Mol, Michel Versluis
Abstract
Introduction: Gaseous microemboli (GME) introduced during cardiac surgery are considered as a potential source of morbidity, which has driven the development of the first bubble counters. Two new generation bubble counters, introduced in the early 2000s, claim correct sizing and counting of GME. This in-vitro study aims to validate the accuracy of two bubble counters using monodisperse bubbles in a highly controlled setting at low GME concentrations.
Methods: Monodisperse GME with a radius of 43 µm were produced in a microfluidic chip. Directly after their formation, they were injected one-by-one into the BCC200 and the EDAC sensors. GME size and count, measured with the bubble counters, were optically verified using high-speed imaging.
Results: During best-case scenarios or low GME concentrations of GME with a size of 43 µm in radius in an in-vitro setup, the BCC200 overestimates GME size by a factor of 2 to 3 while the EDAC underestimates the average GME size by at least a factor of two. The BCC200 overestimates the GME concentration by approximately 20% while the EDAC overestimates the concentration by nearly one order of magnitude. Nevertheless, the calculated total GME volume is only over-predicted by a factor 2 since the EDAC underestimates the actual GME size. For the BCC200, the total GME volume was over-predicted by 25 times due to the over-estimation of GME size.
Conclusions: The measured errors in the absolute sizing/counting of GME do not imply that all results obtained using the bubble counters are insignificant or invalid. A relative change in bubble size or bubble concentration can accurately be measured. However, care must be taken in the interpretation of the results and their absolute values. Moreover, the devices cannot be used interchangeably when reporting GME activity. Nevertheless, both devices can be used to study the relative air removal characteristics of CPB components or for the quantitative monitoring of GME production during CPB interventions.
Keywords: GME; bubble counter; cardiopulmonary bypass; emboli; microbubble; validation.
Conflict of interest statement
Declaration of Conflicting Interests: The authors declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Figures
References
- Barak M, Katz Y. Microbubbles: pathophysiology and clinical implications. Chest 2005; 128: 2918–2932.
- Rodriguez RA, Williams KA, Babaev A, Rubens F, Nathan HJ. Effect of perfusionist technique on cerebral embolization during cardiopulmonary bypass. Perfusion 2005; 20: 3–10.
- Barbut D, Lo Y, Gold J, et al. Impact of embolization during coronary artery bypass grafting on outcome and length of stay. Ann Thorac Surgery 1997; 63: 998–1002.
- Clark RE, Brillman J, Davis DA, Lovell MR, Price TRP, Macgovern GL. Microemboli during coronary artery bypass grafting. J Thorac Cardiovasc Surg 1995; 109: 249–258.
- Raymond PD, Marsh NA. Alteration to haemostasis following cardiopulmonary bypass and the relationship of these changes to neurocognitive morbidity. Blood Coag Fibrinolysis 2001; 12: 601–618.
- Nishi RY. Ultrasonic detection of bubbles with Doppler flow transducers. Ultrasonics 1972; 10: 173–179.
- Medwin H. Counting bubbles acoustically: a review. Ultrasonics 1977; 15: 7–13.
- De Somer F. Impact of oxygenator characteristics on its capability to remove gaseous microemboli. J Extra Corpor Technol 2007; 39: 271–273.
- Stehouwer MC, Boers C, de Vroege R, Kelder C, Yilmaz A, Bruins P. Clinical evaluation of the air removal characteristics of an oxygenator with integrated arterial filter in a minimized extracorporeal circuit. J Artif Organs 2011; 34: 374–382.
- Liu S, Newland RF, Tully PJ, Tuble SC, Baker RA. In vitro evaluation of gaseous microemboli handling of cardiopulmonary bypass circuits with and without integrated arterial line filters. J Extra Corpor Technol 2011; 43: 107–114.
- Potger KC, McMillan D, Ambrose M. Microbubble transmission during cardiotomy infusion of a hardshell venous reservoir with integrated cardiotomy versus a softshell venous reservoir with separated cardiotomy: an in vitro comparison. J Extra Corpor Technol 2013; 45, 77–85.
- Johagen D, Appelblad M, Svenmarker S. Can the oxygenator screen filter reduce gaseous microemboli. J Extra Corpor Technol 2012; 46: 60–66.
- Melchior RW, Rosenthal T, Glatz AC. An in vitro comparison of the ability of three commonly used pediatric cardiopulmonary bypass circuits to filter gaseous microemboli. Perfusion 2010; 25: 255–263.
- Lynch JE, Pouch A, Sanders R, Hinders M, Rudd K, Sevick J. Gaseous microemboli sizing in extracorporeal circuits using ultrasound backscatter. Ultrasound Med Biol 2007; 33: 1661–1675.
- Palanchon P, Klein J, de Jong N. Production of standardized air bubbles: application to embolism studies. Rev Sci Instrum 2003; 74: 2558 –2563.
- Banahan C, Hague JP, Evans DH, Patel R, Ramnarine KV, Chung EML. Sizing gaseous emboli using Doppler embolic signal intensity. Ultraound Med Biol 2012; 38: 824–833.
- Lynch JB, Riley JE. Microemboli detection on extracorporeal bypass circuits. Perfusion 2008; 23: 23–32.
- Jenderka KV, Schultz M, Dietrich G, Fornara P. Detection and sizing of micro bubbles in streaming fluids with ultrasound. Sensors Proceedings of IEEE 2002; 1: 528–531.
- Brand S, Klaua R, Dietrich G, Schultz M. Ultrasonic detection and quantitative analysis of microscopic bubbles and particles in solution: enhanced by attenuation compensation. Proceedings of the IEEE Ultrasonics Symposium 2006; 1: 1840–1843.
- De Somer FM, Vetrano MR, Van Beeck JP, van Nooten GJ. Extracorporeal bubbles: a word of caution. Interact Cardiovasc Thorac Surg 2010; 10: 995–1002.
- Schultz M, Klaua R, Oblonczek G. eComment: Ultrasonic bubble detection: some additional information. Interact Cardiovasc Thorac Surg 2010; 10: 1001–1002.
- Maresca D, Emmer M, van Meer PLMJ, et al. Acoustic sizing of an ultrasound contrast agent. Ultrasound Med Biol 2010; 36: 1713–1721.
- Urbanek S, Tiedtke HJ. Improved methods for measurement of gaseous microbubbles during extracorporeal circulation. Perfusion 2012; 17: 429–434.
- Tan YC, Cristini V, Lee AP. Monodispersed microfluidic droplet generation by shear focusing microfluidic device. Sensors and Actuators B: Chemical 2005; 114: 350–356.
- Duffy DC, McDonald JC, Schueller OJA, Whitesides GM. Rapid prototyping of microfluidic systems in poly(dimethylsiloxane). Anal Chem 1998; 70: 211–217.
- Chung EM, Banahan C, Patel N, et al. Size distribution of air bubbles entering the brain during cardiac surgery. PLoS One 2015; 10: e0122166.
- Segers T, Versluis M. Acoustic bubble sorting for ultrasound contrast agent enrichment. Lab Chip 2014; 14: 1705–1714.
- Szabo T. Diagnostic Ultrasound Imaging: Inside Out. London: Academic Press, 2004.
- Culjat MO, Goldenberg D, Tewari P, Singh R. A review of tissue substitutes for ultrasound imaging. Ultrasound Med Biol 2010; 36: 861–873.
Source: PubMed