VALIDITY OF HAND-HELD DYNAMOMETRY IN MEASURING QUADRICEPS STRENGTH AND RATE OF TORQUE DEVELOPMENT

Joseph Lesnak, Dillon Anderson, Brooke Farmer, Dimitrios Katsavelis, Terry L Grindstaff, Joseph Lesnak, Dillon Anderson, Brooke Farmer, Dimitrios Katsavelis, Terry L Grindstaff

Abstract

Background: A hand-held dynamometer (HHD) offers a reliable and valid method to quantify quadriceps strength in a clinical environment. While measures of peak strength provide functional insights, most daily activities are performed quickly and do not require maximum strength. Rate of torque development (RTD) measures better reflect both the demands of daily activity and athletic movements. The capacity to obtain RTD measures in clinical settings is possible with an HHD, but the validity of RTD measures has not been quantified.

Hypothesis/purpose: To determine the validity of an HHD to measure quadriceps isometric strength metrics compared to isometric strength measures obtained on an isokinetic dynamometer. It was hypothesized that the HHD would be a valid measure of peak torque and RTD at all time intervals when compared to the isokinetic dynamometer.

Study design: Descriptive laboratory study.

Methods: Twenty healthy participants (12 male, 8 female) (age = 23.7 ± 2.9 years, height = 174.6 ± 10.1 cm, mass = 76.4 ± 15.9 kg, and Tegner = 6.7 ± 1.2) performed maximum isometric quadriceps contractions on an isokinetic dynamometer and with an HHD. Outcome measures included quadriceps peak torque and RTD at three intervals (0-100, 0-250 ms, and average). Pearson product-moment correlation coefficients and Spearman's rank correlation coefficient were used to determine relationships between devices. Bland-Altman Plots with Limits of Agreement (LOA) calculations were used to quantify systematic bias between measurement techniques.

Results: There was a significant correlation between the isokinetic dynamometer and the HHD for peak torque (p<.001, r = .894) and all RTD measurements (p<.002, r = .807; ρ = .502-.604). Bland-Altman plot LOA indicated the HHD overestimated peak torque values (19.4 ± 53.2 Nm) and underestimated all RTD measurements (-55.2 ± 190.7 Nm/s to -265.2 ± 402.6 Nm/s).

Conclusion: These results show it is possible to obtain valid measures of quadriceps peak torque and late RTD using an HHD. Measures of early RTD and RTDAvg obtained with an HHD were more variable and should be viewed with caution.

Level of evidence: Diagnostic, Level 3.

Keywords: dynamometry; explosive strength; knee extension; lower limb.

Figures

Figure 1.
Figure 1.
Participants performed testing on an isokinetic dynamometer with hips flexed to 85 ° and knees flexed to 90 ° The chest and pelvis were secured with straps, and a padded ankle strap was placed 5 cm proximal to the distal aspect of the lateral malleolus. Participants kept arms crossed during testing to isolate the quadriceps muscle.
Figure 2.
Figure 2.
HHD testing was performed in a seated position with legs hanging from the edge of a wooden treatment table, with the knee in approximately 90 ° of flexion (A). A small wedge bolster was placed at the posterior aspect of the distal thigh to minimize posterior thigh discomfort, and a strap (standard gait belt) was used to stabilize the thighs to the table. Participants kept arms crossed during testing to isolate the quadriceps muscle. A foam pad was placed across the anterior shin, in a location consistent with the isokinetic dynamometer (5 cm proximal to lateral malleolus), with the strap looped through the pad and the HHD secured against the leg of the table (B). A small elastic wrap was positioned between the table leg and triceps surae muscle to minimize belt slack (i.e., compliance).
Figure 3.
Figure 3.
Bland-Altman Plots and Limits of Agreement for Isokinetic Dynamometer versus Handheld Dynamometer. Black line indicates average difference, red and yellow lines indicate upper and lower Limits of Agreement.

Source: PubMed