Survival of 11,390 Continuum cups in primary total hip arthroplasty based on data from the Finnish Arthroplasty Register

Matias Hemmilä, Mikko Karvonen, Inari Laaksonen, Markus Matilainen, Antti Eskelinen, Jaason Haapakoski, Ari-Pekka Puhto, Jukka Kettunen, Mikko Manninen, Keijo T Mäkelä, Matias Hemmilä, Mikko Karvonen, Inari Laaksonen, Markus Matilainen, Antti Eskelinen, Jaason Haapakoski, Ari-Pekka Puhto, Jukka Kettunen, Mikko Manninen, Keijo T Mäkelä

Abstract

Background and purpose - The use of trabecular metal (TM) cups for primary total hip arthroplasty (THA) is increasing. Some recent data suggest that the use of TM in primary THA might be associated with an increased risk of revision. We compared implant survival of Continuum acetabular cups with other commonly used uncemented cups. Patients and methods - Data on 11,390 primary THAs with the Continuum cup and 30,372 THAs with other uncemented cups (reference group) were collected from the Finnish Arthroplasty Register. Kaplan-Meier survival estimates were calculated; the endpoint was revision for any reason, for infection, or for dislocation. Revision risks were assessed with adjusted Cox multiple regression models. A subgroup analysis on the use of neutral or elevated liners in the Continuum group was made. Results - The 7-year survivorship of the Continuum group was 94.6% (95% CI 94.0-95.2) versus 95.6% (CI 95.3-95.8) in the reference group for revision for any reason. The risk for revision was higher in the Continuum group than in the reference group both for revision for any reason (HR 1.3 [CI 1.2-1.5)]) and for revision for dislocation (HR 1.9 [CI 1.5-2.3]). There was no difference in the rates of revision because of infection (HR 0.99 [CI 0.78-1.3]). Use of a neutral liner increased the risk for revision due to dislocation in comparison with the use of an elevated rim liner in the Continuum group (HR 1.7 [CI 1.2-2.5]). Interpretation - THA with Continuum cups is associated with an increased risk of revision compared with other uncemented cups, mainly due to revisions because of dislocation. Our results support the use of an elevated liner when Continuum cups are used for primary THA.

Figures

Figure 1.
Figure 1.
Kaplan–Meier survival for Continuum group and reference group with revision for any reason as the endpoint. 95% CI levels presented in blue and red.
Figure 2.
Figure 2.
Kaplan–Meier survival for Continuum group and reference group with revision for infection as endpoint. 95% CI levels presented in blue and red.
Figure 3.
Figure 3.
Kaplan–Meier survival for Continuum group and reference group with revision for dislocation as endpoint. 95% CI levels presented in blue and red.
Figure 4.
Figure 4.
Kaplan–Meier survival by subgroup analysis of Continuum THA with or without elevated liner. Endpoint: revision for dislocations. 95% CI levels presented in blue and red.

References

    1. Baad-Hansen T, Kold S, Nielsen P T, Laursen M B, Christensen P H, Soballe K. Comparison of trabecular metal cups and titanium fiber-mesh cups in primary hip arthroplasty: a randomized RSA and bone mineral densitometry study of 50 hips. Acta Orthop 2011; 82(2): 155-60.
    1. Carter A H, Sheehan E C, Mortazavi S M J, Purtill J J, Sharkey P F, Parvizi J. Revision for recurrent instability: what are the predictors of failure? J Arthroplasty 2011; 26(6): 46–52.
    1. Charnley J. Low friction arthroplasty of the hip: theory and practice. New York: Springer Verlag; 1979.
    1. Cobb T K, Morrey B F, Ilstrup D M. The elevated-rim acetabular liner in total hip arthroplasty: relationship to postoperative dislocation. J Bone Joint Surg Am 1996; 78(1): 80–6.
    1. Cobb T K, Morrey B F, Ilstrup D M. Effect of the elevated-rim acetabular liner on loosening after total hip arthroplasty. J Bone Joint Surg Am 1997; 79(9): 1361–4.
    1. Dale H, Fenstad A M, Hallan G, Havelin L I, Furnes O, Overgaard S, Pedersen A B, Kärrholm J, Garellick G, Pulkkinen P, Eskelinen A, Mäkelä K, Engesaeter L B. Increasing risk of prosthetic joint infection after total hip arthroplasty. Acta Orthop 2012; 83(5): 449–58.
    1. Davies J H, Laflamme GY, Delisle J, Fernandes J. Trabecular metal used for major bone loss in acetabular hip revision. J Arthroplasty 2011; 26(8): 1245–50.
    1. De Martino I, De Santis V, Sculco P K, D’Apolito R, Poultsides L A, Gasparini G. Long-term clinical and radiographic outcomes of porous tantalum monoblock acetabular component in primary hip arthroplasty: a minimum of 15-year follow-up. J Arthroplasty 2016; 31(9): 110–14.
    1. FAR. Finnish Arthroplasty Register. ; 2018.
    1. Howard J L, Kremers H M, Loechler Y A, Schleck C D, Harmsen W S, Berry D J, Cabanela M E, Hanssen A D, Pagnano M W, Trousdale R T, Lewallen D G. Comparative survival of uncemented acetabular components following primary total hip arthroplasty. J Bone Joint Surg Am 2011; 93(17): 1597–604.
    1. Huotari K, Peltola M, Jämsen E. The incidence of late prosthetic joint infections. Acta Orthop 2015; 86(3): 21-5.
    1. Insull P J, Cobbett H, Frampton C M, Munro J T. The use of a lipped acetabular liner decreases the rate of revision for instability after total hip replacement: a study using data from the New Zealand joint registry. Bone Joint J 2014; 96-B(7): 884-8
    1. Krushell R J, Burke D W, Harris W H. Elevated-rim acetabular components: effect on range of motion and stability in total hip arthroplasty. J Arthroplasty 1991; 6 Suppl: S53–S8.
    1. Kurtz S M, Lau E, Watson H, Schmier J K, Parvizi J. Economic burden of periprosthetic joint infection in the United States. J Arthroplasty 2012; 27(8 Suppl): 61–5.e1.
    1. Laaksonen I, Lorimer M, Gromov K, Rolfson O, Mäkelä K T, Graves S E, Malchau H, Mohaddes M. Does the risk of rerevision vary between porous tantalum cups and other cementless designs after revision hip arthroplasty? Clin Orthop Relat Res 2017; 475(12): 3015–22.
    1. Laaksonen I, Lorimer M, Gromov K, Eskelinen A, Rolfson O, Graves S E, Malchau H, Mohaddes M. Trabecular metal acetabular components in primary total hip arthroplasty: higher risk for revision compared with other uncemented cup designs in a collaborative register study including 93,709 hips. Acta Orthop 2018; 89(3): 259-64.
    1. Levine B, Della Valle C J, Jacobs J J. Applications of porous tantalum in total hip arthroplasty. J Am Acad Orthop Surg 2006; 14(12): 646–55.
    1. Macheras G A, Kateros K, Koutsostathis S D, Tsakotos G, Galanakos S, Papadakis S A. The trabecular metal monoblock acetabular component in patients with high congenital hip dislocation: a prospective study. J Bone Joint Surg Br 2010; 92(5): 624–8.
    1. McCollum D E, Gray W J. Dislocation after total hip arthroplasty: causes and prevention. Clin Orthop Relat Res 1990; (261): 159–70.
    1. Meneghini R M, Ford K S, McCollough C H, Hanssen A D, Lewallen D G. Bone remodeling around porous metal cementless acetabular components. J Arthroplasty 2010; 25(5): 741–7.
    1. Mohaddes M, Rolfson O, Kärrholm J. Short-term survival of the trabecular metal cup is similar to that of standard cups used in acetabular revision surgery: analysis of 2,460 first-time cup revisions in the Swedish Hip Arthroplasty Register. Acta Orthop 2015; 86(1): 26–31.
    1. Paavolainen P, Hämäläinen M, Mustonen H, Slätis P. Registration of arthroplasties in Finland. Acta Orthop 1991; 241: 27–30.
    1. Pakarinen O, Neuvonen P, Eskelinen A. Luksaatioiden ilmaantuvuus ja riskitekijät lonkan ensitekonivelleikkauksissa— 1381 leikkauksen aineisto Tekonivelsairaala Coxasta [article in Finnish, abstract in English]. Suom Ortop ja Traumatol 2018; 41(2): 142-8.
    1. Sultan P G, Tan V, Lai M, Garino J P. Independent contribution of elevated-rim acetabular liner and femoral head size to the stability of total hip implants. J Arthroplasty 2002; 17(3): 289–92.
    1. Tokarski A T, Novack T A, Parvizi J. Is tantalum protective against infection in revision total hip arthroplasty? Bone Joint J 2015; 97-B(1): 45–9.
    1. Weeden S H, Schmidt R H. The use of tantalum porous metal implants for Paprosky 3A and 3B defects. J Arthroplasty 2007; 22(6 Suppl): 151–5.
    1. Wegrzyn J, Kaufman K R, Hanssen A D, Lewallen D G. Performance of porous tantalum vs. titanium cup in total hip arthroplasty: randomized trial with minimum 10-year follow-up. J Arthroplasty 2015; 30(6): 1008–13.

Source: PubMed