Algorithm for automatic forced spirometry quality assessment: technological developments

Umberto Melia, Felip Burgos, Montserrat Vallverdú, Filip Velickovski, Magí Lluch-Ariet, Josep Roca, Pere Caminal, Umberto Melia, Felip Burgos, Montserrat Vallverdú, Filip Velickovski, Magí Lluch-Ariet, Josep Roca, Pere Caminal

Abstract

We hypothesized that the implementation of automatic real-time assessment of quality of forced spirometry (FS) may significantly enhance the potential for extensive deployment of a FS program in the community. Recent studies have demonstrated that the application of quality criteria defined by the ATS/ERS (American Thoracic Society/European Respiratory Society) in commercially available equipment with automatic quality assessment can be markedly improved. To this end, an algorithm for assessing quality of FS automatically was reported. The current research describes the mathematical developments of the algorithm. An innovative analysis of the shape of the spirometric curve, adding 23 new metrics to the traditional 4 recommended by ATS/ERS, was done. The algorithm was created through a two-step iterative process including: (1) an initial version using the standard FS curves recommended by the ATS; and, (2) a refined version using curves from patients. In each of these steps the results were assessed against one expert's opinion. Finally, an independent set of FS curves from 291 patients was used for validation purposes. The novel mathematical approach to characterize the FS curves led to appropriate FS classification with high specificity (95%) and sensitivity (96%). The results constitute the basis for a successful transfer of FS testing to non-specialized professionals in the community.

Conflict of interest statement

Competing Interests: The authors have declared that no competing interests exist.

Figures

Figure 1. Spirometer Zones.
Figure 1. Spirometer Zones.
An example of FV curve with five zones described.
Figure 2. Spirometer metrics.
Figure 2. Spirometer metrics.
Metrics involved in the traditional criteria: (A) FVC and FEV1, (B) PEFT.
Figure 3. Zone Z1 analysis.
Figure 3. Zone Z1 analysis.
An example of a FV curve that presents irregularity on the ascent to the PEF.
Figure 4. Zone Z2 analysis.
Figure 4. Zone Z2 analysis.
Examples of FV curves that present (A) bimodal peak; (B) flat peak and (C) slow peak.
Figure 5. Zone Z3 analysis.
Figure 5. Zone Z3 analysis.
Examples of FV curves that present irregularity in the descent from PEF.
Figure 6. Zone Z4 analysis.
Figure 6. Zone Z4 analysis.
Example of FT curve that present irregularity in the final part.
Figure 7. Zone Z5 analysis.
Figure 7. Zone Z5 analysis.
Examples of FV curve with peak and valley.

References

    1. Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, et al. (2005) Standardisation of spirometry. Eur Respir J. 26:319–338.
    1. Gardner RM, Hankinson JL (1988) Standardization of spirometry ATS update American Thoracic Society. J Occup Med March. 30 (3):272–3.
    1. Medical Section of the American Lung Association, Standardization of spirometry: 1994 update. (1994) Am Rev Respir Dis. 152:1107–1136.
    1. Burgos F, Roca J (2013) Is Plug-and-Play Spirometry the Future? Buyers' Guide To Respiratory Care Products. 14:11–22.
    1. Burgos F, Disdier C, de Santamaria EL, Galdiz B, Roger N, et al. (2012) Telemedicine enhances quality of forced spirometry in primary care. Eur Respir J. 39 (6):1313–8.
    1. Müller-Brandes U, Krämer M, Gappa G, Seitner-Sorge A, Hüls A. (2014) LUNOKID: can numeric ATS/ERS quality criteria replace visual inspection of spirometry? Eur Respir J. 43:1347–1356.
    1. Burgos F (2013) Impact of Information and Communication Technologies on Remote Testing: Forced Spirometry as a User Case. PhD Thesis, University of Barcelona 2013. .
    1. Sankar AB, Selvi JAV, Kumar D, Lakshmi KS (2013) Effective enhancement of classification of respiratory states using feed forward back propagation neural networks. Sadhana 38(3):377–395.
    1. de Oliveira Camargo-Brunetto MA, Gonçalves AR (2013) Diagnosing Chronic Obstructive Pulmonary Disease with Artificial Neural Networks using Health Expert Guidelines. In HEALTHINF 1:207–214.
    1. Devi GM, Venkatesan P (2013) Least Squares Support Vector Regression for Spirometric Forced Expiratory Volume (FEV1) Values. International Journal of Science and Technology 3:1.
    1. Polak AG (1998) A forward model for maximum expiration. Comp. Biol. Med. 28:613–625.
    1. Pardaens J, van de Woestijne KP, Clément J (1975) Simulation of regional lung emptying during slow and forced expirations. J. Appl. Physiol. 39:191–198.
    1. Lambert RK (1989) A new computational model for expiratory flow from nonhomogeneous human lungs. ASME Trans. J. Biomech. Eng. 111:200–205.
    1. Quanjer PH, Stanojevic S, Cole TJ, Baur X, Hall GL, et al. (2012) Multi-ethnic reference values for spirometry for the 3–95-yr age range: the global lung function 2012 equations. European Respiratory Journal 40(6):1324–1343.
    1. Enright PL (2003) How to make sure your spirometry tests are of good quality. Respir Care. 48(8):773–776.
    1. Enright PL, Skloot GS, Cox-Ganser JM, Udasin IG, Herbert R (2010) Quality of spirometry performed by 13,599 participants in the World Trade Center Worker and Volunteer Medical Screening Program. Respir Care 55(3):303–309.
    1. Masa JF, Gonzalez MT, Pereira R, Mota M, Riesco JA, et al. (2011) Validity of spirometry performed online. Eur Respir J. 37 (4):911–918.
    1. Bellia V, Pistelli R, Catalano F, Antonelli I, Grassi V, et al. (2000) Quality control of spirometry in the elderly, Respiratory Health in the Elderly. Am J Respir Crit Care Med. 161:1094–100.
    1. Spirometry Monitoring Technology - SPIROLA. NIOSH - Spirometry in Occupational Medicine. U.S. Department of Health and Human Services, Public Health Service, Centers for Disease Control, National Institute for Occupational Safety and Health. .
    1. de Montes OM, Talamo C, Perez-Padilla R, Jardim JR, Muino A, et al. (2008) Chronic obstructive pulmonary disease and body mass index in five Latin America cities: the PLATINO study. Respir Med. 102 (5):642–50.
    1. Janssens W, Liu Y, Liu D, Kesten S, Tashkin DP, et al. (2013) Quality and reproducibility of spirometry in COPD patients in a randomized trial (UPLIFT), Respir Med. 25.
    1. Perez-Padilla R, Vazquez-Garcia JC, Marquez MN, Menezes AM (2008) Spirometry quality-control strategies in a multinational study of the prevalence of chronic obstructive pulmonary disease. Respir Care 53 (8):1019–26.
    1. Sala T, Rubies C, Gallego C, Muñoz P, Burgos F, et al. (2011) Technical Requirements of Spirometers in the Strategy for Guaranteeing the Access to Quality Spirometry. Arch Bronconeumol. 47(9):466–469.

Source: PubMed

3
Subskrybuj