Long-term safety of posterior chamber implantable phakic contact lens for the correction of myopia

Gitansha Sachdev, Dandapani Ramamurthy, Gitansha Sachdev, Dandapani Ramamurthy

Abstract

Purpose: To review the safety of the implantable phakic contact lens (IPCL V1, Caregroup Sight Solutions, India) for correction of myopia and myopic astigmatism.

Methods: Retrospective interventional case series including 134 eyes which underwent IPCL implantation for correction of myopia and myopic astigmatism at a tertiary eye care hospital in South India. Intraoperative and postoperative complications were analyzed. The following parameters were analyzed at preoperative and 1, 6 and 12 month postoperative visits: visual acuity (in logMAR) and manifest refraction, slit lamp bio microscopy, dilated fundus examination, IOP and endothelial cell density measurement.

Results: No intraoperative complications were noted. Cataract was the most common postoperative complication, wherein three eyes (2.2%) underwent observation for visually insignificant anterior subcapsular cataract and one eye (0.7%) required surgical intervention. Four eyes developed transient increased IOP due to steroid response (3 eyes) and pupillary block glaucoma (1 eye). The endothelial cell loss noted over a 1-year follow-up was 2.01%±4.12% (P=0.71). One eye developed hypopyon on the third postoperative day, which resolved subsequently with an unaided vision of 20/20. No vision threatening complications were noted. The mean follow-up in the cohort was 25.66±16.45 months.

Conclusion: The IPCL is a safe and effective treatment modality for correction of myopia and myopic astigmatism.

Keywords: ICL; IPCL; implantable collamer lens; implantable phakic contact lens; myopia; myopic astigmatism; phakic intraocular lens; refractive correction.

Conflict of interest statement

Disclosure The authors report no conflicts of interest in this work.

Figures

Figure 1
Figure 1
Refractive outcomes following IPCL implantation. Notes: (A) UDVA. (B) UDVA vs CDVA. (C) Change in CDVA. (D) Spherical equivalent refraction attempted vs achieved. (E) Spherical equivalent refraction accuracy. (F) Spherical equivalent refraction stability. (G) Refractive astigmatism. Abbreviations: CDVA, corrected distance visual acuity; IPCL, implantable phakic contact lens; preop, preoperative; postop, postoperative; SEQ, spherical equivalent; UDVA, uncorrected distance visual acuity.

References

    1. Kamiya K, Igarashi A, Shimizu K, Matsumura K, Komatsu M. Visual performance after posterior chamber phakic intraocular lens implantation and wavefront-guided laser in situ keratomileusis for low to moderate myopia. Am J Ophthalmol. 2012;153(6):1178–1186.
    1. Sanders D, Vukich JA. Comparison of implantable collamer lens (ICL) and laser-assisted in situ keratomileusis (LASIK) for low myopia. Cornea. 2006;25(10):1139–1146.
    1. Igarashi A, Kamiya K, Shimizu K, Komatsu M. Visual performance after implantable collamer lens implantation and wavefront-guided laser in situ keratomileusis for high myopia. Am J Ophthalmol. 2009;148(1):164–170.
    1. Pineda-Fernández A, Jaramillo J, Vargas J, Jaramillo M, Jaramillo J, Galíndez A. Phakic posterior chamber intraocular lens for high myopia. J Cataract Refract Surg. 2004;30(11):2277–2283.
    1. Sanders DR, Doney K, Poco M. ICL in Treatment of Myopia Study Group. United States Food and Drug Administration clinical trial of the Implantable Collamer Lens (ICL) for moderate to high myopia: three-year follow-up. Ophthalmology. 2004;111(9):1683–1692.
    1. Igarashi A, Shimizu K, Kamiya K. Eight-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Am J Ophthalmol. 2014;157(3):532–539.
    1. Kamiya K, Shimizu K, Igarashi A, Hikita F, Komatsu M. Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Arch Ophthalmol. 2009;127(7):854–850.
    1. Alfonso JF, Baamonde B, Fernández-Vega L, Fernandes P, González-Méijome JM, Montés-Micó R. Posterior chamber collagen copolymer phakic intraocular lenses to correct myopia: five-year follow-up. J Cataract Refract Surg. 2011;37(5):873–880.
    1. Fernandes P, González-Méijome JM, Madrid-Costa D, Ferrer-Blasco T, Jorge J, Montés-Micó R. Implantable collamer posterior chamber intra-ocular lenses: a review of potential complications. J Refract Surg. 2011;27(10):765–776.
    1. Alfonso JF, Lisa C, Abdelhamid A, Fernandes P, Jorge J, Montés-Micó R. Three-year follow-up of subjective vault following myopic implantable collamer lens implantation. Graefes Arch Clin Exp Ophthalmol. 2010;248(12):1827–1835.
    1. Gonvers M, Bornet C, Othenin-Girard P. Implantable contact lens for moderate to high myopia: relationship of vaulting to cataract formation. J Cataract Refract Surg. 2003;29(5):918–924.
    1. Sanders DR. Anterior subcapsular opacities and cataracts 5 years after surgery in the visian implantable collamer lens FDA trial. J Refract Surg. 2008;24(6):566–570.
    1. Shiratani T, Shimizu K, Fujisawa K, Uga S, Nagano K, Murakami Y. Crystalline lens changes in porcine eyes with implanted phakic IOL (ICL) with a central hole. Graefes Arch Clin Exp Ophthalmol. 2008;246(5):719–728.
    1. Fujisawa K, Shimizu K, Uga S, et al. Changes in the crystalline lens resulting from insertion of a phakic IOL (ICL) into the porcine eye. Graefes Arch Clin Exp Ophthalmol. 2007;245(1):114–122.
    1. Kawamorita T, Shimizu K, Shoji N. Effect of hole size on fluid dynamics of a posterior-chamber phakic intraocular lens with a central perforation by using computational fluid dynamics. Graefes Arch Clin Exp Ophthalmol. 2016;254(4):739–744.
    1. Alfonso JF, Baamonde B, Fernández-Vega L, Fernandes P, González-Méijome JM, Montés-Micó R. Posterior chamber collagen copolymer phakic intraocular lenses to correct myopia: five-year follow-up. J Cataract Refract Surg. 2011;37(5):873–880.
    1. Lee J, Kim Y, Park S, et al. Long-term clinical results of posterior chamber phakic intraocular lens implantation to correct myopia. Clin Exp Ophthalmol. 2016;44(6):481–487.
    1. Shimizu K, Kamiya K, Igarashi A, Kobashi H. Long-term comparison of posterior chamber phakic intraocular lens with and without a central hole (hole ICL and conventional ICL) Implantation for Moderate to High Myopia and Myopic Astigmatism: Consort-Compliant Article. Medicine. 2016;95(14):e3270.
    1. Allan BD, Argeles-Sabate I, Mamalis N. Endophthalmitis rates after implantation of the intraocular collamer lens: survey of users between 1998 and 2006. J Cataract Refract Surg. 2009;35(4):766–769.
    1. Goukon H, Kamiya K, Shimizu K, Igarashi A. Comparison of corneal endothelial cell density and morphology after posterior chamber phakic intraocular lens implantation with and without a central hole. Br J Ophthalmol. 2017;101(11):1461–1465.

Source: PubMed

3
Subskrybuj