Improved pharmacokinetic and pharmacodynamic profile of rapid-acting insulin using needle-free jet injection technology

Elsemiek E C Engwerda, Evertine J Abbink, Cees J Tack, Bastiaan E de Galan, Elsemiek E C Engwerda, Evertine J Abbink, Cees J Tack, Bastiaan E de Galan

Abstract

Objective: Insulin administered by jet injectors is dispensed over a larger subcutaneous area than insulin injected with a syringe, which may facilitate a more rapid absorption. This study compared the pharmacologic profile of administration of insulin aspart by jet injection to that by conventional insulin pen.

Research design and methods: Euglycemic glucose clamp tests were performed in 18 healthy volunteers after subcutaneous administration of 0.2 units/kg body wt of aspart, either administered by jet injection or by conventional pen, using a randomized, double-blind, double-dummy, cross over study design. Pharmacodynamic and pharmacokinetic profiles were derived from the glucose infusion rate (GIR) needed to maintain euglycemia and from plasma insulin levels, respectively.

Results: The time to maximal GIR was significantly shorter when insulin was injected with the jet injector compared with conventional pen administration (51 ± 3 vs. 105 ± 11 min, P < 0.0001). The time to peak insulin concentration was similarly reduced (31 ± 3 vs. 64 ± 6 min, P < 0.0001) and peak insulin concentrations were increased (108 ± 13 vs. 79 ± 7 mU/L, P = 0.01) when insulin was injected by jet injection compared with conventional pen injection. Jet injector insulin administration reduced the time to 50% glucose disposal by ∼40 min (P < 0.0001). There were no differences in maximal GIR, total insulin absorption, or total insulin action between the two devices.

Conclusions: Administration of insulin aspart by jet injection enhances insulin absorption and reduces the duration of glucose-lowering action. This profile resembles more closely the pattern of endogenous insulin secretion and may help to achieve better meal insulin coverage and correction of postprandial glucose excursions.

Figures

Figure 1
Figure 1
Mean GIR after administration of rapid-acting insulin by the jet injector (closed symbols, black line) or the conventional insulin pen (open symbols, dashed line) during the euglycemic glucose clamp.
Figure 2
Figure 2
Mean plasma insulin levels after administration of rapid-acting insulin by the jet injector (closed symbols, black line) or the conventional insulin pen (open symbols, dashed line) during the euglycemic glucose clamp.

References

    1. Mitragotri S. Current status and future prospects of needle-free liquid jet injectors. Nat Rev Drug Discov 2006;5:543–548
    1. Weller C, Linder M. Jet injection of insulin vs the syringe-and-needle method. JAMA 1966;195:844–847
    1. Taylor R, Home PD, Alberti KG. Plasma free insulin profiles after administration of insulin by jet and conventional syringe injection. Diabetes Care 1981;4:377–379
    1. Pehling GB, Gerich JE. Comparison of plasma insulin profiles after subcutaneous administration of insulin by jet spray and conventional needle injection in patients with insulin-dependent diabetes mellitus. Mayo Clin Proc 1984;59:751–754
    1. Hallé JP, Lambert J, Lindmayer I, et al. . Twice-daily mixed regular and NPH insulin injections with new jet injector versus conventional syringes: pharmacokinetics of insulin absorption. Diabetes Care 1986;9:279–282
    1. Malone JI, Lowitt S, Grove NP, Shah SC. Comparison of insulin levels after injection by jet stream and disposable insulin syringe. Diabetes Care 1986;9:637–640
    1. Kerum G, Profozić V, Granić M, Skrabalo Z. Blood glucose and free insulin levels after the administration of insulin by conventional syringe or jet injector in insulin treated type 2 diabetics. Horm Metab Res 1987;19:422–425
    1. Lucas A, Ribas L, Salinas I, Audí L, Sanmartí A, Foz M. Insulin levels after injection by jet stream and disposable syringe. Diabetes Care 1988;11:298–299
    1. Sarno MJ, Bell J, Edelman SV. Pharmacokinetics and glucodynamics of rapid-, short-, and intermediate-acting insulins: comparison of jet injection to needle syringe. Diabetes Technol Ther 2002;4:863–866
    1. Mudaliar SR, Lindberg FA, Joyce M, et al. . Insulin aspart (B28 asp-insulin): a fast-acting analog of human insulin: absorption kinetics and action profile compared with regular human insulin in healthy nondiabetic subjects. Diabetes Care 1999;22:1501–1506
    1. Howey DC, Bowsher RR, Brunelle RL, Woodworth JR. [Lys(B28), Pro(B29)]-human insulin. A rapidly absorbed analogue of human insulin. Diabetes 1994;43:396–402
    1. Heise T, Weyer C, Serwas A, et al. . Time-action profiles of novel premixed preparations of insulin lispro and NPL insulin. Diabetes Care 1998;21:800–803
    1. DeFronzo RA, Tobin JD, Andres R. Glucose clamp technique: a method for quantifying insulin secretion and resistance. Am J Physiol 1979;237:E214–E223
    1. Luijf YM, DeVries JH. Dosing accuracy of insulin pens versus conventional syringes and vials. Diabetes Technol Ther 2010;12(Suppl. 1):S73–S77
    1. Thurman JE. Insulin pen injection devices for management of patients with type 2 diabetes: considerations based on an endocrinologist’s practical experience in the United States. Endocr Pract 2007;13:672–678
    1. Heinemann L, Anderson JH, Jr. Measurement of insulin absorption and insulin action. Diabetes Technol Ther 2004;6:698–718
    1. Abumrad NN, Rabin D, Diamond MP, Lacy WW. Use of a heated superficial hand vein as an alternative site for the measurement of amino acid concentrations and for the study of glucose and alanine kinetics in man. Metabolism 1981;30:936–940
    1. Abbink EJ, van der Wal PS, Sweep CG, Smits P, Tack CJ. Compared to glibenclamide, repaglinide treatment results in a more rapid fall in glucose level and beta-cell secretion after glucose stimulation. Diabetes Metab Res Rev 2004;20:466–471
    1. Abbink EJ, Walker AJ, van der Sluijs HA, Tack CJ, Smits P. No role of calcium- and ATP-dependent potassium channels in insulin-induced vasodilation in humans in vivo. Diabetes Metab Res Rev 2002;18:143–148
    1. Asakura T, Seino H, Nakano R, et al. . A comparison of the handling and accuracy of syringe and vial versus prefilled insulin pen (FlexPen). Diabetes Technol Ther 2009;11:657–661
    1. Bremseth DL, Pass F. Delivery of insulin by jet injection: recent observations. Diabetes Technol Ther 2001;3:225–232
    1. Borg R, Kuenen JC, Carstensen B, et al. ; ADAG Study Group. HbA1(c) and mean blood glucose show stronger associations with cardiovascular disease risk factors than do postprandial glycaemia or glucose variability in persons with diabetes: the A1C-Derived Average Glucose (ADAG) study. Diabetologia 2011;54:69–72

Source: PubMed