In Vivo Evaluation of Operative Torque Generated by Two Nickel-Titanium Rotary Instruments during Root Canal Preparation

Gianluca Gambarini, Massimo Galli, Marco Seracchiani, Dario Di Nardo, Marco A Versiani, Lucila Piasecki, Luca Testarelli, Gianluca Gambarini, Massimo Galli, Marco Seracchiani, Dario Di Nardo, Marco A Versiani, Lucila Piasecki, Luca Testarelli

Abstract

Objectives: This in vivo study evaluated the operative torque and preparation time of ProTaper NEXT (Dentsply Maillefer; Ballaigues, Switzerland) and EdgeFile X7 (EdgeEndo; Albuquerque, New Mexico, United States) rotary systems during root canal preparation of maxillary premolars.

Materials and methods: Ten double-rooted maxillary premolars with independent canals were selected. Each canal in each tooth was prepared with one of the rotary systems (n = 10), ProTaper NEXT or EdgeFile X7. The instruments were rotated at 300 rpm with maximum torque set at 2 N.cm using an electric motor (KaVo; Biberach, Germany) that automatically recorded torque values at every 1/10th of a second (ds).

Statistical analysis: Operative torque (N.cm) and preparation time (s) of the first shaping instrument (size 17/.04) of both rotary systems were recorded and statistically compared using the Mann-Whiney U test with a significance level set at 5%.

Results: No instrument exhibited flute deformation or underwent intracanal failure. No differences were found between the instruments regarding the maximum (peak) torque values (p > 0.05). EdgeFile X7 17/.04 required significantly less preparation time (3.75 seconds interquartile range [IQR]: 3.2-9.0) than ProTaper NEXT X1 (15.45 seconds IQR: 8.35-21.1) (p < 0.05). The median operative torque values of ProTaper NEXT X1 (0.26 N.cm; IQR: 0.18-0.49) were significantly higher compared with EdgeFile X7 17/.04 (0.09 N.cm; IQR: 0.05-0.17) (p < 0.05).

Conclusions: Although no difference was found between the median peak torque values of ProTaper NEXT X1 and EdgeFile X7 17/.04 instruments, the operative torque and instrumentation time results were impacted by their different designs and alloys during clinical preparation of root canals.

Conflict of interest statement

None declared.

Dental Investigation Society.

Figures

Fig. 1
Fig. 1
Cross-sectional aspect of the ProTaper NEXT X1 (A) and Edge–File X7 (B).
Fig. 2
Fig. 2
The instrumentation adopted for the study: motor and handpiece.
Fig. 3
Fig. 3
Operative clinical torque (N.cm) and time (in deciseconds) measured during preparation of the root canals of a same tooth using ProTaper NEXT X1 (A) and EdgeFile X7 17/.04 (B) instruments.

References

    1. Plotino G, Grande N M, Mercadé Bellido M, Testarelli L, Gambarini G. Influence of temperature on cyclic fatigue resistance of ProTaper Gold and ProTaper Universal Rotary files. J Endod. 2017;43(02):200–202.
    1. Testarelli L, Plotino G, Al-Sudani D. Bending properties of a new nickel-titanium alloy with a lower percent by weight of nickel. J Endod. 2011;37(09):1293–1295.
    1. Pedullà E, Grande N M, Plotino G, Gambarini G, Rapisarda E. Influence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments. J Endod. 2013;39(02):258–261.
    1. Gambarini G, Plotino G, Piasecki L, Al-Sudani D, Testarelli L, Sannino G. Deformations and cyclic fatigue resistance of nickel-titanium instruments inside a sequence. Ann Stomatol (Roma) 2015;6(01):6–9.
    1. Gao Y, Shotton V, Wilkinson K, Phillips G, Johnson W B. Effects of raw material and rotational speed on the cyclic fatigue of ProFile Vortex rotary instruments. J Endod. 2010;36(07):1205–1209.
    1. Dentsply Sirona ProTaper NEXT directions for useAvailable at:Accessed July 20th2018
    1. EdgeEndo. EdgeFile X7 directions for use. Available at:Accessed July 20th2018
    1. Dosanjh A, Paurazas S, Askar M. The effect of temperature on cyclic fatigue of nickel-titanium rotary endodontic instruments. J Endod. 2017;43(05):823–826.
    1. Elnaghy A M. Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files. Int Endod J. 2014;47(11):1034–1039.
    1. Nguyen H H, Fong H, Paranjpe A, Flake N M, Johnson J D, Peters O A. Evaluation of the resistance to cyclic fatigue among ProTaper Next, ProTaper Universal, and Vortex Blue rotary instruments. J Endod. 2014;40(08):1190–1193.
    1. Yared G M, Bou Dagher F E, Machtou P. Cyclic fatigue of ProFile rotary instruments after clinical use. Int Endod J. 2000;33(03):204–207.
    1. Lee W, Song M, Kim E, Lee H, Kim H C. A survey of experience-based preference of Nickel-Titanium rotary files and incidence of fracture among general dentists. Restor Dent Endod. 2012;37(04):201–206.
    1. Iqbal M K, Kohli M R, Kim J S. A retrospective clinical study of incidence of root canal instrument separation in an endodontics graduate program: a PennEndo database study. J Endod. 2006;32(11):1048–1052.
    1. Gambarini G, Piasecki L, Di Nardo D et al.Incidence of deformation and fracture of Twisted File Adaptive instruments after repeated clinical use. J Oral Maxillofac Res. 2016;7(04):e5.
    1. Sattapan B, Nervo G J, Palamara J E, Messer H H. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26(03):161–165.
    1. Wu J, Lei G, Yan M, Yu Y, Yu J, Zhang G. Instrument separation analysis of multi-used ProTaper Universal rotary system during root canal therapy. J Endod. 2011;37(06):758–763.
    1. Gambarini G, Seracchiani M, Piasecki L et al. Measurement of torque generated during intra-canal instrumentation in vivo. Int Endod J. 2019;52(05):737–745.
    1. Gambarini G, Seracchiani M, Piasecki L et al.The effect of a brushing motion inside a sequence; an in vivo study. Ann Stomatol (Roma) 2018;9:72–76.
    1. Gambarini G, Tucci E, Bedini R et al.The effect of brushing motion on the cyclic fatigue of rotary nickel titanium instruments. Ann Ist Super Sanita. 2010;46(04):400–404.
    1. Schneider S W. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971;32(02):271–275.
    1. Yared G M, Bou Dagher F E, Machtou P. Influence of rotational speed, torque and operator’s proficiency on ProFile failures. Int Endod J. 2001;34(01):47–53.
    1. Schrader C, Peters O A. Analysis of torque and force with differently tapered rotary endodontic instruments in vitro. J Endod. 2005;31(02):120–123.
    1. Blum J Y, Machtou P, Micallef J P. Location of contact areas on rotary Profile instruments in relationship to the forces developed during mechanical preparation on extracted teeth. Int Endod J. 1999;32(02):108–114.
    1. Jamleh A, Komabayashi T, Ebihara A. Root surface strain during canal shaping and its influence on apical microcrack development: a preliminary investigation. Int Endod J. 2015;48(12):1103–1111.
    1. Sattapan B, Palamara J E, Messer H H. Torque during canal instrumentation using rotary nickel-titanium files. J Endod. 2000;26(03):156–160.
    1. Baek S H, Lee C J, Versluis A, Kim B M, Lee W, Kim H C. Comparison of torsional stiffness of nickel-titanium rotary files with different geometric characteristics. J Endod. 2011;37(09):1283–1286.
    1. Gambarini G, Testarelli L, Milana V et al.Angular deflection of rotary nickel titanium files: a comparative study. Ann Ist Super Sanita. 2009;45(04):423–426.
    1. Boessler C, Paque F, Peters O A. The effect of electropolishing on torque and force during simulated root canal preparation with ProTaper shaping files. J Endod. 2009;35(01):102–106.
    1. Gambarini G, Giansiracusa Rubini A, Sannino G et al.Cutting efficiency of nickel-titanium rotary and reciprocating instruments after prolonged use. Odontology. 2016;104(01):77–81.
    1. Gambarini G, Pompa G, Di Carlo S, De Luca M, Testarelli L. An initial investigation on torsional properties of nickel-titanium instruments produced with a new manufacturing method. Aust Endod J. 2009;35(02):70–72.
    1. Dane A, Capar I D, Arslan H, Akçay M, Uysal B. Effect of different torque settings on crack formation in root dentin. J Endod. 2016;42(02):304–306.
    1. Kim H C, Cheung G S, Lee C J, Kim B M, Park J K, Kang S I. Comparison of forces generated during root canal shaping and residual stresses of three nickel-titanium rotary files by using a three-dimensional finite-element analysis. J Endod. 2008;34(06):743–747.
    1. Gambarini G. Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors. J Endod. 2001;27(12):772–774.
    1. Kwak S W, Ha J H, Cheung G S, Kim H C, Kim S K. Effect of the glide path establishment on the torque generation to the files during instrumentation: an in vitro measurement . J Endod. 2018;44(03):496–500.
    1. Gambarini G, Testarelli L, Galli M, Tucci E, De Luca M. The effect of a new finishing process on the torsional resistance of twisted nickel-titanium rotary instruments. Minerva Stomatol. 2010;59(07)(08):401–406.
    1. Yared G, Bou Dagher F, Kulkarni K. Influence of torque control motors and the operator’s proficiency on ProTaper failures. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003;96(02):229–233.

Source: PubMed

3
Prenumerera