Aortic root surgery with the CARDIOROOT vascular graft: results of a prospective multicenter post-market surveillance study

Giordano Tasca, Jaroslav Lindner, Laurent Barandon, Petr Santavy, Carlo Antona, Jan Burkert, Amando Gamba, Giordano Tasca, Jaroslav Lindner, Laurent Barandon, Petr Santavy, Carlo Antona, Jan Burkert, Amando Gamba

Abstract

Background: Sparing Aortic Valve procedure requires to reproduction of the geometry of the physiologic anatomy of the aortic root. Thus, the materials adopted may make a difference. CARDIOROOT is a one-piece collagen-coated woven vascular graft with pseudo-sinuses, which was designed for use in the treatment of aortic root disease. We report the results of a prospective, multicenter, observational post-market surveillance study evaluating the safety and performance of the CARDIOROOT in patients requiring aortic root surgery.

Methods: Patients with aortic root disease suitable for treatment with a vascular graft with pseudo-sinuses CARDIOROOT graft were eligible for participation. The enrolled patients were assessed intraoperatively, post-operatively, at discharge and at 1-year. Sites assessed complications at each visit, and recorded any reported adverse events. The study endpoint was mortality and complications through 1-year post-procedure.

Results: Fifty-two patients were enrolled from 6 European centers. All procedures were technically successful. Operative mortality was 1.9%: one patient suffered hemorrhagic shock unrelated to the graft 1 day following surgery. At 1-year follow-up the survival rate was 96.2%, with a late death due to pneumonia 5 months post-procedure. Eleven serious adverse events occurred in 7 patients, which included cardiac complications (pericardial effusion, myocardial infarction and ventricular arrhythmia), infection (pericardial infection, deep sternal infection and superficial sternal infection), vascular disorders, including hemorrhagic shock and pleural effusion requiring drainage. Nine of the 11 events were deemed procedure-related by the local investigator, and all were deemed unrelated to the device. There were no reports of graft-related adverse events, infection, occlusion or graft failure.

Conclusions: The results of this 1-year follow-up study showed that the CARDIOROOT vascular graft is safe and effective for the treatment of aneurysmal aortic root in immediate and mid-term follow-up. However, longer term follow-up is needed before conclusions can be made on the long-term safety and effectiveness.

Trial registration: ClinicalTrials.gov Identifier: NCT01609270. Registered 31 May 2012.

Keywords: Aortic root surgery; CARDIOROOT; Sinus of Valsalva; Vascular graft.

Conflict of interest statement

The authors declare that they have no competing interests.

Figures

Fig. 1
Fig. 1
CARDIOROOT Device
Fig. 2
Fig. 2
Cardioroot graft adopted in a Florida-Sleeve sparing aortic valve procedure

References

    1. David TE. Aortic valve sparing operations: outcomes at 20 years. Ann Cardiothorac Surg. 2013;2(1):24–29.
    1. De Paulis R, De Matteis GM, Nardi P, Scaffa R, Buratta MM, Chiariello L. Opening and closing characteristics of the aortic valve after valve-sparing procedures using a new aortic root conduit. Ann Thorac Surg. 2001;72(2):487–494. doi: 10.1016/S0003-4975(01)02747-3.
    1. Matsumori M, Tanaka H, Kawanishi Y, Onishi T, Nakagiri K, Yamashita T, et al. Comparison of distensibility of the aortic root and cusp motion after aortic root replacement with two reimplantation techniques: Valsalva graft versus tube graft. Interact Cardiovasc Thorac Surg. 2007;6(2):177–181. doi: 10.1510/icvts.2006.143289.
    1. Alimi Y, Juhan C, Morati N, Girard N, Cohen S. Dilation of woven and knitted aortic prosthetic grafts: CT scan evaluation. Ann Vasc Surg. 1994;8(3):238–242. doi: 10.1007/BF02018170.
    1. Franke U, Jurmann MJ, Uthoff K, Kohler A, Jurmann B, Wahlers T, et al. In vivo morphology of woven, collagen-sealed Dacron prostheses in the thoracic aorta. Ann Thorac Surg. 1997;64(4):1096–1098. doi: 10.1016/S0003-4975(97)00802-3.
    1. Utoh J, Goto H, Hirata T, Hara M, Kitamura N. Dilatation of sealed Dacron vascular prostheses: a comparison of Gelseal and Hemashield. J Cardiovasc Surg. 1998;39(2):179–180.
    1. Singh C, Wong CS, Wang X. Medical textiles as vascular implants and their success to mimic natural arteries. Journal of functional biomaterials. 2015;6(3):500–525. doi: 10.3390/jfb6030500.
    1. Hirt SW, Aoki M, Demertzis S, Siclari F, Haverich A, Borst HG. Comparative in vivo study on the healing qualities of four different presealed vascular prostheses. J Vasc Surg. 1993;17(3):538–545. doi: 10.1067/mva.1993.39245.
    1. Gamba A, Tasca G, Giannico F, Lobiati E, Skouse D, Galanti A, et al. Early and medium term results of the sleeve valve-sparing procedure for aortic root ectasia. Ann Thorac Surg. 2015;99(4):1228–1233. doi: 10.1016/j.athoracsur.2014.10.044.
    1. Tasca G, Selmi M, Votta E, Redaelli P, Sturla F, Redaelli A, et al. Aortic root biomechanics after sleeve and David sparing techniques: a finite element analysis. Ann Thorac Surg. 2017;103(5):1451–1459. doi: 10.1016/j.athoracsur.2016.08.003.
    1. Tasca G, Selmi M, Votta E. Reply. Ann Thorac Surg. 2018;105(3):984–985. doi: 10.1016/j.athoracsur.2017.08.058.
    1. De Paulis R, Tomai F, Bertoldo F, Ghini AS, Scaffa R, Nardi P, Chiariello L. Coronary flow characteristics after a Bentall procedure with or without sinuses of Valsalva. Eur J Cardiothorac Surg. 2004;26(1):66–72. doi: 10.1016/j.ejcts.2004.04.031.
    1. Weltert L, De Paulis R, Scaffa R, Maselli D, Bellisario A, D'Alessandro S. Re-creation of a sinuslike graft expansion in Bentall procedure reduces stress at the coronary button anastomoses: a finite element study. J Thorac Cardiovasc Surg. 2009;137(5):1082–1087. doi: 10.1016/j.jtcvs.2008.10.013.
    1. Barannyk O, Oshkai P. The influence of the aortic root geometry on flow characteristics of a prosthetic heart valve. J Biomech Eng. 2015;137(5):051005. doi: 10.1115/1.4029747.
    1. Toninato R, Salmon J, Susin FM, Ducci A, Burriesci G. Physiological vortices in the sinuses of Valsalva: an in vitro approach for bio-prosthetic valves. J Biomech. 2016;49(13):2635–2643. doi: 10.1016/j.jbiomech.2016.05.027.