The BpTRU automatic blood pressure monitor compared to 24 hour ambulatory blood pressure monitoring in the assessment of blood pressure in patients with hypertension

Linda Beckett, Marshall Godwin, Linda Beckett, Marshall Godwin

Abstract

Background: Increasing evidence suggests that ABPM more closely predicts target organ damage than does clinic measurement. Future guidelines may suggest ABPM as routine in the diagnosis and monitoring of hypertension. This would create difficulties as this test is expensive and often difficult to obtain. The purpose of this study is to determine the degree to which the BpTRU automatic blood pressure monitor predicts results on 24 hour ambulatory blood pressure monitoring (ABPM).

Methods: A quantitative analysis comparing blood pressure measured by the BpTRU device with the mean daytime blood pressure on 24 hour ABPM. The study was conducted by the Centre for Studies in Primary Care, Queen's University, Kingston, Ontario, Canada on adult primary care patients who are enrolled in two randomized controlled trials on hypertension. The main outcomes were the mean of the blood pressures measured at the three most recent office visits, the initial measurement on the BpTRU-100, the mean of the five measurements on the BpTRU monitor, and the daytime average on 24 hour ABPM.

Results: The group mean of the three charted clinic measured blood pressures (150.8 (SD10.26) / 82.9 (SD 8.44)) was not statistically different from the group mean of the initial reading on BpTRU (150.0 (SD21.33) / 83.3 (SD12.00)). The group mean of the average of five BpTRU readings (140.0 (SD17.71) / 79.8 (SD 10.46)) was not statistically different from the 24 hour daytime mean on ABPM (141.5 (SD 13.25) / 79.7 (SD 7.79)). Within patients, BpTRU average correlated significantly better with daytime ambulatory pressure than did clinic averages (BpTRU r = 0.571, clinic r = 0.145). Based on assessment of sensitivity and specificity at different cut-points, it is suggested that the initial treatment target using the BpTRU be set at <135/85 mmHG, but achievement of target should be confirmed using 24 hour ABPM.

Conclusion: The BpTRU average better predicts ABPM than does the average of the blood pressures recorded on the patient chart from the three most recent visits. The BpTRU automatic clinic blood pressure monitor should be used as an adjunct to ABPM to effectively diagnose and monitor hypertension.

References

    1. 2003 World Health Organization(WHO)/International Society of Hypertension(ISH) statement on management of hypertension. Journal of Hypertension. 2003;21:1983–1992. doi: 10.1097/00004872-200311000-00002.
    1. The Canadian Hypertension recommendations Working group The 2001 Canadian Hypertension Recommendations. Perspectives in Cardiology. 2002. pp. 38–46.
    1. Verdecchia P. Reference values for ambulatory blood pressure and self-measured blood pressure based on prospective outcome data. Blood Pressure Monitoring. 2001;6:323–327. doi: 10.1097/00126097-200112000-00011.
    1. Stassen JA, Thijs L, Fagard R, O'Brien ET, Clement D, de Leeuw PW, Mancia G, Nachev C, Palatini P, Parati G, Tuomilehto J, Webster J. Predicting cardiovascular risk using conventional vs ambulatory blood pressure in older patients with systolic hypertension. Systolic Hypertension in Europe Trial Investigators. JAMA. 1999;282:539–546. doi: 10.1001/jama.282.6.539.
    1. Redon J, Campos C, Narciso ML, Rodico JL, Pascual JM, Ruilope LM. Prognostic value of ambulatory blood pressure monitoring in refractory hypertension: a prospective study. Hypertension. 1998;31:712–718.
    1. Ohkubo T, Imai Y, Tsuji I, Nagai K, Watanabe N, Minami N, Itoh O, Bando T, Sakuma M, Fukao A, Satoh H, Hisamichi S, Abe K. Prediction of mortality by ambulatory blood pressure monitoring versus screening blood pressure measurements: a pilot study in Ohasama. Jour of Hypertension. 1997;15:357–364. doi: 10.1097/00004872-199715040-00006.
    1. Devereux RB, Pickering TG. Relationship between the level, pattern and variability of ambulatory blood pressure and target organ damage in hypertension. J Hypertens Suppl. 1991;9:S34–8.
    1. Parati G, Pomidossi G, Albini F, Malaspina D, Mancia G. Relationship of 24-hour blood pressure mean and variability to severity of target-organ damage in hypertension. Journ Hypertension. 1987;5:93–98.
    1. Zanchetti A, Crepaldi G, Bond MG, Gallus GV, Veglia F, Ventura A, Mancia G, Baggio G, Sampieri L, Rubba P, Collatina S, Serrotti E. Systolic and pulse blood pressures (but not diastolic blood pressure and serum cholesterol) are associated with alterations in carotid intima-media thickness in the moderately hypercholesterolaemic hypertensive patients of the Plaque Hypertension Lowering Italian Study. Journ Hypertens. 2001;19:79–88. doi: 10.1097/00004872-200101000-00011.
    1. Zanchetti A, Bond MG, Hennig M, Neiss A, Mancia G, Dal Palu C, Hansson L, Magnani B, Rahn KH, Reid J, Rodicio J, Safar M, Eckes L, Ravinetto R. Risk factors associated with alterations on carotid intima-media thickness in hypertension: baseline data from the European Lacidipine Study on atherosclerosis. Journ Hypertens. 1998;16:949–961. doi: 10.1097/00004872-199816070-00008.
    1. Vakopoulos NA, Kotsis VT, Pitiriga VC, Toumanidis ST, Lekakis JP, Nanas SN, Vemmos KN, Stamatelopoulos SF, Moulopoulos SD. White coat effect in normotension and hypertension. Blood Press Monit. 2002;7:271–276. doi: 10.1097/00126097-200210000-00004.
    1. Verdecchia P, Gianpaola R, Porcellati C, Schillaci G, Pede S, Bentivoglio M, Angeli F, Norgiolini S, Ambrosio G. Risk of cardiovascular disease in relation to achieved office and ambulatory blood pressure control in treated hypertensive subjects. J Amer Coll Card. 2002;39:878–885. doi: 10.1016/S0735-1097(01)01827-7.
    1. Verdecchia P, Schillaci G, Reboldi G, Franklin SS, Porcellati C. Ambulatory monitoring for prediction of cardiac and cerebral events. Clin and Research Develop. 2001;6:211–216.
    1. Verdecchia P, Angeli F, Gattobigio R. Clinical usefulness of ambulatory blood pressure monitoring. J AM Soc Nephrol. 2004;15:S30–S33. doi: 10.1097/01.ASN.0000093241.62751.95.
    1. Little P, Barnet J, Barnsley L, Marjoram J, Fitzgerald-Barron A, Mant D. Comparison of agreement between different measures of blood pressure in primary care and daytime ambulatory blood pressure. BMJ. 2002;7:293–300.
    1. Myers MG, Valdivieso MA. Use of an automated blood pressure recording device, the BpTRU, to reduce the "white coat effect" in routine practice. Amer Journ Hypertens. 2003;16:494–497. doi: 10.1016/S0895-7061(03)00058-X.
    1. Pavek K, Nilsson G. Patient-specific differences between blood pressure estimated from 24 h ambulatory measurements and serial office self -recordings. Blood Pressure Monitoring. 2002;7:163–168. doi: 10.1097/00126097-200206000-00004.
    1. Little P, Barnett J, Barnsley L, Marjoram J, Fitzgerald-Barron A, Mant D. Comparison of acceptability and preferences for different methods of measuring blood pressure in primary care. BMJ. 2002;325:258–259. doi: 10.1136/bmj.325.7358.258.
    1. Mattu GS, Perry TL, Jr, Wright JM. Comparisons of oscillometric blood pressure monitor (BPM-100Beta) with the auscultatory mercury sphygmomanometer. Blood Pressure Monitoring. 2001;6:153–159. doi: 10.1097/00126097-200106000-00007.
    1. Wright JM, Mattu GS, Perry TL, Jr, Gelfer ME, Strange KD, Zorn A, Chen Y. Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor. Blood Pressure Monitoring. 2001;6:161–165. doi: 10.1097/00126097-200106000-00008.
    1. Mattu GS, Heran BS, Wright JM. Overall accuracy of the BpTRU – an automated electronic blood pressure device. Blood Press Monit. 2004;9:47–52. doi: 10.1097/00126097-200402000-00009.
    1. Palatini P, Frigo G, Bertolo O, Roman E, Da Corta R, Winnicki M. Validation of the A&D TM-2430 device for ambulatory blood pressure monitoring and evaluation of performance according to subjects' characteristics. Blood Press Monit. 1998;3:255–260.

Source: PubMed

3
Prenumerera