Three-Year Results of the IN.PACT SFA Japan Trial Comparing Drug-Coated Balloons With Percutaneous Transluminal Angioplasty
Yoshimitsu Soga, Osamu Iida, Kazushi Urasawa, Shigeru Saito, Michael R Jaff, Hong Wang, Hiroko Ookubo, Hiroyoshi Yokoi, Yoshimitsu Soga, Osamu Iida, Kazushi Urasawa, Shigeru Saito, Michael R Jaff, Hong Wang, Hiroko Ookubo, Hiroyoshi Yokoi
Abstract
Purpose: To evaluate the 3-year safety and effectiveness of the MDT-2113 (IN.PACT Admiral) drug-coated balloon (DCB) vs percutaneous transluminal angioplasty (PTA) in a Japanese population with femoropopliteal occlusive disease.
Materials and methods: The multicenter, prospective, IN.PACT SFA Japan randomized controlled trial (ClinicalTrials.gov identifier NCT01947478) was an independently adjudicated study evaluating Japanese participants randomized 2:1 to DCB (n=68) or PTA (n=32). The effectiveness endpoint was primary patency through 36 months, defined as freedom from clinically-driven target lesion revascularization (CD-TLR) and freedom from restenosis (by duplex ultrasound). The effectiveness endpoint was evaluated using the Kaplan-Meier method; estimates are presented with the 95% confidence intervals (CIs). The safety composite endpoint was freedom from device- and procedure-related death through 30 days and freedom from major target limb amputation and clinically-driven target vessel revascularization through 36 months.
Results: Primary patency by Kaplan-Meier estimate was higher in the DCB group (68.9%, 95% CI 57.5% to 80.2%) vs the PTA group (46.9%, 95% CI 29.6% to 64.2%) at 36 months (log-rank p=0.001). The CD-TLR rates were 14.9% (10/67) for the DCB group and 20.7% (6/29) for PTA (p=0.554). The safety composite endpoint occurred in 83.6% (56/67) of DCB participants and 75.9% (22/29) of PTA participants (p=0.402). All-cause death was similar between groups at 36 months [DCB 6.0% (4/67) vs PTA 6.9% (2/29), p>0.999), with no device- or procedure-related deaths in either group.
Conclusion: The final report of the IN.PACT SFA Japan trial showed that the IN.PACT Admiral DCB is safe and had durable outcomes through 3 years in Japanese participants with femoropopliteal occlusive disease.
Keywords: amputation; balloon angioplasty; drug-coated balloon; femoropopliteal segment; mortality; restenosis; target lesion revascularization.
Conflict of interest statement
Declaration of Conflicting Interests: The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: Hong Wang and Hiroko Ookubo are full-time employees of Medtronic. Michael R. Jaff is a compensated advisor for Biotronix, Medtronic, Philips/Spectranetics, Sanofi, and Vactronix and is an equity investor of PQ Bypass.
Figures
References
- Gerhard-Herman MD, Gornik HL, Barrett C, et al. 2016 AHA/ACC guideline on the management of patients with lower extremity peripheral artery disease: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. J Am Coll Cardiol. 2017;69:1465–1508.
- Bailey SR, Beckman JA, Dao TD, et al. ACC/AHA/SCAI/SIR/SVM 2018 appropriate use criteria for peripheral artery intervention: a report of the American College of Cardiology Appropriate Use Criteria Task Force, American Heart Association, Society for Cardiovascular Angiography and Interventions, Society of Interventional Radiology, and Society for Vascular Medicine. J Am Coll Cardiol. 2019; 73:214–237.
- Feldman DN, Armstrong EJ, Aronow HD, et al. SCAI consensus guidelines for device selection in femoral-popliteal arterial interventions. Catheter Cardiovasc Interv. 2018;92:124–140.
- Tepe G, Laird J, Schneider P, et al. Drug-coated balloon versus standard percutaneous transluminal angioplasty for the treatment of superficial femoral and popliteal peripheral artery disease: 12-month results from the IN.PACT SFA randomized trial. Circulation. 2015;131:495–502.
- Laird JR, Schneider PA, Tepe G, et al. Durability of treatment effect using a drug-coated balloon for femoropopliteal lesions: 24-month results of IN.PACT SFA. J Am Coll Cardiol. 2015;66:2329–2338.
- Schneider PA, Laird JR, Tepe G, et al. Treatment effect of drug-coated balloons is durable to 3 years in the femoropopliteal arteries: long-term results of the IN.PACT SFA randomized trial. Circ Cardiovasc Interv. 2018;11:e005891.
- Micari A, Brodmann M, Keirse K, et al. Drug-coated balloon treatment of femoropopliteal lesions for patients with intermittent claudication and ischemic rest pain: 2-year results from the IN.PACT Global Study. JACC Cardiovasc Interv. 2018;11:945–953.
- Scheinert D, Micari A, Brodmann M, et al. Drug-coated balloon treatment for femoropopliteal artery disease. Circ Cardiovasc Interv. 2018;11:e005654.
- Brodmann M, Keirse K, Scheinert D, et al. Drug-coated balloon treatment for femoropopliteal artery disease: the IN.PACT Global Study de novo in-stent restenosis imaging cohort. JACC Cardiovasc Interv. 2017;23:2113–2123.
- Ansel GM, Brodmann M, Keirse K, et al. Drug-coated balloon treatment of femoropopliteal lesions typically excluded from clinical trials: twelve-month findings from the IN.PACT Global Study. J Endovasc Ther. 2018;25:673–682.
- Micari A, Vadalà G, Castriota F, et al. 1-year results of paclitaxel-coated balloons for long femoropopliteal artery disease: evidence from the SFA-Long study. JACC Cardiovasc Interv. 2016;9:950–956.
- Micari A, Nerla R, Vadalà G, et al. 2-year results of paclitaxel-coated balloons for long femoropopliteal artery disease: evidence from the SFA-Long study. JACC Cardiovasc Interv. 2017;10:728–734.
- Krankenberg H, Tübler T, Ingwersen M, et al. Drug-coated balloon versus standard balloon for superficial femoral artery in-stent restenosis: the randomized Femoral Artery In-stent Restenosis (FAIR) trial. Circulation. 2015;132:2230–2236.
- Chen Z, Guo W, Jiang W, et al. IN.PACT SFA clinical study using the IN.PACT Admiral drug-coated balloon in a Chinese patient population. J Endovasc Ther. 2019;26:471–478.
- Grotti S, Liistro F, Angioli P, et al. Paclitaxel-eluting balloon vs standard angioplasty to reduce restenosis in diabetic patients with in-stent restenosis of the superficial femoral and proximal popliteal arteries: three-year results of the DEBATE-ISR study. J Endovasc Ther. 2016;23:52–57.
- Rosenfield K, Jaff MR, White CJ, et al. Trial of a paclitaxel-coated balloon for femoropopliteal artery disease. N Engl J Med. 2015;373:145–153.
- Krishnan P, Faries P, Niazi K, et al. Stellarex drug-coated balloon for treatment of femoropopliteal disease: twelve-month outcomes from the randomized ILLUMENATE pivotal and pharmacokinetic studies. Circulation. 2017;136: 1102–1113.
- Schroë H, Holden AH, Goueffic Y, et al. Stellarex drug-coated balloon for treatment of femoropopliteal arterial disease-The ILLUMENATE Global Study: 12-month results from a prospective, multicenter, single-arm study. Catheter Cardiovasc Interv. 2018;91:497–504.
- Schroeder H, Werner M, Meyer DR, et al. Low-dose paclitaxel-coated versus uncoated percutaneous transluminal balloon angioplasty for femoropopliteal peripheral artery disease: one-year results of the ILLUMENATE European randomized clinical trial (randomized trial of a novel paclitaxel-coated percutaneous angioplasty balloon). Circulation. 2017;135:2227–2236.
- Brodmann M, Werner M, Meyer DR, et al. Sustainable antirestenosis effect with a low-dose drug-coated balloon: the ILLUMENATE European randomized clinical trial 2-year results. JACC Cardiovasc Interv. 2018;11:2357–2364.
- Xu Y, Jia X, Zhang J, et al. Drug-coated balloon angioplasty compared with uncoated balloons in the treatment of 200 Chinese patients with severe femoropopliteal lesions: 24-month results of AcoArt I. JACC Cardiovasc Interv. 2018;11:2347–2353.
- Bausback Y, Willfort-Ehringer A, Sievert H, et al. Six-month results from the initial randomized study of the Ranger paclitaxel-coated balloon in the femoropopliteal segment. J Endovasc Ther. 2017;24:459–467.
- Tepe G, Gögebakan Ö, Redlich U, et al. Angiographic and clinical outcomes after treatment of femoro-popliteal lesions with a novel paclitaxel-matrix-coated balloon catheter. Cardiovasc Intervent Radiol. 2017;40:1535–1544.
- Tepe G, Schnorr B, Albrecht T, et al. Angioplasty of femoral-popliteal arteries with drug-coated balloons: 5-year follow-up of the THUNDER trial. JACC Cardiovasc Interv. 2015;8:102–108.
- Iida O, Soga Y, Urasawa K, et al. Drug-coated balloon vs standard percutaneous transluminal angioplasty for the treatment of atherosclerotic lesions in the superficial femoral and proximal popliteal arteries: one-year results of the MDT-2113 SFA Japan randomized trial. J Endovasc Ther. 2018;25:109–117.
- Torsello G, Stavroulakis K, Brodmann M, et al. Three-year sustained clinical efficacy of drug-coated balloon angioplasty in a real-world femoropopliteal cohort [published online June 25, 2020]. J Endovasc Ther. doi:10.1177/1526602820931477.
- Mathews SJ. Stellarex in the treatment of the SFA and popliteal: late-breaking 3-year data. Paper presented at: the New Cardiovascular Horizons (NCVH) 2019 Annual Conference; May 29–31, 2019; New Orleans, LA, USA.
- Laird JA, Schneider PA, Jaff MR, et al. Long-term clinical effectiveness of a drug-coated balloon for the treatment of femoropopliteal lesions. Circ Cardiovasc Interv. 2019;12: e007702.
- Jia X, Guo W. New long-term information on PTX coated DCB 5-year results from the AcoArt I study. Abstract presented at: LINC 2020; January 28–31, 2020; Leipzig, Germany.
- Iida O, Soga Y, Urasawa K, et al. Drug-coated balloon versus uncoated percutaneous transluminal angioplasty for the treatment of atherosclerotic lesions in the superficial femoral and proximal popliteal artery: 2-year results of the MDT-2113 SFA Japan randomized trial. Catheter Cardiovasc Interv. 2019;93:664–672.
- Bausback Y, Wittig T, Schmidt A, et al. Drug-eluting stent versus drug-coated balloon revascularization in patients with femoropopliteal arterial disease. J Am Coll Cardiol. 2019;73: 667–679.
- Bunte MC, Cohen DJ, Jaff MR, et al. Long-term clinical and quality of life outcomes after stenting of femoropopliteal artery stenosis: 3-year results from the STROLL study. Catheter Cardiovasc Interv. 2018;92:106–114.
- Laird JR, Katzen BT, Scheinert D, et al. Nitinol stent implantation vs. balloon angioplasty for lesions in the superficial femoral and proximal popliteal arteries of patients with claudication: three-year follow-up from the RESILIENT randomized trial. J Endovasc Ther. 2012;19:1–9.
- Müller-Hülsbeck S, Keirse K, Zeller T, et al. Long-term results from the MAJESTIC trial of the Eluvia paclitaxel-eluting stent for femoropopliteal treatment: 3-year follow-up. Cardiovasc Intervent Radiol. 2017;40:1832–1838.
- Rocha-Singh KJ, Bosiers M, Schultz G, et al. A single stent strategy in patients with lifestyle limiting claudication: 3-year results from the Durability II trial. Catheter Cardiovasc Interv. 2015;86:164–170.
Source: PubMed