Re-engineering ventilatory support to decrease days and improve resource utilization

O C Kirton, C B DeHaven, J Hudson-Civetta, J P Morgan, J Windsor, J M Civetta, O C Kirton, C B DeHaven, J Hudson-Civetta, J P Morgan, J Windsor, J M Civetta

Abstract

Objective: The objective of this study was to describe the development of a cost-effective ventilatory strategy using a portable microprocessor-controlled respiratory monitor (Bicore CP-100; Allied Healthcare Products, Riverside, CA.)

Summary background data: Until recently, clinicians have had to accept the uncertainties of clinical judgment, which unfortunately, often biased the patient to a prolonged ventilatory course to avoid extubation failures, necessitating reintubation.

Methods: Over a 4-year period, the authors attempted to re-engineer the process of ventilatory support based on measured work of breathing (WOB), including physiologic (WOBPhys), imposed (WOBImp) and total (WOBTot).

Results: The authors made 90 determinations of WOB in 31 patients. The coefficient of determination (i2) of WOBTot, with the breathing frequency was 0.35, with tidal volume was 0.10, and with the rapid shallow breathing index (f/V(tau)) was 0.23; therefore, the authors discarded them as reliable inferences. Of 27 patients ventilated for > 2 days with satisfactory blood gases, but with breathing frequency > 30 breaths/minute, 6 had WOBTot < 0.8 J/L and were extubated successfully. In 21 patients, WOBTot was elevated to 1.6 +/- 0.83 J/L, WOBImp was 1.1 +/- 0.64 J/L, approximately twice the WOBPhys (0.5 +/- 0.26 J/L), a normal value. Extubation was successful in 20 of those 21 patients. This approach was extended to the spontaneous breathing pre-extubation trial. In addition, the ventilator was adjusted so that the patient sustained a WOBTot of 0.6 to 1 J/L during the ventilatory support. This evolution was tracked for 18 months in a series of 838 trauma intensive care unit patients. Average duration of ventilation decreased from 8.2 to 4.2 days (49%; p < 0.01). This translated into approximately 2400 decreased ventilator days per year.

Conclusion: Objective measurement to guide the adequacy of ventilatory support and interpret apparent clinical weaning failures decreased total ventilatory time by 50%, permitting extubation in nearly 20% of patients previously considered failures.

References

    1. Crit Care Med. 1990 Jan;18(1):103-13
    1. Crit Care Med. 1989 Jul;17(7):671-7
    1. N Engl J Med. 1991 May 23;324(21):1445-50
    1. Am Rev Respir Dis. 1991 Sep;144(3 Pt 1):531-7
    1. Chest. 1992 Dec;102(6):1642-3
    1. Crit Care Med. 1993 Feb;21(2):171-3
    1. Crit Care Med. 1993 Feb;21(2):183-90
    1. Crit Care Med. 1993 May;21(5):790-5
    1. Crit Care Med. 1993 Sep;21(9):1333-8
    1. Chest. 1993 Nov;104(5):1445-51
    1. Crit Care Med. 1994 Mar;22(3):515-23
    1. Chest. 1995 Oct;108(4):1021-5
    1. Chest. 1995 Nov;108(5):1338-44
    1. Crit Care Med. 1996 Jun;24(6):976-80
    1. Surg Gynecol Obstet. 1973 Mar;136(3):367-70
    1. Crit Care Med. 1977 Nov-Dec;5(6):264-8
    1. Am Rev Respir Dis. 1982 Nov;126(5):788-91
    1. Arch Intern Med. 1984 May;144(5):1012-6
    1. Crit Care Med. 1986 Feb;14(2):92-4
    1. Anesth Analg. 1986 Aug;65(8):853-9
    1. Chest. 1988 Mar;93(3):499-505
    1. Chest. 1988 Apr;93(4):795-9
    1. Chest. 1988 Aug;94(2):227-8
    1. Am J Med. 1990 Mar;88(3):268-78

Source: PubMed

3
Prenumerera