Comparing Laboratory Blood Glucose Results With HemoCue Glucose 201 RT (CFRD)
To Compare the Clinical Accuracy of Glucose Measurement During Oral Glucose Tolerance Test Using Two Methods: i) Laboratory Standard Technique, ii) HemoCue Glucose 201 RT System
There are very few studies comparing the accuracy, sensitivity and specificity of HemoCue glucose 201 RT system with the laboratory gold standard.
The incidence of cystic fibrosis related diabetes (CFRD) has risen significantly as patients' survival improves. Around 30% of all cystic fibrosis (CF) adult patients have CFRD. Early diagnosis of CFRD is important to slow the deterioration of lung function and nutritional status, both of which increase mortality.
The oral glucose tolerance test (OGTT) is the accepted method for detecting CFRD and the Cystic Fibrosis Trust guidelines recommend that patients with CF over the age of twelve years should be screened annually.
The World Health Organisation has pointed out that due to the absence of a more specific biological marker to define diabetes, plasma glucose estimation remains the basis of diagnostic criteria. WHO recommends that venous plasma glucose should be the standard method for measuring and reporting glucose concentration in blood. However, it has also recognised that there is a widespread use of capillary sampling. Although fasting values for venous and capillary plasma glucose are the same, in the nonfasting state capillary samples will give higher results than venous samples, and glucose values require conversion which can be problematic.
HemoCue Glucose 201 RT (HGS 201 RT) analyzer with plasma conversion provides laboratory quality test results at the point of care for the diagnosis, screening and monitoring of diabetes mellitus. The analyzer is easy to use and the system can be operated by nonlaboratory personnel, also the results can be presented immediately, leading to significant time and resource saving. The patient also benefits from having the result immediately instead of waiting for several hours. With immediate result, the doctor and health care professionals can provide education, support and reassurance, and implement care plans, if appropriate, at the earliest opportunity.
The aim of the study is to compare the clinical accuracy of HemoCue Glucose 201 RT with the laboratory standard method in the analysis of blood glucose concentration during oral glucose tolerance test.
Study Overview
Detailed Description
All CF patients attending annual review who require an OGTT will have a venous and a capillary blood sample taken simultaneously before (fasting = 0 min) and 2 hours (post glucose load = 120 min) after the administration of 75 grams Glucose solution orally.
The capillary sample will be analysed immediately using HGS 201 RT. The venous sample will be collected in a Fluoride Oxalate tube (WHO guideline) and sent to the Trust's laboratory for analysis as per usual practice.
Study Type
Enrollment (Actual)
Contacts and Locations
Study Locations
-
-
-
London, United Kingdom, SW3 6NP
- Royal Brompton Hospital
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
The first 70 consecutive suitable consenting adult CF patients, who are 16 years of age and over, attending annual review at the Royal Brompton Hospital from June 2013 to March 2014, are eligible for the study
Exclusion Criteria:
Patients with an existing diagnosis of CFRD are exclude.
Study Plan
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
Intervention / Treatment |
|---|---|
|
Laboratory blood glucose
Blood glucose value from Lab
|
The glucose value from the HemoCue Glucose 201 RT will be compared with the Trust Laboratory value.
Other Names:
|
|
HemoCue Glucose 201 RT
Blood glucose value from HemoCue Glucose 201 RT
|
The glucose value from the HemoCue Glucose 201 RT will be compared with the Trust Laboratory value.
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Does HemoCue Glucose 201 RT provide lab-quality results in the analysis of glucose values during oral glucose tolerance tests?
Time Frame: One year
|
The glucose value from the HemoCue Glucose 201 RT will be compared with the Trust Laboratory value.
The sensitivity, specificity and clinical significance will be calculated.
|
One year
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Is HemoCue Glucose 201 RT method cost saving relative to the traditional laboratory method for analysing glucose values taken during oral glucose tolerance tests?
Time Frame: One year
|
The cost of using the HemoCue Glucose 201 RT will be compared with Trust Laboratory cost for analysing the glucose value from OGTT.
|
One year
|
Other Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
2. To see if the HemoCue Glucose 201 RT offers sufficient advantages to make it a viable alternative to conventional testing
Time Frame: One year
|
If using HemoCue Glucose 201 RT is shown to offer cost and clinical advantages over conventional testing then look to implement as standard test method and report more widely.
|
One year
|
Collaborators and Investigators
Investigators
- Principal Investigator: Nicholas J Simmonds, MD MRCP, Royal Brompton & Harefield NHS Foundation Trust
Study record dates
Study Major Dates
Study Start
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Estimate)
Study Record Updates
Last Update Posted (Estimate)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- 2012CF007B
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.