Evaluation of ExSpiron™ Measurements Derived Without Patient Specific Calibration (RMI UVU)
Evaluation of ExSpiron™ Non-Invasive Respiratory Volume Monitor Measurements Derived Without Patient Specific Calibration
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
In order to evaluate the performance of the device under the range of conditions encountered in the clinical environment, the respiratory volume monitor (RVM) was tested throughout a range of respiratory rates and also during erratic breathing patterns. At 4 and 36 breaths per minute (the extremes of the rates anticipated to be encountered in the clinical setting), and during erratic breathing, RVM and spirometry measurements were found to be highly-correlated (r=0.98 ±0.01 (mean ±95%CI) at 4 breaths/min, 0.97 ±0.01 at 36, and 0.93 ±0.04 during erratic breathing). These studies also included calibration of the ExSpiron to the spirometer.
The primary objective of the study is to compare the minute ventilation (MV), tidal volume (TV) and respiratory rate (RR) measurements collected and reported by the ExSpiron, without prior ExSpiron calibration, against acquired MV, TV, and RR measurements from an ExSpiron with prior individual specific calibration to a spirometer using either a Wright Spirometer or a Morgan Flow Volume Loop (FVL) Pneumotachograph Spirometer. A secondary objective will be to confirm the PadSet functionality after 24 hours of wear in accordance with current labeling.
Study Type
Enrollment (Actual)
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- Males and Females aged 21-90
- Able to provide written informed consent
- Able to participate in a two day study
Exclusion Criteria:
- Men and Women aged <21 years and >90 years of age
- Inability to follow study and/or study staff instructions
- Inability to perform study procedures, e.g. spirometry
- Allergy to the electrodes or components of the electrodes
Study Plan
How is the study designed?
Design Details
- Observational Models: Cohort
- Time Perspectives: Prospective
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The primary outcome measure of the study is the minute ventilation (MV), in liters per minute (L/min), collected and reported by the ExSpiron, with and without prior ExSpiron calibration
Time Frame: 3 months
|
Calibrated ExSpiron measurements compared against uncalibrated ExSpiron measurements
|
3 months
|
|
The primary outcome measure of the study is the tidal volume (TV), in liters (L), collected and reported by the ExSpiron, with and without prior ExSpiron calibration
Time Frame: 3 months
|
Calibrated ExSpiron measurements compared against uncalibrated ExSpiron measurements
|
3 months
|
|
The primary outcome measure of the study is the respiratory rate (RR), in breaths per minute (bpm), collected and reported by the ExSpiron, with and without prior ExSpiron calibration
Time Frame: 3 months
|
Calibrated ExSpiron measurements compared against uncalibrated ExSpiron measurements
|
3 months
|
Collaborators and Investigators
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Estimate)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Other Study ID Numbers
- RMI UVU 2015-0052
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.