System Accuracy Evaluation of 2 CE-marked Blood Glucose Monitoring System

The objective of this study is the evaluation of system accuracy following ISO 15197:2013 (E), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated.

However in this study, system accuracy evaluation will be performed for Hemocue Glucose 201+ and Hemocue Glucose 201 RT which are both glucose monitoring systems for professional use only on behalf of the IDT with one reagent system lot for each BGMS.

Main goal is the evaluation of system accuracy of both of the hemocue Systems. (english)

Study Overview

Status

Completed

Conditions

Study Type

Interventional

Enrollment (Actual)

109

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Ulm, Germany, 89081
        • Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Male or female, with type 1 diabetes, type 2 diabetes or subjects without diabetes
  • Signed informed consent form
  • Minimum age of 18 years
  • Subjects are legally competent and capable to understand character, meaning and consequences of the study.
  • If blood glucose values < 80 mg/dl or > 300 mg/dl shall be measured after short term alteration in insulin therapy:
  • Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.
  • Signature of subjects to document consent with these procedures on informed consent form.

Exclusion Criteria:

  • Pregnancy or lactation period
  • Severe acute disease (at the study physician's discretion)
  • Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
  • Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
  • Being unable to give informed consent
  • < 18 years
  • Legally incompetent
  • Being committed to an institution (e.g. psychiatric clinic)
  • Language barriers potentially compromising an adequate compliance with study procedures
  • Dependent on investigator or sponsor
  • If blood glucose values < 80 mg/dl shall be measured after short term alteration in insulin therapy (concentration category 2, see Fehler! Verweisquelle konnte nicht gefunden werden.), subjects with type 1 diabetes, suffering from:
  • Coronary heart disease
  • Condition after myocardial infarction
  • Condition after cerebral events
  • Peripheral arterial occlusive disease
  • Hypoglycemia unawareness

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
System accuracy criteria (see description)
Time Frame: For each subject, the experimental phase has an expected duration of up to 6 hours

System accuracy criteria

  • Following ISO 15197:2013 (E), the BGMS shall meet both of the following criteria:
  • Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
  • Each lot shall pass acceptability criterion A.
  • Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes.
For each subject, the experimental phase has an expected duration of up to 6 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

March 1, 2016

Primary Completion (Actual)

April 1, 2016

Study Completion (Actual)

May 1, 2016

Study Registration Dates

First Submitted

March 11, 2016

First Submitted That Met QC Criteria

March 11, 2016

First Posted (Estimate)

March 17, 2016

Study Record Updates

Last Update Posted (Estimate)

August 31, 2016

Last Update Submitted That Met QC Criteria

August 30, 2016

Last Verified

August 1, 2016

More Information

Terms related to this study

Other Study ID Numbers

  • IDT-1542-IU

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Diabetes

Clinical Trials on Blood glucose monitoring systems for point-of-care-testing

3
Subscribe