Clinical System Accuracy Evaluation With OneTouch® Verio Flex Blood Glucose Monitoring System (BGMS)

The objective of this study is the evaluation of system accuracy based on EN ISO 15197:2015 (ISO 15197:2013), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated. ISO 15197:20132 was harmonized with the regulations of the European Union as EN ISO 15197:2015. This harmonization had no impact on the requirements and procedu¬res in ISO 15197:2013; changes were made to the foreword and an informative annex. In this study, system accuracy evaluation will be performed for OneTouch® Verio Flex (LifeScan Europe) on behalf of LifeScan Scotland Limited (customer) with three different reagent system lots.

Study Overview

Study Type

Interventional

Enrollment (Actual)

109

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Baden-Württemberg
      • Ulm, Baden-Württemberg, Germany, 89081
        • Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

Yes

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Male or female, with type 1 diabetes, type 2 diabetes (or subjects without diabetes)
  • Signed informed consent form
  • Minimum age of 18 years
  • Subjects are legally competent and capable to understand character, meaning and consequences of the study.
  • If blood glucose values < 80 mg/dl or > 300 mg/dl shall be measured after short term alteration in insulin therapy:
  • Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.
  • Signature of subjects to document consent with these procedures on informed consent form.

Exclusion Criteria:

  • Pregnancy or lactation period
  • Severe acute disease (at the study physician's discretion)
  • Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
  • Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
  • Being unable to give informed consent
  • < 18 years
  • Legally incompetent
  • Being committed to an institution (e.g. psychiatric clinic)
  • Language barriers potentially compromising an adequate compliance with study procedures
  • Dependent on investigator or sponsor
  • If blood glucose values < 80 mg/dl shall be measured after short term alteration in insulin therapy (concentration category 2), subjects with type 1 diabetes, suffering from:
  • Coronary heart disease
  • Condition after myocardial infarction
  • Condition after cerebral events
  • Peripheral arterial occlusive disease
  • Hypoglycemia unawareness

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: OTHER
  • Allocation: NA
  • Interventional Model: SINGLE_GROUP
  • Masking: NONE

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
EXPERIMENTAL: Blood glucose measurement

Measurement procedure

  • Subjects will be asked to wash and dry their hands before the measurements.
  • The measurement procedure for each reagent system lot of a BGMS will be as follows:
  • Study personnel will collect a capillary blood sample from the subject's fingertip by skin puncture.
  • Duplicate measurements on the sample will be consecutively performed with 2 test meters using reagent system units from the same vial / package.
  • Capillary blood for measurements with all 3 test strip lots and for both comparison measurements will be taken from the same finger puncture.
  • The sample will be applied to the reagent system units following specifications in the manufacturer's labelling.
Other Names:
  • OneTouch® Verio Flex

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
System accuracy criteria
Time Frame: For each subject, the experimental phase has an expected duration of up to 6 hours

Based on EN ISO 15197:2015 (ISO 15197:2013) (E), the BGMS shall meet both of the following criteria:

Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).

Each lot shall pass acceptability criterion A. Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes.

For each subject, the experimental phase has an expected duration of up to 6 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (ACTUAL)

March 16, 2017

Primary Completion (ACTUAL)

April 3, 2017

Study Completion (ACTUAL)

June 1, 2017

Study Registration Dates

First Submitted

March 15, 2017

First Submitted That Met QC Criteria

March 15, 2017

First Posted (ACTUAL)

March 21, 2017

Study Record Updates

Last Update Posted (ACTUAL)

September 26, 2017

Last Update Submitted That Met QC Criteria

September 25, 2017

Last Verified

September 1, 2017

More Information

Terms related to this study

Other Study ID Numbers

  • IDT-1653-LI

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Diabetes Mellitus

Clinical Trials on Blood glucose monitoring systems for self-testing (BGMS)

Subscribe