Evaluation of a Digital Microscope for Malaria (EasyScanGo)

September 1, 2020 updated by: University of Oxford

A Multi-Centric Evaluation of a Device for Automated Malaria Microscopy (EasyScan Go)

Light microscopy, which is based on century-old technology, remains a key indicator in drug efficacy testing performed in the context of clinical trials for monitoring existing antimalarial drugs or in the context of regulatory clinical trials for registration of new drugs. It is one of the main diagnostic methods for malaria diagnosis in general, as in an ideal setting it can provide low-cost accurate diagnosis, determine the density of parasites in the blood, and accurately differentiate between different malaria parasite species, characteristics vital to the implementation of global plans for drug efficacy monitoring. Malaria rapid tests (RDTs), while useful for rapid diagnosis and case management, do not provide information on the parasite density nor the species differentiation necessary for research and drug efficacy assessment. Microscopy therefore retains key advantages over a number of newer technologies, but its reliability is severely impeded by dependence on high technical competence of the human operators as well as availability of high quality equipment and reagents. Recent studies have demonstrated frequent poor specificity and sensitivity associated with manual microscopy diagnostics in operational conditions. These drawbacks constitute a major limiting factor to effective monitoring and preservation of vital anti-malarial medicines.

Advances in digital microscopy performance and affordability have now opened the door to potentially significant improvements in the performance of malaria microscopy, overcoming serious deficiencies in current drug efficacy assessment, and more broadly in malaria diagnosis and management. Global Good (GG)/Intellectual Ventures Laboratory (IVL) sponsored by the Global Good Fund, has developed a microscope prototype consisting of low cost components to scan and capture images from Giemsa-stained thick blood films on slides. The captured images are analyzed with custom image analysis software developed at GG/IVL, using algorithms that are designed for automatic malaria diagnosis, without user input. Versions of a prototype of the device were first tested in field settings in Thailand in 2014-2015 at clinics operated by the Shoklo Malaria Research Unit (SMRU) and then again in 2016-2017. When compared to expert microscopy at SMRU, the performance of the device with respect to diagnostic sensitivity (87.8%), species identification (85.6% species correctly identified) and parasite density estimation (44% of estimates within +/-25% of reference microscopy result) corresponded to WHO Competence Level 2. The device and the accompanying image analysis algorithms have since been further developed and a new, third version of the prototype is now available for testing in diverse settings with varying malaria prevalence and user expertise.

Study Overview

Status

Completed

Conditions

Detailed Description

The primary purpose of this evaluation is to quantify the diagnostic performance of the EasyScan Go prototype in various field settings. The performance of the EasyScan Go prototype will be assessed by scanning of negative and positive slides with the EasyScan Go and comparing the results with expert microscopy. Plasmodium genus- and species-specific PCR will also be performed on samples collected at some sites as an additional confirmatory test for the detection of malaria parasites and their species if present. Testing by microscopy and EasyScan Go will be performed in field clinic settings on Giemsa-stained slides prepared from febrile patient blood collected from a finger-prick. Further work will be undertaken at the WWARN laboratory in Bangkok for data analyses and for quality assurance.

Funder: Intellectual Ventures Lab/Global Good (2018) Sponser: University of Oxford Grant refernce number:The Global Good Fund I, LLC PA No.5

Study Type

Observational

Enrollment (Actual)

2250

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Tak
      • Mae Sot, Tak, Thailand
        • Shoklo Malaria Research Unit

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

6 months to 75 years (Child, Adult, Older Adult)

Accepts Healthy Volunteers

N/A

Genders Eligible for Study

All

Sampling Method

Probability Sample

Study Population

The multi-centric evaluation is currently planned for implementation at relatively few sites/countries and the primary criteria for the selection of sites will be based on operational feasibility of the project, i.e., availability of resources to perform the study procedures, adequate expected numbers of malaria positive cases, etc.

Potential sites in Senegal, Kenya, Tanzania, Uganda, Congo, South Africa, Burkina Faso, Brazil, Thailand, Indonesia, Bangladesh, Myanmar and Cambodia are being considered for the study.

A minimum of 80 malaria cases confirmed by expert microscopy and a minimum of 80 malaria-negative cases per study site.

Description

Inclusion Criteria:

  • Male or female subjects, age ≥ 6 months to 75 years
  • Febrile at presentation or history of fever in the past 48 hours (≥ 37.5 ºC) and no other obvious diagnosis or cause for fever, warranting malaria investigation under routine clinical practice.
  • Individual informed assent/consent obtained

Exclusion Criteria:

- Signs of severe malaria as defined by WHO

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Time Frame
Diagnostic sensitivity for malaria parasite detection
Time Frame: 6 months
6 months
Diagnostic specificity for malaria parasite detection
Time Frame: 6 months
6 months
Bland-Altman plots for parasite density estimation
Time Frame: 6 months
6 months
Kappa statistic for parasite species identification
Time Frame: 6 months
6 months

Secondary Outcome Measures

Outcome Measure
Time Frame
Reliability (comparison between 2 devices and repeat reads)
Time Frame: 6 months
6 months
Cost-effectiveness as compared with routine methods
Time Frame: 6 months
6 months
Prevalence of parasite genetic markers of resistance to antimalarials by location and time period
Time Frame: 6 months
6 months
Prevalence of parasites carrying deletions of pfhrp2/3 genes by location and time period
Time Frame: 6 months
6 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

June 25, 2018

Primary Completion (Actual)

October 31, 2019

Study Completion (Actual)

June 30, 2020

Study Registration Dates

First Submitted

April 18, 2018

First Submitted That Met QC Criteria

April 18, 2018

First Posted (Actual)

May 1, 2018

Study Record Updates

Last Update Posted (Actual)

September 3, 2020

Last Update Submitted That Met QC Criteria

September 1, 2020

Last Verified

September 1, 2020

More Information

Terms related to this study

Other Study ID Numbers

  • MAL18002

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Malaria

3
Subscribe