Tato stránka byla automaticky přeložena a přesnost překladu není zaručena. Podívejte se prosím na anglická verze pro zdrojový text.

Validation of Noninvasive Hemoglobin, Comparison of PVI to Standard Methods, and Validation of Clinical Efficacy of ORI

23. září 2021 aktualizováno: Masimo Corporation

Validation of a Pulse Oximetry Based Hemoglobin Measurement System. Comparison of Plethysmographic Variability Index to Standard Methods of Guiding Fluid and Transfusion Management in the Critically Ill Adult Patients. Validation of Clinical Efficacy of ORI for Supplemental Oxygen Titration

The study goal is to establish the correlation of hemoglobin values determined by the noninvasive pulse oximetry-determined hemoglobin (SpHb) system with hemoglobin as measured by other validated devices. The study will also evaluation oxygen reserve index (ORI) to correlate with partial pressure of oxygen (PaO2), and to evaluate pleth variability index (PVI) with stroke volume variation (SVV) or pulse pressure variation (PPV) values.

Přehled studie

Postavení

Dokončeno

Podmínky

Intervence / Léčba

Typ studie

Intervenční

Zápis (Aktuální)

50

Kontakty a umístění

Tato část poskytuje kontaktní údaje pro ty, kteří studii provádějí, a informace o tom, kde se tato studie provádí.

Studijní místa

    • California
      • Loma Linda, California, Spojené státy, 92354
        • Loma Linda University Medical Center

Kritéria účasti

Výzkumníci hledají lidi, kteří odpovídají určitému popisu, kterému se říká kritéria způsobilosti. Některé příklady těchto kritérií jsou celkový zdravotní stav osoby nebo předchozí léčba.

Kritéria způsobilosti

Věk způsobilý ke studiu

18 let a starší (Dospělý, Starší dospělý)

Přijímá zdravé dobrovolníky

Ne

Pohlaví způsobilá ke studiu

Všechno

Popis

Inclusion Criteria:

  • Admitted into the ICU
  • Aged 18 years or older
  • Mechanically ventilated
  • Arterial line placed
  • Vigileo/Flotrac System being used for guidance of fluid management

Exclusion Criteria:

  • Pregnancy
  • Prisoner status
  • Extreme hemodynamic instability
  • Multiple vasopressors in use with questionable peripheral blood flow
  • Lack of appropriate sites for sensor placement
  • Patient or patient's legal representative refusal

Studijní plán

Tato část poskytuje podrobnosti o studijním plánu, včetně toho, jak je studie navržena a co studie měří.

Jak je studie koncipována?

Detaily designu

  • Primární účel: Jiný
  • Přidělení: N/A
  • Intervenční model: Přiřazení jedné skupiny
  • Maskování: Žádné (otevřený štítek)

Zbraně a zásahy

Skupina účastníků / Arm
Intervence / Léčba
Experimentální: Test group
All subjects will be enrolled into the test group and will receive Masimo Pulse CO-Oximeter for measurement of various physiological parameters.
Noninvasive pulse oximeter with various features for measurement of hemoglobin, pleth variability index, and oxygen reserve index

Co je měření studie?

Primární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Noninvasive Pulse Oximetry-determined SpHb Bias
Časové okno: Duration of study participation is up to three days.
Accuracy of SpHb will be determined by comparing it to the hemoglobin value obtained from a reference blood sample and calculating the accuracy root mean square (Arms) error value. In order to obtain the Arms value, the blood sample hemoglobin value is subtracted from SpHb value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.
Duration of study participation is up to three days.
Correlation of PVI Measurement Compared With PPV or SVV
Časové okno: Duration of study participation is up to three days.
Correlation of PVI measurement compared with pulse pressure variation PPV or stroke volume variation (SVV) measurement from arterial catheter
Duration of study participation is up to three days.
Correlation of "ORI" With Partial Pressure of Oxygen (PaO2) "PaO2" Measurement From Blood Gas Analysis
Časové okno: Duration of study participation is up to three days.
Concordance will be determined by comparing the trending ability of the Oxygen Reserve Index (ORI) measurement of the sensor with that of Partial Pressure of Oxygen in Blood (PaO2).
Duration of study participation is up to three days.

Spolupracovníci a vyšetřovatelé

Zde najdete lidi a organizace zapojené do této studie.

Sponzor

Termíny studijních záznamů

Tato data sledují průběh záznamů studie a předkládání souhrnných výsledků na ClinicalTrials.gov. Záznamy ze studií a hlášené výsledky jsou před zveřejněním na veřejné webové stránce přezkoumány Národní lékařskou knihovnou (NLM), aby se ujistily, že splňují specifické standardy kontroly kvality.

Hlavní termíny studia

Začátek studia (Aktuální)

9. března 2017

Primární dokončení (Aktuální)

11. října 2017

Dokončení studie (Aktuální)

11. října 2017

Termíny zápisu do studia

První předloženo

13. července 2016

První předloženo, které splnilo kritéria kontroly kvality

21. července 2016

První zveřejněno (Odhad)

22. července 2016

Aktualizace studijních záznamů

Poslední zveřejněná aktualizace (Aktuální)

21. října 2021

Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality

23. září 2021

Naposledy ověřeno

1. září 2021

Více informací

Termíny související s touto studií

Další identifikační čísla studie

  • DORO0001

Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .