Smartphone Versus Back Range of Motion: Validity and Reliability in Lumbar Range of Motion Assessment
Clinimetric Properties of Smartphone Application as a Suitable Alternative to Back Range of Motion Device in Assessing the Back Range of Motion
Přehled studie
Postavení
Postavení
Podmínky
Podmínky
Intervence / Léčba
Intervence / Léčba
Detailní popis
Typ studie
Typ studie
Zápis (Aktuální)
Zápis
Kontakty a umístění
Studijní místa
-
-
-
Giza, Egypt, 12613
- Cairo University
-
-
Kritéria účasti
Kritéria způsobilosti
Kritéria způsobilosti
Věk způsobilý ke studiu
- Dospělý
Přijímá zdravé dobrovolníky
Metoda odběru vzorků
Studijní populace
Popis
Inclusion Criteria:
- Age 18-25 years old
- Healthy Male university students
- All participants have normal BMI
Exclusion Criteria:
- Recent back, pelvic, or abdominal surgeries, traumatic injury to the back .
- Lower limb deformities, e.g., leg length discrepancy.
- Adolescent idiopathic scoliosis
- Any overweight or obese participants
Studijní plán
Jak je studie koncipována?
Detaily designu
Co je měření studie?
Primární výstupní opatření
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
assessment of flexion and extension range of motion using inclinometer
Časové okno: at baseline
|
During lumbar flexion and extension measurements using the Back Range of Motion (BROM) device, patients stand upright with extended knees and feet approximately 17 centimeters apart.
For flexion, they bend forward to touch the floor, while for extension, they bend backward with crossed arms.
An inclinometer is secured over the sacrum, and its position is centered to record readings in degrees.
An L-shaped slide arm measures the distance between the T12 and S1 spinous processes.
The device isolates lumbar motion from pelvic motion, and the stationary flexion/extension unit minimizes errors, enhancing measurement accuracy.
|
at baseline
|
|
assessment of lateral flexion range of motion using inclinometer
Časové okno: at baseline
|
For measuring lumbar lateral flexion, the subject's sides are parallel to a wall to prevent compensatory forward trunk flexion.
An inclinometer is secured over the sacral region at S1, using a belt for stability.
A gravity goniometer is placed on the upper thoracic region to capture trunk angular displacement during side bending.
The inclinometer is levelled and zeroed before movement, allowing for precise measurement of lateral lumbar movement while reducing thoracic and pelvic compensations.
Accurate placement of the device is essential for reliable results.
The subject slides their hand down the thigh, keeping weight on the opposite leg to ensure the measurement reflects true lateral flexion.measures]
|
at baseline
|
Sekundární výstupní opatření
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
assessment of flexion and extension range of motion using smartphone
Časové okno: at baseline
|
To measure lumbar flexion range of motion (ROM) using the iPhone Level app, participants are positioned standing upright with feet shoulder-width apart.
T12-L1 and S1-S2 spinal levels are identified through palpation and marked on the skin.
The iPhone is placed at these levels to record angles during maximum lumbar flexion.
Total lumbar flexion ROM is calculated by subtracting the angle at S1-S2 from T12-L1.
Subjects perform maximum lumbar flexion and extension with straight knees, and measurements are taken across three trials.
|
at baseline
|
|
assessment of lateral flexion range of motion using smartphone
Časové okno: at baseline
|
During the measurement of lumbar side bending using the iPhone Level application, the subject will stand in a neutral upright position with feet shoulder-width apart and hands placed on the hips to prevent arm movement from influencing the measurement.
The spinous processes of T12 and S2 are identified and marked on the skin.
The iPhone is first placed and zeroed at the T12 level, and the subject is instructed to bend sideways (toward one side) as far as possible without lifting the heels or rotating the trunk.
The device remains aligned at T12 to capture the angular displacement.
The same procedure is repeated with the device positioned at S2, and the final lumbar side bending range of motion is calculated by subtracting the angle at S2 (pelvic movement) from that at T12 (trunk movement).
This method isolates lumbar lateral flexion and has shown good reliability and validity in clinical settings
|
at baseline
|
Spolupracovníci a vyšetřovatelé
Sponzor
Sponzor
Spolupracovníci
Spolupracovníci
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Aktuální)
Začátek studia
Primární dokončení (Aktuální)
Primární dokončení
Dokončení studie (Aktuální)
Dokončení studie
Termíny zápisu do studia
První předloženo
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
První zveřejněno
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Poslední zveřejněná aktualizace
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Naposledy ověřeno
Více informací
Termíny související s touto studií
Další relevantní podmínky MeSH
Další identifikační čísla studie
Další identifikační čísla studie
- P.T.REC/012/006199
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .