Agreement Among Three Methods of Intraoperative Core Temperature Monitoring: Nasopharyngeal Probe, TaiDoc TD-1261 Infrared Tympanic Thermometer, and 3M SpotOn™ Skin Sensor (TEMP-AGREE)
A Prospective Observational Method-Comparison Study of the Agreement Among Nasopharyngeal Temperature Probe, TaiDoc TD-1261 Infrared Tympanic Thermometer, and 3M SpotOn™ Zero-Heat-Flux Skin Sensor for Intraoperative Core Temperature Monitoring in Adult Patients Undergoing Elective Surgery Under General Anesthesia
Keeping a patient's body temperature normal during surgery is an important part of safe anesthetic care. When the body temperature drops below 36°C (called inadvertent perioperative hypothermia), patients face higher risks of bleeding, wound infection, heart problems, and a slower recovery. To prevent these complications, anesthesiologists monitor core body temperature continuously during surgery.
Three monitoring methods are commonly used in clinical practice: (1) a nasopharyngeal probe - a small soft sensor inserted through the nose and placed at the back of the throat, which provides continuous readings and is considered the gold standard; (2) an infrared tympanic thermometer (TaiDoc TD-1261) - a handheld device that measures ear temperature in about one second; and (3) the 3M SpotOn™ sensor - a non-invasive adhesive patch placed on the forehead that continuously estimates core temperature without requiring insertion into any body cavity.
Although all three methods are used interchangeably at Srinagarind Hospital, no study has directly compared all three simultaneously in the same patient to determine how well they agree with each other. If significant differences exist between methods, a clinician who switches from one device to another may unknowingly under-detect or over-estimate hypothermia.
This study will enroll 92 adult patients scheduled for elective surgery under general anesthesia at Srinagarind Hospital, Khon Kaen University. All three temperature sensors will be applied at the same time, and readings will be recorded simultaneously every 15 minutes throughout the operation. The main goal is to determine whether each device gives results close enough to the nasopharyngeal probe to be safely used as an alternative. The study is observational - no medications or procedures will be added to routine care.
Přehled studie
Postavení
Postavení
Podmínky
Podmínky
Detailní popis
Background: Inadvertent perioperative hypothermia (IPH), defined as a core body temperature < 36.0°C, affects 20-90% of patients undergoing general anesthesia and is associated with coagulopathy, surgical site infection, morbid cardiac events, and prolonged drug metabolism. Accurate intraoperative temperature monitoring is mandated by professional society guidelines.
Study design: Prospective, single-center, triple-method observational agreement study at Srinagarind Hospital, Khon Kaen University, Thailand. All three thermometry modalities are applied simultaneously in each participant. No intervention is assigned; temperature monitoring is additional to standard perioperative care.
Methods: After induction of general anesthesia and tracheal intubation, the nasopharyngeal (NP) probe (reference standard), TaiDoc TD-1261 infrared tympanic thermometer (test method 1), and 3M SpotOn™ zero-heat-flux forehead sensor (test method 2) are placed. Simultaneous temperature readings are recorded at 15-minute intervals from 15 minutes after induction until surgery end. For the TD-1261, two consecutive readings are taken at each time point and averaged; pinna retraction and fresh disposable probe cover are used at every measurement. The SpotOn™ sensor requires a 10-minute equilibration period before the first recorded value.
Statistical analysis: Primary analysis uses repeated-measures Bland-Altman analysis (Myles & Cui, 2007 framework) to estimate bias and limits of agreement (LoA) for each method pair against the NP probe. Pre-specified clinical acceptability criterion: LoA within ±0.5°C (Sessler, 2008). Variance of differences is decomposed into between-subject (σ²_B) and within-subject (σ²_W) components. Secondary analyses: proportional bias assessment, sensitivity/specificity for hypothermia detection (NP < 36.0°C), intraclass correlation coefficient (ICC, two-way mixed-effects model), and multivariable linear mixed-effects regression for clinical covariates. Sample size: 92 patients (Lu et al., 2016 method; 80% power; α = 0.05; SD of differences = 0.30°C for NP vs. TM; LoA acceptability threshold ±0.50°C; 15% dropout allowance).
This study provides the first institution-specific, population-specific evidence on temperature monitoring agreement in Thai surgical patients and constitutes the first published intraoperative validation of the TaiDoc TD-1261 infrared tympanic thermometer.
Typ studie
Typ studie
Zápis (Odhadovaný)
Zápis
Kontakty a umístění
Studijní kontakt
Studijní kontakt
- Jméno: Thepakorn Sathitkarnmanee, MD
- Telefonní číslo: 66819547622
- E-mail: thepakorns@gmail.com
Studijní záloha kontaktů
- Jméno: Sirirat Tribuddharat, MD, PhD
- Telefonní číslo: 66816205920
- E-mail: sirirat.tribuddharat@gmail.com
Studijní místa
-
-
-
Khon Kaen, Thajsko, 40002
- Faculty of Medicine, Khon Kaen University
-
-
Kritéria účasti
Kritéria způsobilosti
Kritéria způsobilosti
Věk způsobilý ke studiu
- Dospělý
- Starší dospělý
Přijímá zdravé dobrovolníky
Metoda odběru vzorků
Studijní populace
Popis
Inclusion Criteria:
- Age ≥ 18 years
- We will include 3 independent groups of patients receiving general anesthesia with tracheal intubation:
Group 1: Cardiac surgery Group 2: Thoracic surgery Group 3: General surgery, including neurosurgery
- Expected surgical duration ≥ 60 minutes
- ASA Physical Status Classification I- III
- Able to provide written informed consent
Exclusion Criteria:
- Prior nasal surgery, nasal polyps, deviated nasal septum, or significant epistaxis history precluding safe nasopharyngeal probe insertion
- Otitis media, tympanic membrane perforation, significant cerumen impaction, or external auditory canal deformity that would prevent reliable infrared tympanic temperature measurement
- Facial or forehead skin lesions, burns, or known adhesive hypersensitivity at the planned 3M SpotOn™ application site, craniotomy, and head and neck surgery.
- Baseline core temperature < 35.0°C or > 38.5°C at preoperative assessment
- Planned use of neuraxial anesthesia as the primary or adjunct technique
- Emergency procedures, pregnancy, or procedures involving deliberate hypothermia or cardiopulmonary bypass.
Studijní plán
Jak je studie koncipována?
Detaily designu
Počet skupin / kohort
Kohorty a intervence
Skupina / kohortaSkupina / kohorta |
|---|
|
Surgical Patients - All Three Monitoring Methods
Adult patients (age ≥ 18 years) undergoing elective surgery under general anesthesia with tracheal intubation at Srinagarind Hospital and Queen Sirikit Heart Center of the Northeast, Khon Kaen University, with an expected surgical duration ≥ 60 minutes and ASA Physical Status I-III.
All participants have all three temperature monitoring devices (NP probe, TaiDoc TD-1261 IR tympanic thermometer, and 3M SpotOn™ ZHF sensor) applied simultaneously following induction of anesthesia.
|
Co je měření studie?
Primární výstupní opatření
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Limits of Agreement Between Each Test Method and the Nasopharyngeal Probe
Časové okno: From 15 minutes after anesthesia induction until end of surgery; measurements recorded at 15-minute intervals (estimated total observation window: 60-195 minutes per participant)
|
The 95% limits of agreement (LoA = Bias ± 1.96 × SD_total) between (i) the TaiDoc TD-1261 infrared tympanic thermometer and the nasopharyngeal probe, and (ii) the 3M SpotOn™ sensor and the nasopharyngeal probe, estimated by repeated-measures Bland-Altman analysis.
Pre-specified clinical acceptability criterion: LoA must be contained within ±0.5°C with both 95% CI limits of the upper and lower LoA falling within this range.
|
From 15 minutes after anesthesia induction until end of surgery; measurements recorded at 15-minute intervals (estimated total observation window: 60-195 minutes per participant)
|
Sekundární výstupní opatření
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Systematic Bias (Mean Difference) of Each Test Method vs. NP Probe
Časové okno: From 15 minutes after anesthesia induction until surgery end (15-minute intervals)
|
Mean difference (°C) with 95% CI between each test method and the NP probe across all paired observations, estimated from the repeated-measures Bland-Altman model
|
From 15 minutes after anesthesia induction until surgery end (15-minute intervals)
|
|
Proportional Bias Assessment
Časové okno: From 15 minutes after anesthesia induction until surgery end
|
Linear regression of the difference (test - reference) on the mean of both methods, with cluster-robust standard errors; Pitman's test for correlation between sum and difference
|
From 15 minutes after anesthesia induction until surgery end
|
|
Sensitivity and Specificity for Hypothermia Detection
Časové okno: From 15 minutes after anesthesia induction until surgery end
|
Sensitivity, specificity, PPV, and NPV of each test method for detecting intraoperative hypothermia defined as NP temperature < 36.0°C, with AUROC estimated by DeLong's method
|
From 15 minutes after anesthesia induction until surgery end
|
|
Intraclass Correlation Coefficient
Časové okno: From 15 minutes after anesthesia induction until surgery end
|
ICC (two-way mixed-effects model, absolute agreement) with 95% CI for each method pair
|
From 15 minutes after anesthesia induction until surgery end
|
|
Percentage of Paired Measurements Within ±0.5°C
Časové okno: From 15 minutes after anesthesia induction until surgery end
|
Percentage of all paired simultaneous readings in which the test method falls within ±0.5°C of the simultaneous NP reference reading
|
From 15 minutes after anesthesia induction until surgery end
|
|
Clinical Covariate Association With Method Difference
Časové okno: Baseline characteristics collected before surgery; temperature data from 15 minutes post-induction until surgery end
|
Association between clinical covariates (age, sex, BMI, ASA status, surgical duration, anesthetic technique, active warming use, ambient OR temperature, baseline core temperature) and the absolute method difference, estimated by multivariable linear mixed-effects regression
|
Baseline characteristics collected before surgery; temperature data from 15 minutes post-induction until surgery end
|
|
Incidence of Device-Related Adverse Events
Časové okno: From device placement until surgery end
|
Rate of epistaxis (NP probe), aural discomfort or canal injury (TD-1261), and adhesive skin reaction (SpotOn™) during the monitoring period
|
From device placement until surgery end
|
Spolupracovníci a vyšetřovatelé
Sponzor
Sponzor
Publikace a užitečné odkazy
Obecné publikace
- Sessler DI. Perioperative thermoregulation and heat balance. Lancet. 2016 Jun 25;387(10038):2655-2664. doi: 10.1016/S0140-6736(15)00981-2. Epub 2016 Jan 8.
- Torossian A. Thermal management during anaesthesia and thermoregulation standards for the prevention of inadvertent perioperative hypothermia. Best Pract Res Clin Anaesthesiol. 2008 Dec;22(4):659-68. doi: 10.1016/j.bpa.2008.07.006.
- Sessler DI. Temperature monitoring and perioperative thermoregulation. Anesthesiology. 2008 Aug;109(2):318-38. doi: 10.1097/ALN.0b013e31817f6d76.
- Lu MJ, Zhong WH, Liu YX, Miao HZ, Li YC, Ji MH. Sample Size for Assessing Agreement between Two Methods of Measurement by Bland-Altman Method. Int J Biostat. 2016 Nov 1;12(2):/j/ijb.2016.12.issue-2/ijb-2015-0039/ijb-2015-0039.xml. doi: 10.1515/ijb-2015-0039.
- Matsukawa T, Sessler DI, Sessler AM, Schroeder M, Ozaki M, Kurz A, Cheng C. Heat flow and distribution during induction of general anesthesia. Anesthesiology. 1995 Mar;82(3):662-73. doi: 10.1097/00000542-199503000-00008.
- Cork RC, Vaughan RW, Humphrey LS. Precision and accuracy of intraoperative temperature monitoring. Anesth Analg. 1983 Feb;62(2):211-4.
- Wang J, Liang H, Tian C, Rong G, Shao X, Ran C. Agreement of zero-heat-flux thermometry compared with infrared tympanic temperature monitoring in adults undergoing major surgery. Eur J Med Res. 2025 Jan 30;30(1):60. doi: 10.1186/s40001-025-02317-9.
- Iden T, Horn EP, Bein B, Bohm R, Beese J, Hocker J. Intraoperative temperature monitoring with zero heat flux technology (3M SpotOn sensor) in comparison with sublingual and nasopharyngeal temperature: An observational study. Eur J Anaesthesiol. 2015 Jun;32(6):387-91. doi: 10.1097/EJA.0000000000000232.
- Wan L, Shen PY, Zhang SX, Wang LZ. Agreement of infrared ear temperature with nasopharyngeal temperature and diagnostic performance on hypothermia in general anesthetized patients. J Chin Med Assoc. 2022 Nov 1;85(11):1093-1097. doi: 10.1097/JCMA.0000000000000770. Epub 2022 Jul 6.
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Odhadovaný)
Začátek studia
Primární dokončení (Odhadovaný)
Primární dokončení
Dokončení studie (Odhadovaný)
Dokončení studie
Termíny zápisu do studia
První předloženo
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
První zveřejněno
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Poslední zveřejněná aktualizace
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Naposledy ověřeno
Více informací
Termíny související s touto studií
Klíčová slova
Další identifikační čísla studie
Další identifikační čísla studie
- HE691250
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Popis plánu IPD
Časový rámec sdílení IPD
Kritéria přístupu pro sdílení IPD
Typ podpůrných informací pro sdílení IPD
- PROTOKOL STUDY
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .