Similar migration in computer-assisted and conventional total knee arthroplasty

Gunnar Petursson, Anne Marie Fenstad, Øystein Gøthesen, Kristin Haugan, Gro Sævik Dyrhovden, Geir Hallan, Stephan M Röhrl, Arild Aamodt, Kjell G Nilsson, Ove Furnes, Gunnar Petursson, Anne Marie Fenstad, Øystein Gøthesen, Kristin Haugan, Gro Sævik Dyrhovden, Geir Hallan, Stephan M Röhrl, Arild Aamodt, Kjell G Nilsson, Ove Furnes

Abstract

Background and purpose - Computer-assisted surgery (CAS) in total knee arthroplasty (TKA) has been used in recent years in the hope of improving the alignment and positioning of the implant, thereby achieving a better functional outcome and durability. However, the role of computer navigation in TKA is still under debate. We used radiostereometric analysis (RSA) in a randomized controlled trial (RCT) to determine whether there are any differences in migration of the tibial component between CAS- and conventionally (CONV-) operated TKA. Patients and methods - 54 patients (CAS, n = 26; CONV, n = 28) with a mean age of 67 (56-78) years and with osteoarthritis or arthritic disease of the knee were recruited from 4 hospitals during the period 2009-2011. To estimate the mechanical stability of the tibial component, the patients were examined with RSA up to 24 months after operation. The following parameters representing tibial component micromotion were measured: 3-D vector of the prosthetic marker that moved the most, representing the magnitude of migration (maximum total point motion, MTPM); the largest negative value for y-translation (subsidence); the largest positive y-translation (lift-off); and prosthetic rotations. The precision of the RSA measurements was evaluated and migration in the 2 groups was compared. Results - Both groups had most migration within the first 3 months, but there was no statistically significant difference in the magnitude of the migration between the CAS group and the CONV group. From 3 to 24 months, the MTPM (in mm) was 0.058 and 0.103 (p = 0.1) for the CAS and CON groups, respectively, and the subsidence (in mm) was 0.005 and 0.011 (p = 0.3). Interpretation - Mean MTPM, subsidence, lift-off, and rotational movement of tibial trays were similar in CAS- and CONV-operated knees.

Figures

Figure 1.
Figure 1.
Distribution of the 0.8-mm and 1.0-mm tantalum markers in the polyethylene component of the prosthesis.
Figure 2.
Figure 2.
Flow chart of the patients.
Figure 3.
Figure 3.
Maximum total point motion.
Figure 4.
Figure 4.
Subsidence, i.e. largest negative value for y-axis translation.
Figure 5.
Figure 5.
Lift-off, i.e. largest positive value for y-axis translation.

References

    1. Australian Orthopaedic Association National Joint Replacement Registry Annual Report. Adelaide: AOA; 2015.
    1. Bauwens K, Matthes G, Wich M, Gebhard F, Hanson B, Ekkernkamp A, Stengel D.. Navigated total knee replacement. A meta-analysis. J Bone Joint Surg (Am) 2007; 89 (2):261–9.
    1. Bellemans J. Neutral mechanical alignment: a requirement for successful TKA: opposes. Orthopedics 2011; 34: 507–9.
    1. Blakeney W G, Khan R J K, Wall S J.. Computer-assistet technique versus conventional guides for component alignment in total knee arthroplasty. A randomized controlled trial. J Bone Joint Surg (Am) 2011; 93: 1377–84.
    1. Blakeney W G, Khan R J, Palmer J L.. Functional outcomes following total knee arthroplasty: a randomised trial comparing computer-assisted surgery with conventional techniques. Knee 2014;21(2): 364–8.
    1. Campbell M J, Gardner M J.. Calculating confidence intervals for some non-parametric analyses. Br Med J (Clin Res Ed) 1988; 296(6634): 1454–6.
    1. Chauhan S K, Scott R G, Breidahl W, Beaver RJ .. Computer-assisted knee arthroplasty versus a conventional jig-based technique. A randomised, prospective trial. J Bone Joint Surg (Br) 2004; 86-B: 372–7.
    1. Cheng T, Shao S, Peng X, Shang X.. Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 2012; 20: 1307–22.
    1. Choong P F, Dowsey M M, Stoney J D.. Does Accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. J Arthroplasty 2009; 24: 560–9.
    1. Dennis DA, Komistek RD, Kim RH, Sharma A.. Gap balancing versus measured resection technique for total knee arthroplasty. Clin Orthop Relat Res 2010; 468: 102–7.
    1. Dyrhovden G S, Gøthesen Ø, Lygre S H L, Fenstad A M, Sørås T E, Halvorsen S, Jellestad T, Furnes O.. Is the use of computer navigation in total knee arthroplasty improving implant positioning and function? A comparative study of 198 knees operated at a Norwegian district hospital. BMC Musculoskelet Disord 2013; 14: 321.
    1. Fehring T K, Odum S, Friffin W L, Mason J B, Nadaud M.. Early failures in total knee arthroplasty. Clin Orthop Relat Res 2001; 392: 315–8.
    1. Gøthesen Ø, Espehaug B, Havelin L, Petursson G, Furnes O.. Short-term outcome of 1,465 computer-navigated primary total knee replacements 2005-2008. A report from the Norwegian Arthroplasty Register. Acta Orthop 2011; 82(3): 293–300.
    1. Gøthesen Ø, Espehaug B, Havelin LI, Petursson G, Hallan G, Strøm E, Dyrhovden GS, Furnes O.. Functional Outcome and alignment in computer assisted and conventionally operated total knee replacements. A multicenter parallel group randomised controlled trial. Bone Joint J 2014; 96-B: 609–18.
    1. Henricson A, Linder L, Nilsson K G.. A trabecular metal tibial component in total knee replacement in patients younger than 60 years. A two-year radiostereophotogrammetric analysis. J Bone Joint Surg (Br) 2008; 90: 1585–93.
    1. Henricson A, Rösmark D, Nilsson K G.. Trabecular metal tibia still stable at 5 years. An RSA study of 36 patients aged less than 60 years. Acta Orthop 2013; 84(4): 398–405.
    1. Kärrholm J, Borssen B, Lowenhielm G, Snorrason F.. Does early micromotion of femoral stem prostheses matter? 4-7-years stereoradiographic follow-up of 84 cemented prostheses. J Bone Joint Surg (Br) 1994; 76:912–7.
    1. Kärrholm J, Gill R H S, Valstar E R.. The History and Future of Radiosterometric Analysis. Clin Orthop Rel Res 2006: 448: 10–21.
    1. Kim Y H, Kim J S, Choi Y, Kwon O R.. Computer-assisted surgical navigation does not improve the alignment and orientation of the components in total knee arthroplasty. J Bone Joint Surg (Am) 2009; 91(1): 14–9.
    1. Lad D G, Thilak J, Thadi M.. Component alignment and functional outcome following computer assisted and jig based total knee arthroplasty. Indian J Orthop 2013; 47(1): 77–82.
    1. Longstaff L M, Sloan K, Stamp N, Scaddan M, Beaver R.. Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty 2009; 24(4): 570–8.
    1. Lüring C, Bäthis H, Tingart M, Perlick L, Grifka J.. Computer assistance in total knee replacement – a critical assessment of current health care technology. Comput Aided Surg 2006; 11: 77–80.
    1. Lützner J, Dexel J, Kirschner S.. No difference between computer-assisted and conventional total knee arthroplasty: five-year results of a prospective randomised study. Knee Surg Sports Traumatol Arthrosc 2013; 21(10): 2231–7.
    1. Nilsson K G, Kärrholm J, Ekelund L, Magnusson P.. Evaluation of micromotion in cemented vs uncemented knee arthroplasty in osteoarthrosis and rheumatoid arthritis. J Arthroplasty 1991; 6: 265–78.
    1. Norwegian Arthroplasty Register Norwegian National Advisory Unit on Arthroplasty and Hip Fractures. Helse Bergen HF, Department of Orthopaedic Surgery, Haukeland University Hospital 2015: .
    1. Parratte S, Pagnano M W, Trousdale R T, Berry D J.. Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern cemented total knee replacements. J Bone Joint Surg (Am) 2010; 92-A: 2143–9.
    1. Pijls B G, Valstar E R, Nouta K A, Plevier J W M, Fiocco M, Middeldorp S, Nelisson R G G H.. Early migration of tibial components is associated with late revision. A systematic review and meta-analysis of 21,000 knee arthroplasties. Acta Orthop 2012; 83(6): 614–624.
    1. Ritter M A, Faris P M, Keating E M, Meding J B.. Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res 1994; 299: 153–6.
    1. Ryd L, Albrektsson B E J, Carlsson L, Dansgård F, Herberts P, Lindstrand A, Regnèr L, Toksvig-Larsen S.. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses. J Bone Joint Surg (Br) 1995; 77-B: 377–83.
    1. Schulz K F, Altman D G, Moher D, CONSORT Group . CONSORT 2010 statement: updated guidelines for reporting parallel group randomised trials. BMJ 2010; 340: 332.
    1. Shi J, Wei Y, Wang S, Chen F, Wu J, Huang G, Chen J, Wei L, Xia J.. Computer navigation and total knee arthroplasty. Orthopedics 2014; 37(1): e39–43.
    1. Spencer J M, Chauhan S K, Sloan K, Taylor A, Beaver R J.. Computer navigation versus conventional total knee replacement: no difference in functional results at two years. J Bone Joint Surg Br: 2007; 89(4): 477–80.
    1. de Steiger R N, Liu Y L, Graves S E.. Computer navigation for total knee arthroplasty reduces revision rate for patients less than Sixty-five years of age. J Bone Joint Surg (Am) 2015; 97: 635–42.
    1. Tjørnild M, Søballe K, Hansen P M, Holm C, Stilling M.. Mobile- vs. fixed-bearing total knee replacement. A randomized radiostereometric and bone mineral density study. Acta Orthop 2015; 86(2): 208–14.
    1. Valstar E R, Gill R, Ryd L, Flivik G, Borlin N, Karrholm J.. Guidelines for standardization of radiosterometry (RSAI of implants. Acta Orthop 2005; 76: 563–72.
    1. Valstar E R, Kaptein B, Nelissen R. Guest editorial . Radiosterometry and new prostheses. Acta Orthop 2012; 83(2): 103–4.
    1. van Strien T, van der Linden-van der Zwaag E, Kaptein B, van Erkel A, Valstar E, Nelissen R.. Computer assisted versus conventional cemented total knee prostheses alignment accuracy and micromotion of the tibial component. Int Orthop 2009; 33(5): 1255–61.
    1. Whiteside L A. Soft tissue balancing the knee. J Arthroplasty 2002; 17 (4 Suppl1): 23–27.
    1. Whiteside L A, Arima J.. The anteroposterior axis for femoral rotational alignment in valgus total knee arthroplasty. Clin Orthop Relat Res 1995; 321: 168–72.