Use of wearable devices for post-discharge monitoring of ICU patients: a feasibility study

Ryan R Kroll, Erica D McKenzie, J Gordon Boyd, Prameet Sheth, Daniel Howes, Michael Wood, David M Maslove, WEARable Information Technology for hospital INpatients (WEARIT-IN) study group, Ryan R Kroll, Erica D McKenzie, J Gordon Boyd, Prameet Sheth, Daniel Howes, Michael Wood, David M Maslove, WEARable Information Technology for hospital INpatients (WEARIT-IN) study group

Abstract

Background: Wearable devices generate signals detecting activity, sleep, and heart rate, all of which could enable detailed and near-continuous characterization of recovery following critical illness.

Methods: To determine the feasibility of using a wrist-worn personal fitness tracker among patients recovering from critical illness, we conducted a prospective observational study of a convenience sample of 50 stable ICU patients. We assessed device wearability, the extent of data capture, sensitivity and specificity for detecting heart rate excursions, and correlations with questionnaire-derived sleep quality measures.

Results: Wearable devices were worn over a 24-h period, with excellent capture of data. While specificity for the detection of tachycardia was high (98.8%), sensitivity was low to moderate (69.5%). There was a moderate correlation between wearable-derived sleep duration and questionnaire-derived sleep quality (r = 0.33, P = 0.03). Devices were well-tolerated and demonstrated no degradation in quality of data acquisition over time.

Conclusions: We found that wearable devices could be worn by patients recovering from critical illness and could generate useful data for the majority of patients with little adverse effect. Further development and study are needed to better define and enhance the role of wearables in the monitoring of post-ICU recovery.

Trial registration: Clinicaltrials.gov, NCT02527408.

Keywords: Critical care; Heart rate monitoring; Medical informatics; Mobile health technologies; Sleep quality; Validation study; Wearable devices.

Conflict of interest statement

Ethics approval and consent to participate

All participating patients, or substitute decision makers on their behalf, provided written informed consent for participation in this study. The Health Sciences Research Ethics Board at Queen’s University reviewed and approved the study protocol, and the trial was registered with Consent for publication

The patient depicted in Fig. 1 provided consent for the photo to be published in this manuscript.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Figures

Fig. 1
Fig. 1
The Fitbit Charge HR device used in the study (a). The wearable device as worn by a patient on the inpatient ward following ICU discharge (b)
Fig. 2
Fig. 2
Accuracy of wearable-derived heart rates for the detection of tachycardia (HR > 100) or bradycardia (HR

Fig. 3

Correlation between mean score on…

Fig. 3

Correlation between mean score on the Richards-Campbell Sleep Questionnaire (RCSQ) and wearable-derived measure…

Fig. 3
Correlation between mean score on the Richards-Campbell Sleep Questionnaire (RCSQ) and wearable-derived measure of the number of minutes asleep overnight (between 22:00 and 06:00). The Pearson correlation coefficient was 0.33 (95% CI 0.04 - 0.58)
Fig. 3
Fig. 3
Correlation between mean score on the Richards-Campbell Sleep Questionnaire (RCSQ) and wearable-derived measure of the number of minutes asleep overnight (between 22:00 and 06:00). The Pearson correlation coefficient was 0.33 (95% CI 0.04 - 0.58)

References

    1. Salah H, MacIntosh E, Rajakulendran N. Wearable tech: leveraging canadian innovation to improve health: MaRS Market Insights; 2014 Mar 26. p. 1–45. Available from
    1. Savage N. Mobile data: made to measure. Nature. 2015;527:S12–S13. doi: 10.1038/527S12a.
    1. El-Amrawy F, Nounou MI. Are currently available wearable devices for activity tracking and heart rate monitoring accurate, precise, and medically beneficial? Healthc Inform Res. 2015;21(4):315–316. doi: 10.4258/hir.2015.21.4.315.
    1. Case MA, Burwick HA, Volpp KG, Patel MS. Accuracy of smartphone applications and wearable devices for tracking physical activity data. JAMA-J Am Med Assoc. 2015;313(6):625–626. doi: 10.1001/jama.2014.17841.
    1. Wang R, Blackburn G, Desai M, Phelan D, Gillinov L, Houghtaling P, et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2017;2(1):104–106. doi: 10.1001/jamacardio.2016.3340.
    1. Asada HH, Shaltis P, Reisner A, Rhee S. Mobile monitoring with wearable photoplethysmographic biosensors. IEEE Eng Med Biol Mag. 2003;22(3):28–40. doi: 10.1109/MEMB.2003.1213624.
    1. Tamura T, Maeda Y, Sekine M, Yoshida M. Wearable photoplethysmographic sensors—past and present. Electronics. 2014;3(2):282–302. doi: 10.3390/electronics3020282.
    1. Pantelopoulos A, Bourbakis NG. A survey on wearable sensor-based systems for health monitoring and prognosis. IEEE Trans Syst, Man, Cybern C. 2009;40(1):1–12. doi: 10.1109/TSMCC.2009.2032660.
    1. Di Rienzo M, Rizzo F, Parati G, Brambilla G, Ferratini M, Castiglioni P. MagIC system: a new textile-based wearable device for biological signal monitoring. Applicability in daily life and clinical setting. In: Engineering in medicine and biology society, 2005. IEEE-EMBS 2005. 27th annual international conference of the 2005 (pp. 7167–7169). Shanghai: IEEE.
    1. de Zambotti M, Baker FC, Willoughby AR, Godino JG, Wing D, Patrick K, et al. Measures of sleep and cardiac functioning during sleep using a multi-sensory commercially-available wristband in adolescents. Physiol Behav. 2016;158:143–149. doi: 10.1016/j.physbeh.2016.03.006.
    1. Mantua J, Gravel N, Spencer R. Reliability of sleep measures from four personal health monitoring devices compared to research-based actigraphy and polysomnography. Sensors. 2016;16(5):646. doi: 10.3390/s16050646.
    1. Young JS, Bourgeois JA, Hilty DM, Hardin KA. Sleep in hospitalized medical patients, part 1: factors affecting sleep. J Hosp Med. 2008;3(6):473–482. doi: 10.1002/jhm.372.
    1. Kamdar BB, Needham DM, Collop NA. Sleep deprivation in critical illness: its role in physical and psychological recovery. J Intensive Care Med. 2012;27(2):97–111. doi: 10.1177/0885066610394322.
    1. Litton E, Carnegie V, Elliott R, Webb SAR. The efficacy of earplugs as a sleep hygiene strategy for reducing delirium in the ICU: a systematic review and meta-analysis. Crit Care Med. 2016;44(5):992–999. doi: 10.1097/CCM.0000000000001557.
    1. Richards KC, O’Sullivan PS, Phillips RL. Measurement of sleep in critically ill patients. J Nurs Meas. 2000;8(2):131–144.
    1. Kamdar BB, Shah PA, King LM, Kho ME, Zhou X, Colantuoni E, et al. Patient-nurse interrater reliability and agreement of the Richards-Campbell Sleep Questionnaire. Am J Crit Care. 2012;21(4):261–269. doi: 10.4037/ajcc2012111.
    1. Flannery AH, Oyler DR, Weinhouse GL. The impact of interventions to improve sleep on delirium in the ICU: a systematic review and research framework. Crit Care Med. 2016;44(12):2231–2240. doi: 10.1097/CCM.0000000000001952.
    1. Kroll RR, Boyd JG, Maslove DM. Accuracy of a wrist-worn wearable device for monitoring heart rates in hospital inpatients: a prospective observational study. J Med Internet Res. 2016;18(9):e253. doi: 10.2196/jmir.6025.
    1. Herridge MS, Chu LM, Matte A, Tomlinson G, Chan L, Thomas C, et al. The RECOVER program: disability risk groups and 1-year outcome after 7 or more days of mechanical ventilation. Am J Crit Care Med. 2016;194(7):831–844. doi: 10.1164/rccm.201512-2343OC.
    1. McNelly AS, Rawal J, Shrikrishna D, Hopkinson NS, Moxham J, Harridge SD, et al. An exploratory study of long-term outcome measures in critical illness survivors: construct validity of physical activity, frailty, and health-related quality of life measures. Crit Care Med. 2016;44(6):e362–e369. doi: 10.1097/CCM.0000000000001645.
    1. Roney JK, Whitley BE, Maples JC, Futrell LS, Stunkard KA, Long JD. Modified early warning scoring (MEWS): evaluating the evidence for tool inclusion of sepsis screening criteria and impact on mortality and failure to rescue. J Clin Nurs. 2015;24(23–24):3343–3354. doi: 10.1111/jocn.12952.
    1. Semple JL, Armstrong KA. Mobile applications for postoperative monitoring after discharge. Can Med Assoc J. 2016. electronically published ahead of print; 10.1503/cmaj.160195.
    1. Jakicic JM, Davis KK, Rogers RJ, King WC, Marcus MD, Helsel D, et al. Effect of wearable technology combined with a lifestyle intervention on long-term weight loss: the IDEA randomized clinical trial. JAMA. 2016;316(11):1161–1171. doi: 10.1001/jama.2016.12858.
    1. Smith MEB, Chiovaro JC, O'Neil M, Kansagara D, Quiñones AR, Freeman M, et al. Early warning system scores for clinical deterioration in hospitalized patients: a systematic review. Ann Am Thorac Soc. 2014;11(9):1454–1465. doi: 10.1513/AnnalsATS.201403-102OC.
    1. Churpek MM, Yuen TC, Park SY, Meltzer DO, Hall JB, Edelson DP. Derivation of a cardiac arrest prediction model using ward vital signs. Crit Care Med. 2012;40(7):2102–2108. doi: 10.1097/CCM.0b013e318250aa5a.
    1. Churpek MM, Yuen TC, Winslow C, Meltzer DO, Kattan MW, Edelson DP. Multicenter comparison of machine learning methods and conventional regression for predicting clinical deterioration on the wards. Crit Care Med. 2016;44(2):368–374. doi: 10.1097/CCM.0000000000001571.
    1. Churpek MM, Yuen TC, Park SY, Gibbons R, Edelson DP. Using electronic health record data to develop and validate a prediction model for adverse outcomes in the wards. Crit Care Med. 2014;42(4):841. doi: 10.1097/CCM.0000000000000038.
    1. Churpek MM, Yuen TC, Winslow C, Hall J, Edelson DP. Differences in vital signs between elderly and nonelderly patients prior to ward cardiac arrest. Crit Care Med. 2015;43(4):816–822. doi: 10.1097/CCM.0000000000000818.
    1. Bano M, Chiaromanni F, Corrias M, Turco M, De Rui M, Amodio P, et al. The influence of environmental factors on sleep quality in hospitalized medical patients. Front Neurol. 2014;5(4):267.
    1. Fitbit Inc. Fitbit help: how do I track my sleep? [internet]. . [accessed 2016 Oct 15]. Available from:
    1. Elliott R, McKinley S, Cistulli P, Fien M. Characterisation of sleep in intensive care using 24-hour polysomnography: an observational study. Crit Care. 2013;17(2):R46. doi: 10.1186/cc12565.

Source: PubMed

3