High-pitch coronary CT angiography at 70 kVp with low contrast medium volume: comparison of 80 and 100 kVp high-pitch protocols

Long Jiang Zhang, Li Qi, Carlo N De Cecco, Chang Sheng Zhou, James V Spearman, U Joseph Schoepf, Guang Ming Lu, Long Jiang Zhang, Li Qi, Carlo N De Cecco, Chang Sheng Zhou, James V Spearman, U Joseph Schoepf, Guang Ming Lu

Abstract

The purpose of this article is to evaluate image quality and radiation dose of prospectively electrocardiogram (ECG)-triggered high-pitch coronary computed tomography angiography (CCTA) at 70 kVp and 30 mL contrast medium.One hundred fifty patients with a heart rate ≤70 beats per minute (bpm) underwent CCTA using a second-generation dual-source computed tomography (CT) scanner and were randomized into 3 groups according to tube voltage and contrast medium volume (370 mg/mL iodine concentration) (100 kVp group, 100 kVp/60 mL, n = 55; 80 kVp group, 80 kVp/60 mL, n = 44; 70 kVp group, 70 kVp/30 mL, n = 51). Objective and subjective image quality along with the effect of heart rate (HR) and body mass index (BMI) was evaluated and compared between the groups. Radiation dose was estimated for each patient.CT attenuation and image noise were higher in the 80 and 70 kVp groups than in the 100 kVp group (all P < 0.001). Signal-to-noise ratios (SNRs) and contrast-to-noise ratios (CNRs) were lower in the 70 kVp group than in the 80 and 100 kVp groups (all P < 0.05). There was no difference for subjective image quality between the groups (P > 0.05). HR did not affect subjective image quality (all P > 0.05), while patients with BMI <23 kg/m had higher image quality than patients with BMI ≥23 kg/m (P < 0.05). Compared with the 100 kVp group, the radiation dose of the 70 kVp group was reduced by 75%.In conclusion, prospectively ECG-triggered high-pitch 70 kVp/30 mL CCTA can obtain diagnostic image quality with lower radiation dose in selected patients with BMI <23 kg/m compared with 80/100 kVp/60 mL CCTA.

Conflict of interest statement

The authors have no funding and conflicts of interest to disclose.

Figures

FIGURE 1
FIGURE 1
Image examples illustrating the four-point rating scale for image quality. Shown are automatically generated curved multiplanar reformations along the vessel center line of coronary arteries studied with CCTA. Panel A shows a 100 kVp CCTA study which was given a score of 1 because of marked motion artifact in segment 2. Panel B shows a 100 kVp CCTA study which was given a score of 2 because of mild motion artifact in segment 2. Panel C shows a 70 kVp CCTA study which was given a score of 3 because of a mild irregular contour of segment 2. Panel D shows an 80 kVp CCTA study which was given a score of 4.
FIGURE 2
FIGURE 2
Representative CCTA images in three different patients investigated with 70 kVp, 80 kVp, and 100 kVp. All CCTA images were assigned a score of 4 by the two readers. Note severe stenosis (arrows) in the proximal segment of the LAD of the subject in the 100 kVp group. VR = volume rendering; RCA = right coronary artery; LAD = left descending coronary artery; LCX = left circumflex coronary artery.

References

    1. Miller JM, Rochitte CE, Dewey M, et al. Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med. 2008;359:2324–2336.
    1. Meijboom WB, Meijs MF, Schuijf JD, et al. Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study. J Am Coll Cardiol. 2008;52:2135–2144.
    1. Litt HI, Gatsonis C, Snyder B, et al. CT angiography for safe discharge of patients with possible acute coronary syndromes. N Engl J Med. 2012;366:1393–1403.
    1. Hoffmann U, Truong QA, Schoenfeld DA, et al. Coronary CT angiography versus standard evaluation in acute chest pain. N Engl J Med. 2012;367:299–308.
    1. Brenner DJ, Hall EJ. Computed tomography–an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–2284.
    1. Cademartiri F, Maffei E, Arcadi T, et al. CT coronary angiography at an ultra-low radiation dose (<0.1 mSv): feasible and viable in times of constraint on healthcare costs. Eur Radiol. 2013;23:607–613.
    1. Flohr TG, Schoepf UJ, Ohnesorge BM. Chasing the heart: new developments for cardiac CT. J Thorax Imag. 2007;22:4–16.
    1. Yin WH, Lu B, Hou ZH, et al. Detection of coronary artery stenosis with sub-milliSievert radiation dose by prospectively ECG-triggered high-pitch spiral CT angiography and iterative reconstruction. Eur Radiol. 2013;23:2927–2933.
    1. Halliburton SS, Abbara S, Chen MY, et al. SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT. J Cardiovasc Comput Tomogr. 2011;5:198–224.
    1. Cao JX, Wang YM, Lu JG, et al. Radiation and contrast agent doses reductions by using 80-kV tube voltage in coronary computed tomographic angiography: a comparative study. Eur J Radiol. 2014;83:309–314.
    1. Jun BR, Yong HS, Kang EY, et al. 64-slice coronary computed tomography angiography using low tube voltage of 80 kV in subjects with normal body mass indices: comparative study using 120 kV. Acta Radiol. 2012;53:1099–1106.
    1. Schuhbaeck A, Achenbach S, Layritz C, et al. Image quality of ultra-low radiation exposure coronary CT angiography with an effective dose <0.1 mSv using high-pitch spiral acquisition and raw data-based iterative reconstruction. Eur Radiol. 2013;23:597–606.
    1. Neefjes LA, Dharampal AS, Rossi A, et al. Image quality and radiation exposure using different low-dose scan protocols in dual-source CT coronary angiography: randomized study. Radiology. 2011;261:779–786.
    1. Gnannt R, Winklehner A, Goetti R, et al. Low kilovoltage CT of the neck with 70 kVp: comparison with a standard protocol. AJNR Am J Neuroradiol. 2012;33:1014–1019.
    1. Duan Y, Wang X, Yang X, et al. Diagnostic efficiency of low-dose CT angiography compared with conventional angiography in peripheral arterial occlusions. AJR Am J Roentgenol. 2013;201:W906–W914.
    1. Zhang LJ, Qi L, Wang J, et al. Feasibility of prospectively ECG-triggered high-pitch coronary CT angiography with 30 mL iodinated contrast agent at 70 kVp: initial experience. Eur Radiol. 2014;24:1537–1546.
    1. Meinel FG, Canstein C, Schoepf UJ, et al. Image quality and radiation dose of low tube voltage 3(rd) generation dual-source coronary CT angiography in obese patients: a phantom study. Eur Radiol. 2014;24:1643–1650.
    1. Meyer M, Haubenreisser H, Schoepf UJ, et al. Closing in on the K Edge: coronary CT angiography at 100, 80, and 70 kV-Initial comparison of a second- versus a third-generation dual-source CT system. Radiology. 2014;273:373–382.
    1. Morsbach F, Desbiolles L, Plass A, et al. Stenosis quantification in coronary CT angiography: impact of an integrated circuit detector with iterative reconstruction. Invest Radiol. 2013;48:32–40.
    1. Austen WG, Edwards JE, Frye RL, et al. A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation. 1975;51:5–40.
    1. Chen MY, Shanbhag SM, Arai AE. Submillisievert median radiation dose for coronary angiography with a second-generation 320-detector row CT scanner in 107 consecutive patients. Radiology. 2013;267:76–85.
    1. Bongartz G, Golding SJ, Jurik AG, et al. European guidelines for multislice computed tomography: Appendix C Funded by the European Commission. Contract number FIGM-CT2000-20078-CT-TIP. Accessed September 9, 2013.
    1. AAPM Task Group. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations: report of AAPM Task Group 204. Accessed September 9, 2013.
    1. Christner JA, Braun NN, Jacobsen MC, et al. Size-specific dose estimates for adult patients at CT of the torso. Radiology. 2012;265:841–847.
    1. Komatsu S, Kamata T, Imai A, et al. Coronary computed tomography angiography using ultra-low-dose contrast media: radiation dose and image quality. Int J Cardiovasc Imag. 2013;29:1335–1340.
    1. Abbara S, Arbab-Zadeh A, Callister TQ, et al. SCCT guidelines for performance of coronary computed tomographic angiography: a report of the Society of Cardiovascular Computed Tomography Guidelines Committee. J Cardiovasc Comput Tomogr. 2009;3:190–204.
    1. Fei X, Du X, Yang Q, et al. 64-MDCT coronary angiography: phantom study of effects of vascular attenuation on detection of coronary stenosis. AJR Am J Roentgenol. 2008;191:43–49.
    1. Neefjes LA, Rossi A, Genders TS, et al. Diagnostic accuracy of 128-slice dual-source CT coronary angiography: a randomized comparison of different acquisition protocols. Eur Radiol. 2013;23:614–622.
    1. Stolzmann P, Goetti RP, Maurovich-Horvat P, et al. Predictors of image quality in high-pitch coronary CT angiography. AJR Am J Roentgenol. 2011;197:851–858.
    1. Achenbach S, Marwan M, Ropers D, et al. Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J. 2010;31:340–346.
    1. Alkadhi H, Stolzmann P, Desbiolles L, et al. Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode. Heart. 2010;96:933–938.
    1. Wang Y, Vidan E, Bergman GW. Cardiac motion of coronary arteries: variability in the rest period and implications for coronary MR angiography. Radiology. 1999;213:751–758.
    1. Husmann L, Leschka S, Desbiolles L, et al. Coronary artery motion and cardiac phases: dependency on heart rate—implications for CT image reconstruction. Radiology. 2007;245:567–576.
    1. Lee SM, Lee W, Chung JW, et al. Effect of kVp on image quality and accuracy in coronary CT angiography according to patient body size: a phantom study. Int J Cardiovasc Imag. 2013;29suppl 2:83–91.

Source: PubMed

3
Předplatit