Tattoo free setup for partial breast irradiation: A feasibility study

Rachel B Jimenez, Estelle Batin, Drosoula Giantsoudi, William Hazeltine, Kendell Bertolino, Alice Y Ho, Shannon M MacDonald, Alphonse G Taghian, David P Gierga, Rachel B Jimenez, Estelle Batin, Drosoula Giantsoudi, William Hazeltine, Kendell Bertolino, Alice Y Ho, Shannon M MacDonald, Alphonse G Taghian, David P Gierga

Abstract

Purpose: Patients undergoing external beam accelerated partial breast irradiation (APBI) receive permanent tattoos to aid with daily setup alignment and verification. With the advent of three-dimensional (3D) body surface imaging and two-dimensional (2D) x-ray imaging-based matching to surgical clips, tattoos may not be necessary to ensure setup accuracy. We compared the accuracy of conventional tattoo-based setups to a patient setup without tattoos.

Materials/methods: Twenty consecutive patients receiving APBI at our institution from July 10, 2017 to February 13, 2018 were identified. All patients received tattoos per standard of care. Ten patients underwent setup using tattoos for initial positioning followed by surface imaging and 2D matching of surgical clips. The other ten patients underwent positioning using surface imaging followed by 2D matching without reference to tattoos. Overall setup time and orthogonal x-ray-based shifts after surface imaging per fraction were recorded. Shift data were used to calculate systematic and random error.

Results: Among ten patients in the "no tattoo" group, the average setup time per fraction was 6.83 min vs 8.03 min in the tattoo cohort (P < 0.01). Mean 3D vector shifts for patients in the "no tattoo" group were 4.6 vs 5.9 mm in the "tattoo" cohort (P = NS). Mean systematic errors in the "no tattoo" group were: 1.2 mm (1.5 mm SD) superior/inferior, 0.5 mm (1.6 mm SD) right/left, and 2.3 mm (1.9 mm SD) anterior/posterior directions. Mean systematic errors in the "tattoo" group were: 0.8 mm (2.2 mm SD) superior/inferior, 0.3 mm (2.5 mm SD) right/left, and 1.4 mm (4.4 mm SD) anterior/posterior directions. The random errors in the "no tattoo" group ranged from 0.6 to 0.7 mm vs 1.2 to 1.7 mm in the "tattoo" group.

Conclusions: Using both surface imaging and 2D matching to surgical clips provides excellent accuracy in APBI patient alignment and setup verification with reduced setup time relative to the tattoo cohort. Skin-based tattoos may no longer be warranted for patients receiving external beam APBI.

Keywords: APBI; breast radiotherapy; surface imaging; tattoos.

Conflict of interest statement

Dr. Gierga has previously received speaking honoraria for VisionRT. The remaining authors have no conflict of interest to disclose.

© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Figures

Figure 1
Figure 1
Institutional tattoo setup per current standards of care.
Figure 2
Figure 2
(a) Region of interest (ROI) using AlignRT for APBI with (b) associated example daily orthogonal x ray, and (c) digitally reconstructed radiograph.
Figure 3
Figure 3
Initial setup errors detected by surface imaging.
Figure 4
Figure 4
Setup errors detected by x rays following surface imaging.

References

    1. Fobair P, Stewart SL, Chang S, et al. Body image and sexual problems in young women with breast cancer. Psychooncology. 2006;15:579–594.
    1. Clow B, Allen J. Psychosocial impacts of radiation tattooing for breast cancer patients: a critical review. Canadian Woman Studies. 2010;28:46–52.
    1. Fingeret MC, Teo I, Epner DE. Managing body image difficulties of adult cancer patients: lessons from available research. Cancer. 2014;120:633–641.
    1. Sewak S, Graham P, Nankervis J. Tattoo allergy in patients receiving adjuvant radiotherapy for breast cancer. Austr Radiol. 1999;43:558–561.
    1. Bregnhol A, Haedersal M. Q‐switched YAG laser vs. punch biopsy excision for iatrogenic radiation tattoo markers – a randomized controlled trial. J Eur Acam Dermatol Venereol. 2010;24:1183–1186.
    1. Landeg SJ, Kirby AM, Lee SF, et al. A randomized control trial evaluating fluorescent ink versus dark ink tattoos for breast radiotherapy. Br J Radiol. 2016;89:20160288.
    1. Gierga DP, Riboldi M, Turcotte JC, et al. Comparison of target registration errors for multiple image‐guided techniques in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2008;70:1239–1246.
    1. Oh KS, Kong FM, Griffith KA, et al. Planning the breast tumor bed boost: changes in the excision cavity volume and surgical scar location after breast‐conserving surgery and whole breast irradiation. Int J Radiat Oncol Biol Phys. 2006;66:680–686.
    1. Padilla L, Kang H, Washington M, et al. Assessment of interfractional variation of the breast surface following conventional patient positioning for whole‐breast radiotherapy. J Appl Clin Med Phys. 2014;15:4921.
    1. Pashtan IM, Recht A, Ancukiewicz M, et al. External beam accelerated partial‐breast irradiation using 32 Gy in 8 twice‐daily fractions: 5‐year results of a prospective study. Int J Radiat Oncol Biol Phys. 2012;84:e271–1.
    1. Correa C, Harris EE, Leonardi MC, et al. Accelerated partial breast irradiation: executive summary for the update of an ASTRO evidence‐based consensus statement. Pract Radiat Oncol. 2017;7:73–79.
    1. NSABP B‐39, RTOG 0413 . A randomized phase III study of conventional whole breast irradiation versus partial breast irradiation for women with stage 0, I, or II breast cancer. Clin Adv Hematol Oncol. 2006;4:719–721.
    1. Recht A, Ancukiewicz M, Alm El‐Din MA, et al. Lung dose‐volume parameters and the risk of pneumonitis for patients treated with accelerated partial‐breast irradiation using three‐dimensional conformal radiotherapy. J Clin Oncol. 2009;27:3887–3893.
    1. Boer HC, van Sörnsen de Koste JR, Creutzberg Cl, Visser AG, Levendag PC, Heijmen BJ. Electronic portal image assisted reduction of systematic setup errors in head and neck irradiation. Radiother Oncol. 2001;61:299–308.
    1. Bijhold J, Lebesque JV, Hart AAM, Vijlbrief RE. Maximizing setup accuracy using portal images applied to a conformal boost technique for prostatic cancer. Radiother Oncol. 1992;24:261–271.
    1. Chang AJ, Zhao H, Wahab SH, et al. Video surface image guidance for external beam partial breast irradiation. Pract Radiat Oncol. 2012;2:97–105.
    1. Patient message boards. , June 15, 2018