Use of a p64 MW Flow Diverter with Hydrophilic Polymer Coating (HPC) and Prasugrel Single Antiplatelet Therapy for the Treatment of Unruptured Anterior Circulation Aneurysms: Safety Data and Short-term Occlusion Rates

V Hellstern, M Aguilar Pérez, E Henkes, E Donauer, C Wendl, H Bäzner, O Ganslandt, H Henkes, V Hellstern, M Aguilar Pérez, E Henkes, E Donauer, C Wendl, H Bäzner, O Ganslandt, H Henkes

Abstract

Purpose: To assess the safety and short-term occlusion rates in procedures using the p64 MW hydrophilic polymer-coated (HPC) flow diverter (FD) with prasugrel single antiplatelet therapy (SAPT) for the treatment of anterior circulation saccular aneurysms.

Methods: We retrospectively identified patients who underwent treatment of one or more intracranial anterior circulation saccular aneurysms between March 2020 and December 2021 with a p64 MW HPC FD and prasugrel SAPT with verified P2Y12 platelet receptor inhibition. Patients diagnosed with fusiform, dissecting, or recently ruptured aneurysms were excluded. Periprocedural and postprocedural complications, clinical outcomes, and angiographic follow-up results were evaluated.

Results: One hundred and two patients with 132 intracranial aneurysms met the inclusion criteria. Previous or concomitant treatments (e.g., coil occlusion) had been performed on 18 of these aneurysms. The technical success rate (i.e., implantation of the intended FD) was 100% with an average of 1.1 devices implanted per patient. Periprocedural and postprocedural complications occurred in 13.6% and 6.8% of these patients, respectively. No mortality or permanent clinical deterioration (i.e., modified Rankin scale score ≥ 3) were reported. Early follow-up digital subtraction angiography revealed aneurysmal occlusion rates of 72.6% and 83.8% at four and nine months, respectively.

Conclusions: The implantation of a p64 MW HPC FD with prasugrel SAPT is safe and results in rapid, reliable and effective aneurysmal occlusion.

Keywords: Aneurysm; Anti-thrombogenic coating; Flow diverter; Prasugrel; Single antiplatelet therapy.

Conflict of interest statement

VH has a consultancy agreement with the BALT Group; EH has a consulting agreement with phenox GmbH; HH is co-founder and shareholder of phenox GmbH; all other authors report no conflict of interest.

© 2022. The Author(s).

Figures

Fig. 1
Fig. 1
Digital subtraction angiography (DSA). a–d DSA imaging of an aneurysm of the supraclinoid segment of the left internal carotid artery (ICA). a Lateral projection prior to placement of a flow diverter (FD). b Lateral projection after deployment of a 3/15 mm p64 MW HPC; left, a non-contrast-enhanced image demonstrating complete opening of the FD and right, contrast-enhanced image after deployment of the FD showing complete coverage of the aneurysm neck. Follow-up (FU) angiography at c 39 days and d 251 days postprocedure. The lateral view documents the complete occlusion of the aneurysm (OKM D). In panels b–d, the distal and proximal ends of the FD are indicated by the black and blue arrows, respectively. e–i DSA imaging of a large symptomatic aneurysm of the cavernous segment of the right ICA and a small aneurysm at the origin of the choroidal artery. e Posterior-anterior projection before placement of the first FD. (f) Lateral (non-contrast-enhanced) image and anterior–posterior projection after deployment of a 3.5/21 mm p64MW HPC FD (black arrows indicating the ends of the device). g FU angiography at 56 days postprocedure revealed the complete occlusion of the cavernous aneurysm (OKM D) (black arrows indicating the ends of the FD device; the small aneurysm at the choroidal segment of the ICA is as indicated by the red arrow). h This lesion was treated in the same session with a 4/12 mm p64 MW HPC FD (blue arrows indicating the ends of the second FD; non-contrast-enhanced lateral view after the deployment of a second FD). i FU angiography at 91 days postprocedure with a lateral view that revealed complete occlusion of both aneurysms (OKM D) with mild intimal hyperplasia in the distal FD (blue arrows at the ends of the distal FD and a black arrow indicating the distal end of the proximal FD; the proximal end of this latter FD is not shown in this picture)
Fig. 2
Fig. 2
Progressive increase in aneurysmal occlusion over time. Occlusion was evaluated (a) overall, and for (b) ICA aneurysms; (c) MCA aneurysms; (d) MCA bifurcation aneurysms; (e) ACA aneurysms; and (f) AcomA aneurysms. FU, follow-up; OKM, O'Kelly-Marotta; ICA, internal carotid artery; MCA, middle cerebral artery; ACA, anterior cerebral artery; AcomA, anterior communicating artery
Fig. 2
Fig. 2
Progressive increase in aneurysmal occlusion over time. Occlusion was evaluated (a) overall, and for (b) ICA aneurysms; (c) MCA aneurysms; (d) MCA bifurcation aneurysms; (e) ACA aneurysms; and (f) AcomA aneurysms. FU, follow-up; OKM, O'Kelly-Marotta; ICA, internal carotid artery; MCA, middle cerebral artery; ACA, anterior cerebral artery; AcomA, anterior communicating artery

References

    1. Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ. A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke. 2007;38:2346–2352. doi: 10.1161/STROKEAHA.106.479576.
    1. D’Urso PI, Lanzino G, Cloft HJ, Kallmes DF. Flow diversion for intracranial aneurysms. Stroke. 2011;42:2363–2368. doi: 10.1161/STROKEAHA.111.620328.
    1. Skukalek SL, Winkler AM, Kang J, Dion JE, Cawley CM, Webb A, Dannenbaum MJ, Schuette AJ, Asbury B, Tong FC. Effect of antiplatelet therapy and platelet function testing on hemorrhagic and thrombotic complications in patients with cerebral aneurysms treated with the pipeline embolization device: a review and meta-analysis. J Neurointerv Surg. 2016;8:58–65. doi: 10.1136/neurintsurg-2014-011145.
    1. Tonetti DA, Jankowitz BT, Gross BA. Antiplatelet therapy in flow diversion. Neurosurgery. 2020;86:S47–S52. doi: 10.1093/neuros/nyz391.
    1. Aguilar-Perez M, Hellstern V, AlMatter M, Wendl C, Bäzner H, Ganslandt O, Henkes H. The p48 flow modulation device with hydrophilic polymer coating (HPC) for the treatment of acutely ruptured aneurysms: early clinical experience using single antiplatelet therapy. Cardiovasc Intervent Radiol. 2020;43:740–748. doi: 10.1007/s00270-020-02418-4.
    1. Bhogal P, Bleise C, Chudyk J, Lylyk I, Perez N, Henkes H, Lylyk P. The p48_HPC antithrombogenic flow diverter: initial human experience using single antiplatelet therapy. J Int Med Res. 2020;48:300060519879580. doi: 10.1177/0300060519879580.
    1. Bhogal P, Petrov A, Rentsenkhu G, Nota B, Ganzorig E, Regzengombo B, Jagusch S, Henkes E, Henkes H. Early clinical experience with the p48MW HPC and p64MW HPC flow diverters in the anterior circulation aneurysm using single anti-platelet treatment. Interv Neuroradiol. 2021 doi: 10.1177/15910199211029503.
    1. O’Kelly CJ, Krings T, Fiorella D, Marotta TR. A novel grading scale for the angiographic assessment of intracranial aneurysms treated using flow diverting stents. Interv Neuroradiol. 2010;16:133–137. doi: 10.1177/159101991001600204.
    1. Wilson JTL, Hareendran A, Grant M, Baird T, Schulz UGR, Muir KW, Bone I. Improving the assessment of outcomes in stroke. Stroke. 2002;33:2243–2246. doi: 10.1161/.
    1. Cruz JP, Marotta T, O’Kelly C, Holtmannspötter M, Saliou G, Willinsky R, Krings T, Agid R. Enhancing brain lesions after endovascular treatment of aneurysms. AJNR Am J Neuroradiol. 2014;35:1954–1958. doi: 10.3174/ajnr.A3976.
    1. AlMatter M, Aguilar Péreza M, Bhogal P, Hellstern V, Ganslandt O, Henkes H. Results of interdisciplinary management of 693 patients with aneurysmal subarachnoid hemorrhage: clinical outcome and relevant prognostic factors. Clin Neurol Neurosurg. 2018;167:106–111. doi: 10.1016/j.clineuro.2018.02.022.
    1. Chalouhi N, Tjoumakaris S, Starke RM, Gonzalez LF, Randazzo C, Hasan D, McMahon JF, Singhal S, Moukarzel LA, Dumont AS, Rosenwasser R, Jabbour P. Comparison of flow diversion and coiling in large unruptured intracranial saccular aneurysms. Stroke. 2013;44:2150–2154. doi: 10.1161/STROKEAHA.113.001785.
    1. Perez MA, Henkes E, Hellstern V, Candel CS, Wendl C, Bäzner H, Ganslandt O, Henkes H. Endovascular treatment of anterior circulation aneurysms with the p64 flow modulation device: mid- and long-term results in 617 aneurysms from a single center. Neurosurgery. 2021;89:S164. doi: 10.1093/neuros/opaa425_S164.
    1. Bonafe A, Perez MA, Henkes H, Lylyk P, Bleise C, Gascou G, Sirakov S, Sirakov A, Stockx L, Turjman F, Petrov A, Roth C, Narata A-P, Barreau X, Loehr C, Berlis A, Pierot L, Miś M, Goddard T, Clifton A, Klisch J, Wałęsa C, Dall’Olio M, Spelle L, Clarencon F, Yakovlev S, Keston P, Nuzzi NP, Dima S, Wendl C, Willems T, Schramm P (2021) Diversion-p64: results from an international, prospective, multicenter, single-arm post-market study to assess the safety and effectiveness of the p64 flow modulation device. J NeuroIntervent Surg neurintsurg-2021–017809. 10.1136/neurintsurg-2021-017809
    1. Pierot L, Spelle L, Berge J, Januel A-C, Herbreteau D, Aggour M, Piotin M, Biondi A, Barreau X, Mounayer C, Papagiannaki C, Lejeune J-P, Gauvrit J-Y, Derelle A-L, Chabert E, Costalat V. SAFE study (Safety and efficacy analysis of FRED embolic device in aneurysm treatment): 1-year clinical and anatomical results. J Neurointerv Surg. 2019;11:184–189. doi: 10.1136/neurintsurg-2018-014261.
    1. Hanel RA, Kallmes DF, Lopes DK, Nelson PK, Siddiqui A, Jabbour P, Pereira VM, Szikora István I, Zaidat OO, Bettegowda C, Colby GP, Mokin M, Schirmer C, Hellinger FR, Given Ii C, Krings T, Taussky P, Toth G, Fraser JF, Chen M, Priest R, Kan P, Fiorella D, Frei D, Aagaard-Kienitz B, Diaz O, Malek AM, Cawley CM, Puri AS. Prospective study on embolization of intracranial aneurysms with the pipeline device: the PREMIER study 1 year results. J Neurointerv Surg. 2020;12:62–66. doi: 10.1136/neurintsurg-2019-015091.
    1. Taschner CA, Stracke CP, Dorn F, Kadziolka KB, Kreiser K, Solymosi L, Pham M, Buhk JH, Turowski B, Reith W, Elsheikh S, Meckel S, Janssen H, Hammer A, Beuing O, Jansen O, Urbach H, Knauth M, Jenkner C, Chapot R. Derivo embolization device in the treatment of unruptured intracranial aneurysms: a prospective multicenter study. J Neurointerv Surg. 2021;13:541–546. doi: 10.1136/neurintsurg-2020-016303.
    1. Diestro JDB, Adeeb N, Dibas M, Boisseau W, Harker P, Brinjikji W, Xiang S, Joyce E, Shapiro M, Raz E, Parra-Farinas C, Pickett G, Alotaibi NM, Regenhardt RW, Bernstock JD, Spears J, Griessenauer CJ, Burkhardt J-K, Hafeez MU, Kan P, Grandhi R, Taussky P, Nossek E, Hong T, Zhang H, Rinaldo L, Lanzino G, Stapleton CJ, Rabinov JD, Patel AB, Marotta TR, Roy D, Dmytriw AA. Flow diversion for middle cerebral artery aneurysms: an international cohort study. Neurosurgery. 2021;89:1112–1121. doi: 10.1093/neuros/nyab365.
    1. Lobsien D, Clajus C, Behme D, Ernst M, Riedel CH, Abu-Fares O, Götz FG, Fiorella D, Klisch J. Aneurysm treatment in acute SAH with hydrophilic-coated flow diverters under single-antiplatelet therapy: a 3-center experience. AJNR Am J Neuroradiol. 2021;42:508–515. doi: 10.3174/ajnr.A6942.
    1. Guzzardi G, Galbiati A, Stanca C, Del Sette B, Paladini A, Cossandi C, Carriero A. Flow diverter stents with hydrophilic polymer coating for the treatment of acutely ruptured aneurysms using single antiplatelet therapy: Preliminary experience. Interv Neuroradiol. 2020;26:525–531. doi: 10.1177/1591019920950878.
    1. Tan LA, Keigher KM, Munich SA, Moftakhar R, Lopes DK. Thromboembolic complications with pipeline embolization device placement: impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) value. J Neurointerv Surg. 2015;7:217–221. doi: 10.1136/neurintsurg-2014-011111.
    1. Brasiliense LBC, Stanley MA, Grewal SS, Cloft HJ, Sauvageau E, Lanzino G, Miller D, Kallmes DF, Hanel R. Silent ischemic events after Pipeline embolization device: a prospective evaluation with MR diffusion-weighted imaging. J Neurointerv Surg. 2016;8:1136–1139. doi: 10.1136/neurintsurg-2015-012091.
    1. Iosif C, Lecomte J-C, Pedrolo-Silveira E, Mendes G, Boncoeur Martel M-P, Saleme S, Mounayer C. Evaluation of ischemic lesion prevalence after endovascular treatment of intracranial aneurysms, as documented by 3-T diffusion-weighted imaging: a 2-year, single-center cohort study. J Neurosurg. 2018;128:982–991. doi: 10.3171/2016.11.JNS161020.
    1. Safain MG, Roguski M, Heller RS, Malek AM. Flow diverter therapy with the pipeline embolization device is associated with an elevated rate of delayed fluid-attenuated inversion recovery lesions. Stroke. 2016;47:789–797. doi: 10.1161/STROKEAHA.115.010522.
    1. Pikis S, Mantziaris G, Mamalis V, Barkas K, Tsanis A, Lyra S, Karkoulias K, Petrosyan T, Archontakis E. Diffusion weighted image documented cerebral ischemia in the postprocedural period following pipeline embolization device with shield technology treatment of unruptured intracranial aneurysms: a prospective, single center study. J NeuroInterventional Surg. 2020;12:407–411. doi: 10.1136/neurintsurg-2019-015363.
    1. Pierot L, Soize S, Cappucci M, Manceau P-F, Riva R, Eker OF. Surface-modified flow diverter p48-MW-HPC: preliminary clinical experience in 28 patients treated in two centers. J Neuroradiol. 2021;48:195–199. doi: 10.1016/j.neurad.2020.11.006.
    1. de Castro-Afonso LH, Nakiri GS, Abud TG, Monsignore LM, Freitas RK, de Oliveira RS, Colli BO, dos Santos AC, Abud DG. Treatment of distal unruptured intracranial aneurysms using a surface-modified flow diverter under prasugrel monotherapy: a pilot safety trial. J NeuroInterventional Surg. 2021;13:647–651. doi: 10.1136/neurintsurg-2020-017262.
    1. de Castro-Afonso LH, Nakiri GS, Abud TG, Monsignore LM, de Freitas RK, Abud DG. Aspirin monotherapy in the treatment of distal intracranial aneurysms with a surface modified flow diverter: a pilot study. J Neurointerv Surg. 2021;13:336–341. doi: 10.1136/neurintsurg-2020-017024.
    1. Manning NW, Cheung A, Phillips TJ, Wenderoth JD. Pipeline shield with single antiplatelet therapy in aneurysmal subarachnoid haemorrhage: multicentre experience. J Neurointerv Surg. 2019;11:694–698. doi: 10.1136/neurintsurg-2018-014363.
    1. Agnoletto GJ, Aguilar-Salinas P, Santos R, Sauvageau E, Hanel RA. PED flex with shield technology: a feasible alternative for fusiform MCA aneurysms. Stroke Vasc Neurol. 2018;3:185–188. doi: 10.1136/svn-2017-000132.
    1. Trivelato FP, Wajnberg E, Rezende MTS, Ulhôa AC, Piske RL, Abud TG, de Castro-Afonso LH, Abath CGC, Nakiri GS, Araújo JFS, Silva JLJ, Tosello RT, Vanzin JR, Manzato LB, Baccin CE, da Mota BAA, Abud DG. Safety and effectiveness of the pipeline flex embolization device with shield technology for the treatment of intracranial aneurysms: midterm results from a multicenter study. Neurosurgery. 2020;87:104–111. doi: 10.1093/neuros/nyz356.
    1. Rice H, Martínez Galdámez M, Holtmannspötter M, Spelle L, Lagios K, Ruggiero M, Vega P, Sonwalkar H, Chapot R, Lamin S. Periprocedural to 1-year safety and efficacy outcomes with the pipeline embolization device with shield technology for intracranial aneurysms: a prospective, post-market, multi-center study. J Neurointerv Surg. 2020;12:1107–1112. doi: 10.1136/neurintsurg-2020-015943.
    1. Martínez-Galdámez M, Lamin SM, Lagios KG, Liebig T, Ciceri EF, Chapot R, Stockx L, Chavda S, Kabbasch C, Faragò G, Nordmeyer H, Boulanger T, Piano M, Boccardi EP. Treatment of intracranial aneurysms using the pipeline flex embolization device with Shield Technology: angiographic and safety outcomes at 1-year follow-up. J Neurointerv Surg. 2019;11:396–399. doi: 10.1136/neurintsurg-2018-014204.
    1. Martínez-Galdámez M, Lamin SM, Lagios KG, Liebig T, Ciceri EF, Chapot R, Stockx L, Chavda S, Kabbasch C, Farago G, Nordmeyer H, Boulanger T, Piano M, Boccardi EP. Periprocedural outcomes and early safety with the use of the pipeline flex embolization device with shield technology for unruptured intracranial aneurysms: preliminary results from a prospective clinical study. J Neurointerv Surg. 2017;9:772–776. doi: 10.1136/neurintsurg-2016-012896.
    1. Cappuzzo JM, Monteiro A, Taylor MN, Waqas M, Baig AA, Almayman F, Davies JM, Snyder KV, Siddiqui AH, Levy EI. First U.S. experience using the pipeline flex embolization device with shield technology for treatment of intracranial aneurysms. World Neurosurg. 2022;159:e184–e191. doi: 10.1016/j.wneu.2021.12.031.
    1. Hanel RA, Aguilar-Salinas P, Brasiliense LB, Sauvageau E (2017) First US experience with Pipeline Flex with Shield Technology using aspirin as antiplatelet monotherapy. BMJ Case Rep 2017:bcr2017219406. 10.1136/bcr-2017-219406
    1. Guerrero BP, Pacheco CD, Saied A, Joshi K, Rodríguez C, Martínez-Galdámez M, Lopes DK. First human evaluation of endothelial healing after a pipeline flex embolization device with shield technology implanted in posterior circulation using optical coherence tomography. Neurointervention. 2018;13:129–132. doi: 10.5469/neuroint.2018.01032.

Source: PubMed

3
Předplatit