Re-evaluation of low intensity pulsed ultrasound in treatment of tibial fractures (TRUST): randomized clinical trial

TRUST Investigators writing group, Jason W Busse, Mohit Bhandari, Thomas A Einhorn, Emil Schemitsch, James D Heckman, Paul Tornetta 3rd, Kwok-Sui Leung, Diane Heels-Ansdell, Sun Makosso-Kallyth, Gregory J Della Rocca, Clifford B Jones, Gordon H Guyatt, TRUST Investigators writing group, Jason W Busse, Mohit Bhandari, Thomas A Einhorn, Emil Schemitsch, James D Heckman, Paul Tornetta 3rd, Kwok-Sui Leung, Diane Heels-Ansdell, Sun Makosso-Kallyth, Gregory J Della Rocca, Clifford B Jones, Gordon H Guyatt

Abstract

Objective: To determine whether low intensity pulsed ultrasound (LIPUS), compared with sham treatment, accelerates functional recovery and radiographic healing in patients with operatively managed tibial fractures.

Design: A concealed, randomized, blinded, sham controlled clinical trial with a parallel group design of 501 patients, enrolled between October 2008 and September 2012, and followed for one year.

Setting: 43 North American academic trauma centers.

Participants: Skeletally mature men or women with an open or closed tibial fracture amenable to intramedullary nail fixation. Exclusions comprised pilon fractures, tibial shaft fractures that extended into the joint and required reduction, pathological fractures, bilateral tibial fractures, segmental fractures, spiral fractures >7.5 cm in length, concomitant injuries that were likely to impair function for at least as long as the patient's tibial fracture, and tibial fractures that showed <25% cortical contact and >1 cm gap after surgical fixation. 3105 consecutive patients who underwent intramedullary nailing for tibial fracture were assessed, 599 were eligible and 501 provided informed consent and were enrolled.

Interventions: Patients were allocated centrally to self administer daily LIPUS (n=250) or use a sham device (n=251) until their tibial fracture showed radiographic healing or until one year after intramedullary fixation.

Main outcome measures: Primary registry specified outcome was time to radiographic healing within one year of fixation; secondary outcome was rate of non-union. Additional protocol specified outcomes included short form-36 (SF-36) physical component summary (PCS) scores, return to work, return to household activities, return to ≥80% of function before injury, return to leisure activities, time to full weight bearing, scores on the health utilities index (mark 3), and adverse events related to the device.

Results: SF-36 PCS data were acquired from 481/501 (96%) patients, for whom we had 2303/2886 (80%) observations, and radiographic healing data were acquired from 482/501 (96%) patients, of whom 82 were censored. Results showed no impact on SF-36 PCS scores between LIPUS and control groups (mean difference 0.55, 95% confidence interval -0.75 to 1.84; P=0.41) or for the interaction between time and treatment (P=0.30); minimal important difference is 3-5 points) or in other functional measures. There was also no difference in time to radiographic healing (hazard ratio 1.07, 95% confidence interval 0.86 to 1.34; P=0.55). There were no differences in safety outcomes between treatment groups. Patient compliance was moderate; 73% of patients administered ≥50% of all recommended treatments.

Conclusions: Postoperative use of LIPUS after tibial fracture fixation does not accelerate radiographic healing and fails to improve functional recovery.Study registration ClinicalTrialGov Identifier: NCT00667849.

Conflict of interest statement

Competing interests: All authors have completed the ICMJE uniform disclosure form at www.icmje.org/coi_disclosure.pdf and declare: TAE, ES, and MB have received consulting fees from Smith & Nephew, the manufacturer of the study device. PT receives royalties from Smith & Nephew. GJDR is a paid consultant for Bioventus LLC, which is 51% owned by Essex Woodlands and 49% by Smith & Nephew. MB is supported, in part, by a Canada research chair, McMaster University.

Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions.

Figures

https://www.ncbi.nlm.nih.gov/pmc/articles/instance/5080447/bin/busj032767.f1_default.jpg
Fig 1 Recruitment of patients with tibial fractures and follow-up schedule
https://www.ncbi.nlm.nih.gov/pmc/articles/instance/5080447/bin/busj032767.f2_default.jpg
Fig 2 SF-36 physical component summary (PCS) scores over time in patients with tibial fracture according to treatment with low intensity pulsed ultrasound or sham device
https://www.ncbi.nlm.nih.gov/pmc/articles/instance/5080447/bin/busj032767.f3_default.jpg
Fig 3 Kaplan-Meier time to event curve for radiographic fracture healing in patients with tibial fracture according to treatment with low intensity pulsed ultrasound or sham device

References

    1. Heckman JD, Sarasohn-Kahn J. The economics of treating tibia fractures. The cost of delayed unions. Bull Hosp Jt Dis 1997;56:63-72..
    1. Busse JW, Morton E, Lacchetti C, Guyatt GH, Bhandari M. Current management of tibial shaft fractures: a survey of 450 Canadian orthopedic trauma surgeons. Acta Orthop 2008;79:689-94. 10.1080/17453670810016722 .
    1. Sarmiento A, Sharpe FE, Ebramzadeh E, Normand P, Shankwiler J. Factors influencing the outcome of closed tibial fractures treated with functional bracing. Clin Orthop Relat Res 1995;(315):8-24..
    1. Wachovia Capital Markets. Equity research: bone growth stimulation 2008 outlook. 2007.
    1. Griffin XL, Parsons N, Costa ML, Metcalfe D. Ultrasound and shockwave therapy for acute fractures in adults. Cochrane Database Syst Rev 2014;6:CD008579..
    1. Busse JW, Kaur J, Mollon B, et al. Low intensity pulsed ultrasonography for fractures: systematic review of randomised controlled trials. BMJ 2009;338:b351 10.1136/bmj.b351 .
    1. The worldwide orthopaedic market: 2004-2005. Knowledge Enterprises. The Ortho People. October, 2005.
    1. Cassel CK, Guest JA. Choosing wisely: helping physicians and patients make smart decisions about their care. JAMA 2012;307:1801-2. 10.1001/jama.2012.476 .
    1. Lenzer J. Unnecessary care: are doctors in denial and is profit driven healthcare to blame?BMJ 2012;345:e6230 10.1136/bmj.e6230 .
    1. The BMJ's Too Much Medicine Campaign
    1. Moynihan R, Glasziou P, Woloshin S, Schwartz L, Santa J, Godlee F. Winding back the harms of too much medicine. BMJ 2013;346:f1271 10.1136/bmj.f1271 .
    1. Busse JW, Bhandari M, Einhorn TA, et al. The TRUST Investigators. Trial to re-evaluate ultrasound in the treatment of tibial fractures (TRUST): a multicenter randomized pilot study. Trials 2014;15:206 10.1186/1745-6215-15-206 .
    1. McHorney CA, Ware JE Jr, , Lu JF, Sherbourne CD. The MOS 36-item Short-Form Health Survey (SF-36): III. Tests of data quality, scaling assumptions, and reliability across diverse patient groups. Med Care 1994;32:40-66. 10.1097/00005650-199401000-00004 .
    1. Ware JE Jr, ed. SF-36 Physical & Mental Health Summary Scales: A User’s Manual.Quality Metric Inc, 1997.
    1. Bhandari M, Sprague S, Hanson B, et al. Health-related quality of life following operative treatment of unstable ankle fractures: a prospective observational study. J Orthop Trauma 2004;18:338-45. 10.1097/00005131-200407000-00002 .
    1. Busse JW, Bhandari M, Guyatt GH, et al. SPRINT Investigators. Use of both Short Musculoskeletal Function Assessment questionnaire and Short Form-36 among tibial-fracture patients was redundant. J Clin Epidemiol 2009;62:1210-7. 10.1016/j.jclinepi.2009.01.014 .
    1. Kooistra BW, Dijkman BG, Busse JW, Sprague S, Schemitsch EH, Bhandari M. The radiographic union scale in tibial fractures: reliability and validity. J Orthop Trauma 2010;24(Suppl 1):S81-6. 10.1097/BOT.0b013e3181ca3fd1 .
    1. Horsman J, Furlong W, Feeny D, Torrance G. The Health Utilities Index (HUI): concepts, measurement properties and applications. Health Qual Life Outcomes 2003;1:54 10.1186/1477-7525-1-54 .
    1. Stewart AL, Greenfield S, Hays RD, et al. Functional status and well-being of patients with chronic conditions. Results from the Medical Outcomes Study. JAMA 1989;262:907-13. 10.1001/jama.1989.03430070055030 .
    1. O’Brien PC, Fleming TR. A multiple testing procedure for clinical trials. Biometrics 1979;35:549-56. 10.2307/2530245 .
    1. Carpenter J, Kenward M. Multiple Imputation and its Application. 1st ed Wiley, 201310.1002/9781119942283.
    1. Graham JW. Missing data analysis: making it work in the real world. Annu Rev Psychol 2009;60:549-76. 10.1146/annurev.psych.58.110405.085530 .
    1. Tunis SR, Stryer DB, Clancy CM. Practical clinical trials: increasing the value of clinical research for decision making in clinical and health policy. JAMA 2003;290:1624-32. 10.1001/jama.290.12.1624 .
    1. Tsumaki N, Kakiuchi M, Sasaki J, Ochi T, Yoshikawa H. Low-intensity pulsed ultrasound accelerates maturation of callus in patients treated with opening-wedge high tibial osteotomy by hemicallotasis. J Bone Joint Surg Am 2004;86-A:2399-405..
    1. Rue JP, Armstrong DW 3rd, , Frassica FJ, Deafenbaugh M, Wilckens JH. The effect of pulsed ultrasound in the treatment of tibial stress fractures. Orthopedics 2004;27:1192-5..
    1. Osterberg L, Blaschke T. Adherence to medication. N Engl J Med 2005;353:487-97. 10.1056/NEJMra050100 .
    1. Falck-Ytter Y, Francis CW, Johanson NA, et al. American College of Chest Physicians. Prevention of VTE in orthopedic surgery patients: Antithrombotic Therapy and Prevention of Thrombosis, 9th ed: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines. Chest 2012;141(Suppl):e278S-325S. 10.1378/chest.11-2404 .
    1. Rubin C, Bolander M, Ryaby JP, Hadjiargyrou M. The use of low-intensity ultrasound to accelerate the healing of fractures. J Bone Joint Surg Am 2001;83-A:259-70..
    1. Thompson M, Heneghan C, Billingsley M, Cohen D. Medical device recalls and transparency in the UK. BMJ 2011;342:d2973 10.1136/bmj.d2973 .
    1. Marcus HJ, Payne CJ, Hughes-Hallett A, et al. Regulatory approval of new medical devices: cross sectional study. BMJ 2016;353:i2587 10.1136/bmj.i2587 .
    1. In Memoriam—John Ryaby. The Bioelectromagnetics Society.
    1. Heckman JD, Ryaby JP, McCabe J, Frey JJ, Kilcoyne RF. Acceleration of tibial fracture-healing by non-invasive, low-intensity pulsed ultrasound. J Bone Joint Surg Am 1994;76:26-34..
    1. Kristiansen TK, Ryaby JP, McCabe J, Frey JJ, Roe LR. Accelerated healing of distal radial fractures with the use of specific, low-intensity ultrasound. A multicenter, prospective, randomized, double-blind, placebo-controlled study. J Bone Joint Surg Am 1997;79:961-73..
    1. Wilmshurst P. The regulation of medical devices. BMJ 2011;342:d2822 10.1136/bmj.d2822 .

Source: PubMed