Seizures from transcranial magnetic stimulation 2012-2016: Results of a survey of active laboratories and clinics

Adam J Lerner, Eric M Wassermann, Diana I Tamir, Adam J Lerner, Eric M Wassermann, Diana I Tamir

Abstract

Objective: Transcranial magnetic stimulation (TMS) can cause seizures in healthy individuals and patients. However, the rate at which this occurs is unknown. We estimated the risk of seizure and other adverse events with TMS.

Methods: We surveyed laboratories and clinics about seizures and other events observed between 2012 and 2016 (inclusive). Respondents (N = 174) reported an estimated 318,560 TMS sessions.

Results: Twenty-four seizures were reported (.08/1000 sessions). TMS delivered within published guidelines to subjects without recognized risk factors caused 4 seizures (<.02/1000 sessions). High-frequency (>1 Hz) rTMS delivered within published guidelines to individuals without known risk factors was no more likely to cause seizures than low-frequency and single/paired-pulse TMS. Subject risk factors (e.g., brain lesions and epilepsy) increased seizure risk substantially. Seizures appeared more common when safety guidelines were exceeded. Seizures were most likely to occur within the first few exposures to TMS.

Conclusions: TMS delivered within published guidelines to individuals without risk factors appears to cause fewer than 1 seizure per 60,000 sessions. The assumption that repetitive TMS is riskier than single and paired pulses under these conditions should be reevaluated.

Significance: This information should help laboratories, clinics, and regulatory authorities form updated safety policies for TMS.

Keywords: Risk; Seizure; Transcranial magnetic stimulation.

Conflict of interest statement

Conflict of interest

None of the authors have potential conflicts of interest to be disclosed.

Copyright © 2019 International Federation of Clinical Neurophysiology. All rights reserved.

References

    1. Boes AD, Stern AP, Bernstein M, Hooker JE, Connor A, Press DZ, Pascual-Leone A. H-Coil repetitive transcranial magnetic stimulation induced seizure in an adult with major depression: a case report. Brain Stimul 2016;9(4):632–3. 10.1016/j.brs.2016.04.013.
    1. Bravo M, Kamel H, Custer B, Tomasulo P. Factors associated with fainting: before, during and after whole blood donation. Vox Sang 2011;101(4):303–12. 10.1111/j.1423-0410.2011.01494.x.
    1. Chen R, Gerloff C, Classen J, Wassermann EM, Hallett M, Cohen LG. Safety of different inter-train intervals for repetitive transcranial magnetic stimulation and recommendations for safe ranges of stimulation parameters. Electroencephalogr Clin Neurophysiol 1997;105:415–21.
    1. Cullen KR, Jasberg S, Nelson B, Klimes-Dougan B, Lim KO, Croarkin PE. Seizure induced by deep transcranial magnetic stimulation in an adolescent with depression. J Child Adolesc Psychopharmacol 2016;26(7):637–41. 10.1089/cap.2016.0070.
    1. Dhuna AK, Gates JR, Pascual-Leone A. Transcranial magnetic stimulation in patients with epilepsy. Neurology 1991;41:1067–71.
    1. Groiss SJ, Trenado C, Sabel M, Schnitzler A, Wojtecki L. Focal seizure induced by preoperative navigated transcranial magnetic stimulation in a patient with anaplastic oligoastrocytoma. Brain Stimul 2017;10(2):331–2. 10.1016/j.brs.2016.12.006.
    1. Hirtz D, Thurman DJ, Gwinn-Hardy K, Mohamed M, Chaudhuri AR, Zalutsky R. How common are the “common” neurologic disorders? Neurology 2007;68 (5):326–37. 10.1212/01.wnl.0000252807.38124.a3.
    1. Hömberg V, Netz J. Generalized seizures induced by transcranial magnetic stimulation of motor cortex. Lancet 1989;2:1223.
    1. Merton PA, Morton HB. Stimulation of the cerebral cortex in the intact human subject. Nature 1980;285:227.
    1. Pascual-Leone A, Houser CM, Reeves K, Shotland LI, Grafman J, Sato S, et al. Safety of rapid-rate transcranial magnetic stimulation in normal volunteers. Electroencephalogr Clin Neurophysiol 1993;89:120–30.
    1. Rossi S, Hallett M, Rossini PM, Pascual-Leone A. Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and research. Clin Neurophysiol 2009;120(12):2008–39. doi: S1388-2457(09)00519-7[pii] 10.1016/j.clinph.2009.08.016..
    1. Wassermann EM. Risk and safety of repetitive transcranial magnetic stimulation: report and suggested guidelines from the International Workshop on the Safety of Repetitive Transcranial Magnetic Stimulation, June 5–7, 1996. Electroencephalogr Clin Neurophysiol 1998;108(1):1–16.