Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised controlled studies

Roman Hovorka, Kavita Kumareswaran, Julie Harris, Janet M Allen, Daniela Elleri, Dongyuan Xing, Craig Kollman, Marianna Nodale, Helen R Murphy, David B Dunger, Stephanie A Amiel, Simon R Heller, Malgorzata E Wilinska, Mark L Evans, Roman Hovorka, Kavita Kumareswaran, Julie Harris, Janet M Allen, Daniela Elleri, Dongyuan Xing, Craig Kollman, Marianna Nodale, Helen R Murphy, David B Dunger, Stephanie A Amiel, Simon R Heller, Malgorzata E Wilinska, Mark L Evans

Abstract

Objective: To compare the safety and efficacy of overnight closed loop delivery of insulin (artificial pancreas) with conventional insulin pump therapy in adults with type 1 diabetes.

Design: Two sequential, open label, randomised controlled crossover, single centre studies.

Setting: Clinical research facility.

Participants: 24 adults (10 men, 14 women) with type 1 diabetes, aged 18-65, who had used insulin pump therapy for at least three months: 12 were tested after consuming a medium sized meal and the other 12 after consuming a larger meal accompanied by alcohol.

Intervention: During overnight closed loop delivery, sensor measurements of glucose were fed into a computer algorithm, which advised on insulin pump infusion rates at 15 minute intervals. During control nights, conventional insulin pump settings were applied. One study compared closed loop delivery of insulin with conventional pump therapy after a medium sized evening meal (60 g of carbohydrates) at 1900, depicting the scenario of "eating in." The other study was carried out after a later large evening meal (100 g of carbohydrates) at 2030, accompanied by white wine (0.75 g/kg ethanol) and depicted the scenario of "eating out."

Main outcome measures: The primary outcome was the time plasma glucose levels were in target (3.91-8.0 mmol/L) during closed loop delivery and a comparable control period. Secondary outcomes included pooled data analysis and time plasma glucose levels were below target (≤ 3.9 mmol/L).

Results: For the eating in scenario, overnight closed loop delivery of insulin increased the time plasma glucose levels were in target by a median 15% (interquartile range 3-35%), P = 0.002. For the eating out scenario, closed loop delivery increased the time plasma glucose levels were in target by a median 28% (2-39%), P = 0.01. Analysis of pooled data showed that the overall time plasma glucose was in target increased by a median 22% (3-37%) with closed loop delivery (P < 0.001). Closed loop delivery reduced overnight time spent hypoglycaemic (plasma glucose ≤ 3.9 mmol/L) by a median 3% (0-20%), P=0.04, and eliminated plasma glucose concentrations below 3.0 mmol/L after midnight.

Conclusion: These two small crossover trials suggest that closed loop delivery of insulin may improve overnight control of glucose levels and reduce the risk of nocturnal hypoglycaemia in adults with type 1 diabetes. Trial registration ClinicalTrials.gov NCT00910767 and NCT00944619.

Conflict of interest statement

Competing interests: All authors have completed the Unified Competing Interest form at www.icmje.org/coi_disclosure.pdf (available on request from the corresponding author) and declare: in the previous 3 years RH has received speaker honorariums from Minimed Medtronic, Lifescan, and Novo Nordisk, served on advisory panel for Animas and Minimed Medtronic, received licence fees from BBraun and Beckton Dickinson, and served as a consultant to Beckton Dickinson, BBraun, and Profil; CK has served as a consultant to Medtronic International Trading Sàrl and Diabetes Technology Management; SAA has served on advisory boards for Medtronic and Johnson and Johnson and has received speaker honorariums for Medtronic, Animas, and Roche; SRH has received speaker honorariums from NovoNordisk, Eli Lilly, Sanofi-Aventis, and Lifescan and served on advisory panels for NovoNordisk and Eli Lily; MLE has received speaker honorariums from Eli Lilly and served on advisory panels for Medtronic, Sanofi-Aventis, and Cellnovo; HRM has received speaker honorariums from Minimed Medtronic; MEW has received license fees from Becton Dickinson and has served as a consultant to Beckton Dickinson; KK, JH, JMA, DE, DX, and MN declare no competing financial interests. RH, MEW, and DBD report patent applications in the area of closed loop delivery of insulin.

Figures

https://www.ncbi.nlm.nih.gov/pmc/articles/instance/4787992/bin/hovr823856.f1_default.jpg
Fig 1 Design of two studies comparing closed loop delivery of insulin with conventional insulin pump therapy after two meal scenarios: eating in and eating out
https://www.ncbi.nlm.nih.gov/pmc/articles/instance/4787992/bin/hovr823856.f2_default.jpg
Fig 2 Flow of participants through study comparing closed loop delivery of insulin with conventional insulin pump therapy after a medium sized evening meal (eating in scenario), and study comparing closed loop delivery with insulin pump therapy after a large evening meal accompanied by alcohol (eating out scenario)
https://www.ncbi.nlm.nih.gov/pmc/articles/instance/4787992/bin/hovr823856.f3_default.jpg
Fig 3 Profiles (medians and interquartile ranges) of plasma glucose and insulin concentrations and insulin infusion in eating in scenario (12 participants). Outlying squares represent hypoglycaemic events (glucose level <3.0 mmol/L)
https://www.ncbi.nlm.nih.gov/pmc/articles/instance/4787992/bin/hovr823856.f4_default.jpg
Fig 4 Profiles (medians and interquartile ranges) of plasma glucose and insulin concentrations and insulin infusion in the eating out scenario (12 participants). Outlying squares represent hypoglycaemic events (glucose level <3.0 mmol/L)
https://www.ncbi.nlm.nih.gov/pmc/articles/instance/4787992/bin/hovr823856.f5_default.jpg
Fig 5 Distribution of plasma glucose values during closed loop insulin delivery and conventional insulin pump therapy (continuous subcutaneous insulin infusion) combining data collected from midnight until end of the eating in scenario and the eating out scenario. Percentages represent total time plasma glucose level was below, at, and above target from midnight until end of closed loop delivery

References

    1. World Health Organization. Prevention of diabetes mellitus. Report of a WHO study group. WHO, 1994.
    1. Shaw JE, Sicree RA, Zimmet PZ. Global estimates of the prevalence of diabetes for 2010 and 2030. Diabetes Res Clin Pract 2010;87:4-14.
    1. Onkamo P, Vaananen S, Karvonen M, Tuomilehto J. Worldwide increase in incidence of type 1 diabetes—the analysis of the data on published incidence trends. Diabetologia 1999;42:1395-403.
    1. Diabetic Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977-86.
    1. Amiel S, Beveridge S, Bradley C, Gianfrancesco C, Heller S, James P, et al. Training in flexible, intensive insulin management to enable dietary freedom in people with type 1 diabetes: dose adjustment for normal eating (DAFNE) randomised controlled trial. BMJ 2002;325:746-9.
    1. Cryer PE. The barrier of hypoglycemia in diabetes. Diabetes 2008;57:3169-76.
    1. Bode BW, Schwartz S, Stubbs HA, Block JE. Glycemic characteristics in continuously monitored patients with type 1 and type 2 diabetes: normative values. Diabetes Care 2005;28:2361-6.
    1. Klonoff DC. Continuous glucose monitoring: roadmap for 21st century diabetes therapy. Diabetes Care 2005;28:1231-9.
    1. Tamborlane WV, Beck RW, Bode BW, Buckingham B, Chase HP, Clemons R, et al. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 2008;359:1464-76.
    1. Pickup J, Keen H. Continuous subcutaneous insulin infusion at 25 years: evidence base for the expanding use of insulin pump therapy in type 1 diabetes. Diabetes Care 2002;25:593-8.
    1. Bergenstal RM, Tamborlane WV, Ahmann A, Buse JB, Dailey G, Davis SN, et al. Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes. N Engl J Med 2010;363:311-20.
    1. The DCCT Research Group. Epidemiology of severe hypoglycemia in the Diabetes Control and Complications Trial. Am J Med 1991;90:450-9.
    1. Cheyne EH, Sherwin RS, Lunt MJ, Cavan DA, Thomas PW, Kerr D. Influence of alcohol on cognitive performance during mild hypoglycaemia; implications for type 1 diabetes. Diabet Med 2004;21:230-7.
    1. Hovorka R. Continuous glucose monitoring and closed-loop systems. Diabet Med 2006;23:1-12.
    1. Steil GM, Rebrin K. Closed-loop insulin delivery—what lies between where we are and where we are going? Expert Opin Drug Deliv 2005;2:353-62.
    1. Renard E, Costalat G, Chevassus H, Bringer J. Artificial beta-cell: clinical experience toward an implantable closed-loop insulin delivery system. Diabetes Metab 2006;32:497-502.
    1. Kowalski AJ. Can we really close the loop and how soon? Accelerating the availability of an artificial pancreas: a roadmap to better diabetes outcomes. Diabetes Technol Ther 2009;11(suppl 1):S113-9.
    1. Hovorka R, Allen JM, Elleri D, Chassin LJ, Harris J, Xing D, et al. Manual closed-loop insulin delivery in children and adolescents with type 1 diabetes: a phase 2 randomised crossover trial. Lancet 2010;375:743-51.
    1. Bequette BW. A critical assessment of algorithms and challenges in the development of a closed-loop artificial pancreas. Diabetes Technol Ther 2005;7:28-47.
    1. Hovorka R, Shojaee-Moradie F, Carroll PV, Chassin LJ, Gowrie IJ, Jackson NC, et al. Partitioning glucose distribution/transport, disposal, and endogenous production during IVGTT. Am J Physiol 2002;282:E992-1007.
    1. Mazor E, Averbuch A, Bar-Shalom Y, Dayan J. Interacting multiple model methods in target tracking: a survey. IEEE Trans Aerosp Electron Syst 1998;34:103-23.
    1. Service FJ, Molnar GD, Rosevear JW, Ackerman E, Gatewood LC, Taylor WF. Mean amplitude of glycemic excursions, a measure of diabetic instability. Diabetes 1970;19:644-55.
    1. Kovatchev BP, Cox DJ, Gonder-Frederick LA, Young-Hyman D, Schlundt D, Clarke W. Assessment of risk for severe hypoglycemia among adults with IDDM: validation of the low blood glucose index. Diabetes Care 1998;21:1870-5.
    1. Elleri D, Allen JM, Nodale M, Wilinska ME, Acerini CL, Dunger DB, et al. Suspended insulin infusion during overnight closed-loop glucose control in children and adolescents with type 1 diabetes. Diabet Med 2010;27:480-4.
    1. Kovatchev B, Cobelli C, Renard E, Anderson S, Breton M, Patek S, et al. Multinational study of subcutaneous model-predictive closed-loop control in type 1 diabetes mellitus: summary of the results. J Diabetes Sci Technol 2010;4:1374-81.
    1. Atlas E, Nimri R, Miller S, Grunberg EA, Phillip M. MD-logic artificial pancreas system: a pilot study in adults with type 1 diabetes. Diabetes Care 2010;33:1072-6.
    1. Castle JR, Engle JM, El YJ, Massoud RG, Yuen KC, Kagan R, et al. Novel use of glucagon in a closed-loop system for prevention of hypoglycemia in type 1 diabetes. Diabetes Care 2010;33:1282-7.
    1. Turner BC, Jenkins E, Kerr D, Sherwin RS, Cavan DA. The effect of evening alcohol consumption on next-morning glucose control in type 1 diabetes. Diabetes Care 2001;24:1888-93.
    1. El-Khatib FH, Russell SJ, Nathan DM, Sutherlin RG, Damiano ER. A bihormonal closed-loop artificial pancreas for type 1 diabetes. Sci Transl Med 2010;2:27ra27.
    1. Renard E, Place J, Cantwell M, Chevassus H, Palerm CC. Closed-loop insulin delivery using a subcutaneous glucose sensor and intraperitoneal insulin delivery: feasibility study testing a new model for the artificial pancreas. Diabetes Care 2010;33:121-7.

Source: PubMed