Sonoelastography to Predict Rotator Cuff Tears

1. marts 2021 opdateret af: Chang Gung Memorial Hospital

Use Sonoelastography to Predict the Reparability of Large-to-massive Rotator Cuff Tears

Large-to-massive rotator cuff tears accounts for 30% of all rotator cuff tears. These problems can be solved by surgeries, but only part of them can be completely repaired. The prognosis for partial repair is worse than complete repair, so evaluating the possibility of complete repair is so important that it will affect the decision of treatment. More and more recent researches focused on using magnetic resonance imaging (MRI) for evaluation of fatty infiltration of rotator cuff muscles to predict the reparability of large-to-massive rotator cuff tears. However, the availability of MRI is not that good as ultrasound, so some researchers are starting to use ultrasound to predict the reparability of large-to-massive rotator cuff tears. Because it is hard to observe the tissue quality through the general ultrasound, many researchers use sonoelastography to evaluate the tissue elasticity and viability. This aim of this study is to:

  1. check the reliability of sonoelastography.
  2. associate the findings of sonoelastography to the results of MRI.
  3. build a predictive model for the reparability of large-to-massive rotator cuff tears.

Studieoversigt

Status

Afsluttet

Intervention / Behandling

Undersøgelsestype

Observationel

Tilmelding (Faktiske)

50

Kontakter og lokationer

Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.

Studiesteder

      • Kaohsiung, Taiwan, 833
        • Kaohsiung Chang Gung Memorial Hospital

Deltagelseskriterier

Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.

Berettigelseskriterier

Aldre berettiget til at studere

18 år til 80 år (Voksen, Ældre voksen)

Tager imod sunde frivillige

Ingen

Køn, der er berettiget til at studere

Alle

Prøveudtagningsmetode

Ikke-sandsynlighedsprøve

Studiebefolkning

The investigator will enroll participants with diagnosis of large-to-massive rotator cuff tears who will receive surgical repairs later, and sonoelastography and MRI are necessary for preoperative evaluation.

Beskrivelse

Inclusion Criteria:

  1. Patients who are diagnosed with rotator cuff tear by an orthopedist.
  2. The large-to-massive rotator cuff tears need to be confirmed by magnetic resonance imaging or ultrasound. The definitions of this diagnosis include a tear over 3 cm or any full-thickness of tears in more than two rotator cuff muscles.
  3. Being willing to cooperated with the arranged examinations before the operation.

Exclusion Criteria:

  1. Patients who are only diagnosed with partial-thickness rotator cuff muscle tears or small- to large- sized full-thickness tears.
  2. Patients who have acromioclavicular arthritis that needs distal clavicle resection.
  3. Patients who had serious glenohumeral arthritis, pseudoparalysis, or any other shoulder trauma history.

Studieplan

Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.

Hvordan er undersøgelsen tilrettelagt?

Design detaljer

  • Observationsmodeller: Kohorte
  • Tidsperspektiver: Tværsnit

Hvad måler undersøgelsen?

Primære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Sonoelastography
Tidsramme: Baseline

This diagnostic test would be executed by a physician who has experiences of using Siemens Acuson S2000 ultrasound system for more than three year. The examination includes two of four rotator cuff muscles, listed as follows, supraspinatus muscle and infraspinatus muscle. Linear transducer (4-9 MHz) would be used for the whole test. The probe would be positioned along the longitudinal axis of the muscle belly.

Compressive sonoelastography technique is used for semiquantitative analysis. In order to maintain the quality of images, quality factor needs to be over or equal to 60. The quantitative analysis was performed by shear wave sonoelastography. While assessing, the region of interest (ROI) would be divided into four quadrants. The physicians will measure the shear wave velocity for wight times at the center point of each quadrant.

Baseline

Sekundære resultatmål

Resultatmål
Foranstaltningsbeskrivelse
Tidsramme
Passive range of motions
Tidsramme: Baseline
Passive range of motions of internal and external rotation, abduction, extension, and flexion, which are assessed by a goniometer.
Baseline
Muscle strength
Tidsramme: Baseline
Muscle strength assessment include the following motions, which are internal and external rotation, abduction, extension, and flexion. They are measured by a manual dynamometer.
Baseline
American Shoulder and Elbow Surgeons score.
Tidsramme: Baseline
Shoulder function that are scaled by using American Shoulder and Elbow Surgeons score.
Baseline
MRI
Tidsramme: Baseline
MRI was performed on a 1.5T system (Signa Horizon LX, GE Healthcare) equipped with a standard shoulder surface coil. The shoulder MRI protocol was identical for all patients and consisted of axial proton-density-weighted fast spin-echo with fat suppression sequence, the sequences was performed above the level o f the acromioclavicular joint down to below the axillary pouch with the following parameters: Coronal oblique proton-density-weighted with and without fat suppression, with axis parallel to the supraspinatus tendon, and sagittal oblique proton-density--weighted with and without fat suppression, with axis perpendicular to the coronal oblique axis, the ast spin-echo sequences were performed with the following parameters: TR msec/TE msec, 2700-4800 /25-40, echo-train length, 6; matrix, 256 × 256; field of view, 180 mm; section thickness, 2.5 mm with 2-mm gap. No IV or intraarticular gadolinium was administered.
Baseline
Surgical outcome
Tidsramme: Baseline
The investigator will discuss the surgical outcome with the orthopedist to define whether the tear side is repaired partially or completely. If the tendon can be repaired using simple side-to-side tendon repair, horizontal mattress transosseous repair via the bone trough and reduced into an anatomic position, or via the rough area of greater tuberosity (marginal converging repair), it is defined as complete repair of the tendon. If the tendon cannot be reduced to the anatomic footprint and the orthopedist only secures re-attachment of the infraspinatus to the posterior side of greater tuberosity and/or the subscapularis of lesser tuberosity, it is defined as partial repair of the tendon.
Baseline
B-mode ultrasound
Tidsramme: Baseline

The ultrasound machine will be calibrated similarly using the musculoskeletal (MSK) preset and B-mode with autogain. The dynamic range will be set at 70 dB for all participants. The examiners will apply adequate transducer pressure to obtain optimal images for comparisons of muscle and bone echo. The Heckmatt scale will be used to grade the muscle echogenicity within each ROI of supraspinatus and infraspinatus muscles. The grading rule is as follows:

  1. Grade 1: Normal
  2. Grade 2: Increased muscle echo intensity with distinct bone echo
  3. Grade 3: Marked increased muscle echo with reduced bone echo
  4. Grade 4: Very strong muscle echo and complete loss of bone echo
Baseline

Samarbejdspartnere og efterforskere

Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.

Efterforskere

  • Studiestol: Po-Cheng Chen, MD, Chang Gung Memorial Hospital

Publikationer og nyttige links

Den person, der er ansvarlig for at indtaste oplysninger om undersøgelsen, leverer frivilligt disse publikationer. Disse kan handle om alt relateret til undersøgelsen.

Datoer for undersøgelser

Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.

Studer store datoer

Studiestart (Faktiske)

22. oktober 2018

Primær færdiggørelse (Faktiske)

31. december 2020

Studieafslutning (Faktiske)

31. december 2020

Datoer for studieregistrering

Først indsendt

18. september 2018

Først indsendt, der opfyldte QC-kriterier

21. september 2018

Først opslået (Faktiske)

25. september 2018

Opdateringer af undersøgelsesjournaler

Sidste opdatering sendt (Faktiske)

3. marts 2021

Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier

1. marts 2021

Sidst verificeret

1. februar 2020

Mere information

Begreber relateret til denne undersøgelse

Yderligere relevante MeSH-vilkår

Andre undersøgelses-id-numre

  • CMRPG8H0811

Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter

Studerer et amerikansk FDA-reguleret lægemiddelprodukt

Ingen

Studerer et amerikansk FDA-reguleret enhedsprodukt

Ingen

produkt fremstillet i og eksporteret fra U.S.A.

Ingen

Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .