Sonoelastography to Predict Rotator Cuff Tears

1 maart 2021 bijgewerkt door: Chang Gung Memorial Hospital

Use Sonoelastography to Predict the Reparability of Large-to-massive Rotator Cuff Tears

Large-to-massive rotator cuff tears accounts for 30% of all rotator cuff tears. These problems can be solved by surgeries, but only part of them can be completely repaired. The prognosis for partial repair is worse than complete repair, so evaluating the possibility of complete repair is so important that it will affect the decision of treatment. More and more recent researches focused on using magnetic resonance imaging (MRI) for evaluation of fatty infiltration of rotator cuff muscles to predict the reparability of large-to-massive rotator cuff tears. However, the availability of MRI is not that good as ultrasound, so some researchers are starting to use ultrasound to predict the reparability of large-to-massive rotator cuff tears. Because it is hard to observe the tissue quality through the general ultrasound, many researchers use sonoelastography to evaluate the tissue elasticity and viability. This aim of this study is to:

  1. check the reliability of sonoelastography.
  2. associate the findings of sonoelastography to the results of MRI.
  3. build a predictive model for the reparability of large-to-massive rotator cuff tears.

Studie Overzicht

Toestand

Voltooid

Interventie / Behandeling

Studietype

Observationeel

Inschrijving (Werkelijk)

50

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studie Locaties

      • Kaohsiung, Taiwan, 833
        • Kaohsiung Chang Gung Memorial Hospital

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar tot 80 jaar (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Nee

Geslachten die in aanmerking komen voor studie

Allemaal

Bemonsteringsmethode

Niet-waarschijnlijkheidssteekproef

Studie Bevolking

The investigator will enroll participants with diagnosis of large-to-massive rotator cuff tears who will receive surgical repairs later, and sonoelastography and MRI are necessary for preoperative evaluation.

Beschrijving

Inclusion Criteria:

  1. Patients who are diagnosed with rotator cuff tear by an orthopedist.
  2. The large-to-massive rotator cuff tears need to be confirmed by magnetic resonance imaging or ultrasound. The definitions of this diagnosis include a tear over 3 cm or any full-thickness of tears in more than two rotator cuff muscles.
  3. Being willing to cooperated with the arranged examinations before the operation.

Exclusion Criteria:

  1. Patients who are only diagnosed with partial-thickness rotator cuff muscle tears or small- to large- sized full-thickness tears.
  2. Patients who have acromioclavicular arthritis that needs distal clavicle resection.
  3. Patients who had serious glenohumeral arthritis, pseudoparalysis, or any other shoulder trauma history.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Observatiemodellen: Cohort
  • Tijdsperspectieven: Dwarsdoorsnede

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Sonoelastography
Tijdsspanne: Baseline

This diagnostic test would be executed by a physician who has experiences of using Siemens Acuson S2000 ultrasound system for more than three year. The examination includes two of four rotator cuff muscles, listed as follows, supraspinatus muscle and infraspinatus muscle. Linear transducer (4-9 MHz) would be used for the whole test. The probe would be positioned along the longitudinal axis of the muscle belly.

Compressive sonoelastography technique is used for semiquantitative analysis. In order to maintain the quality of images, quality factor needs to be over or equal to 60. The quantitative analysis was performed by shear wave sonoelastography. While assessing, the region of interest (ROI) would be divided into four quadrants. The physicians will measure the shear wave velocity for wight times at the center point of each quadrant.

Baseline

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Passive range of motions
Tijdsspanne: Baseline
Passive range of motions of internal and external rotation, abduction, extension, and flexion, which are assessed by a goniometer.
Baseline
Muscle strength
Tijdsspanne: Baseline
Muscle strength assessment include the following motions, which are internal and external rotation, abduction, extension, and flexion. They are measured by a manual dynamometer.
Baseline
American Shoulder and Elbow Surgeons score.
Tijdsspanne: Baseline
Shoulder function that are scaled by using American Shoulder and Elbow Surgeons score.
Baseline
MRI
Tijdsspanne: Baseline
MRI was performed on a 1.5T system (Signa Horizon LX, GE Healthcare) equipped with a standard shoulder surface coil. The shoulder MRI protocol was identical for all patients and consisted of axial proton-density-weighted fast spin-echo with fat suppression sequence, the sequences was performed above the level o f the acromioclavicular joint down to below the axillary pouch with the following parameters: Coronal oblique proton-density-weighted with and without fat suppression, with axis parallel to the supraspinatus tendon, and sagittal oblique proton-density--weighted with and without fat suppression, with axis perpendicular to the coronal oblique axis, the ast spin-echo sequences were performed with the following parameters: TR msec/TE msec, 2700-4800 /25-40, echo-train length, 6; matrix, 256 × 256; field of view, 180 mm; section thickness, 2.5 mm with 2-mm gap. No IV or intraarticular gadolinium was administered.
Baseline
Surgical outcome
Tijdsspanne: Baseline
The investigator will discuss the surgical outcome with the orthopedist to define whether the tear side is repaired partially or completely. If the tendon can be repaired using simple side-to-side tendon repair, horizontal mattress transosseous repair via the bone trough and reduced into an anatomic position, or via the rough area of greater tuberosity (marginal converging repair), it is defined as complete repair of the tendon. If the tendon cannot be reduced to the anatomic footprint and the orthopedist only secures re-attachment of the infraspinatus to the posterior side of greater tuberosity and/or the subscapularis of lesser tuberosity, it is defined as partial repair of the tendon.
Baseline
B-mode ultrasound
Tijdsspanne: Baseline

The ultrasound machine will be calibrated similarly using the musculoskeletal (MSK) preset and B-mode with autogain. The dynamic range will be set at 70 dB for all participants. The examiners will apply adequate transducer pressure to obtain optimal images for comparisons of muscle and bone echo. The Heckmatt scale will be used to grade the muscle echogenicity within each ROI of supraspinatus and infraspinatus muscles. The grading rule is as follows:

  1. Grade 1: Normal
  2. Grade 2: Increased muscle echo intensity with distinct bone echo
  3. Grade 3: Marked increased muscle echo with reduced bone echo
  4. Grade 4: Very strong muscle echo and complete loss of bone echo
Baseline

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Onderzoekers

  • Studie stoel: Po-Cheng Chen, MD, Chang Gung Memorial Hospital

Publicaties en nuttige links

De persoon die verantwoordelijk is voor het invoeren van informatie over het onderzoek stelt deze publicaties vrijwillig ter beschikking. Dit kan gaan over alles wat met het onderzoek te maken heeft.

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Werkelijk)

22 oktober 2018

Primaire voltooiing (Werkelijk)

31 december 2020

Studie voltooiing (Werkelijk)

31 december 2020

Studieregistratiedata

Eerst ingediend

18 september 2018

Eerst ingediend dat voldeed aan de QC-criteria

21 september 2018

Eerst geplaatst (Werkelijk)

25 september 2018

Updates van studierecords

Laatste update geplaatst (Werkelijk)

3 maart 2021

Laatste update ingediend die voldeed aan QC-criteria

1 maart 2021

Laatst geverifieerd

1 februari 2020

Meer informatie

Termen gerelateerd aan deze studie

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Nee

product vervaardigd in en geëxporteerd uit de V.S.

Nee

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .