- ICH GCP
- Registro de ensayos clínicos de EE. UU.
- Ensayo clínico NCT01003665
Comparison of Continuous Non-invasive Arterial Pressure (CNAP) With Invasive Arterial Pressure Monitoring
Accuracy of the CNAP™ Monitor (Continuous Non-invasive Arterial Pressure) Versus Invasive Radial Arterial Monitoring in Surgical Patients
The purpose of this study was to compare the Continuous Non-invasive Arterial Pressure (CNAP) monitor with the gold standard of invasive arterial pressure monitoring during:
- induction and maintenance of general anaesthesia
- intensive care unit treatment of postoperative patients with an ASA 3 or 4 status
Descripción general del estudio
Estado
Condiciones
Descripción detallada
The CNAP monitor continuously measures blood pressure using a finger cuff. It showed a good agreement to invasive blood pressure measurements during anaesthesia induction and maintenance during surgery. In critical ill patients cardiac arrhythmia is probably a confounding factor affecting accuracy and interchangeability of CNAP. As systolic arterial CNAP pressure in comparison to mean pressure shows no statistical interchangeability with invasive measurements, mean pressure should be considered when making therapy decisions.
A problem is the missing standard criterion for comparison of continuously devices with invasive pressure.
Tipo de estudio
Inscripción (Actual)
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
-
Kiel, Alemania, 24106
- Klinik für Anästhesiologie und Operative Intensivmedizin, Universitätsklinik Schleswig-Holstein, Campus Kiel
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Método de muestreo
Población de estudio
Descripción
Inclusion Criteria:
- Age >18
- BMI < 35 kg.m²
- Patients capable of giving informed consent
- patients undergoing elective surgical procedures in supine position under general anesthesia
- perfusion of the examined arm evidenced by a positive Allen's test
Exclusion Criteria:
- Patients not competent or unwilling to provide informed consent
- Patients with history of neurological, neuromuscular seizure
- Patients where IBP cannula, CNAP™ finger-cuff and CNAP™ upper arm cuff cannot be placed on the same arm
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
Cohortes e Intervenciones
Grupo / Cohorte |
---|
Intensive Care treatement
ASA status 3 or 4 patients with invasive blood pressure measurement on the intensive care unit
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Periodo de tiempo |
---|---|
Bias and percentage error of CNAP compared with invasive arterial pressure measurement
Periodo de tiempo: 1.) Induction of general anaesthesia. 2.) Maintanance of general anaesthesia. 3.) Intensive Care Treatement after surgery
|
1.) Induction of general anaesthesia. 2.) Maintanance of general anaesthesia. 3.) Intensive Care Treatement after surgery
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Silla de estudio: Markus Steinfath, Professor, Klinik für Anästhesiologie und Operative Intensivmedizin, UKSH Kiel
- Director de estudio: Robert Hanss, Professor, Klinik für Anästhesiologie und Operative Intensivmedizin, UKSH Kiel
Publicaciones y enlaces útiles
Publicaciones Generales
- Ilies C, Bauer M, Berg P, Rosenberg J, Hedderich J, Bein B, Hinz J, Hanss R. Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth. 2012 Feb;108(2):202-10. doi: 10.1093/bja/aer394. Epub 2011 Dec 12.
- Ilies C, Grudev G, Hedderich J, Renner J, Steinfath M, Bein B, Haake N, Hanss R. Comparison of a continuous noninvasive arterial pressure device with invasive measurements in cardiovascular postsurgical intensive care patients: a prospective observational study. Eur J Anaesthesiol. 2015 Jan;32(1):20-8. doi: 10.1097/EJA.0000000000000136.
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Estimar)
Actualizaciones de registros de estudio
Última actualización publicada (Estimar)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Otros números de identificación del estudio
- CNAP-2009-1
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .