- ICH GCP
- Registro de ensayos clínicos de EE. UU.
- Ensayo clínico NCT01221649
Monitoring of Fluid Status in Heart Failure Patients by Intrathoracic Impedance Measurement in Japan (J-HomeCARE II)
Monitoring of Fluid Status in Heart Failure Patients by Intrathoracic Impedance Measurement in Japan
Descripción general del estudio
Estado
Condiciones
Intervención / Tratamiento
Descripción detallada
The new BIOTRONIK ICD family Lumax 540 is capable of measuring the intrathoracic impedance on a daily basis and transmission of impedance trend data via the Home Monitoring technology. In future, this will allow for automatic early detection of imminent decompensation at a presymptomatic stage and immediate initialization of clinical meaningful therapeutic interventions preventing heart failure related hospitalizations. The aim of this investigation is to record long-term impedance trends in heart failure patients with and without HF related events (hospitalizations due to acute decompensation) in order to develop algorithms for device-based early detection and warning of HF deterioration accompanied by pulmonary congestion.
Lumax is the name of a family of implantable ICDs. The primary objective of the therapy is the prevention of sudden cardiac death. The aim is to automatically detect and terminate cardiac arrest caused by ventricular tachyarrhythmia. All major therapeutical approaches from the field of cardiology and electrophysiology are contained within the Lumax family.
Furthermore, the device is capable of treating bradycardia arrhythmias and congestive heart failure. Congestive heart failure is treated with cardiac resynchronization therapy with multisite ventricular pacing known as cardiac resynchronization therapy (CRT).
The integrated Home Monitoring component can provide information about occurring rhythm disturbances and delivered therapies close to real time as well as by intracardiac electrogram (IEGM) Online HD®. Furthermore, statistical data about the patient's condition as well as information about the integrity status of the implant itself are sent.
The implantation of an ICD is a symptomatic therapy with the following objectives:
Termination of spontaneous ventricular fibrillation (VF) through shock delivery Termination of spontaneous ventricular tachycardia (VT) by means of antitachycardia pacing (ATP); in case of ineffective ATP or hemodynamically not tolerated VT with shock delivery Cardiac resynchronization by multisite ventricular pacing (triple-chamber implant) Compensation of bradycardia through ventricular (single-chamber implant) or atrioventricular sequential pacing (dual- and triple-chamber implant)
Tipo de estudio
Inscripción (Actual)
Fase
- Fase 4
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
-
Hiroshima, Japón
- Hiroshima Shimin Hospital
-
Tokyo, Japón
- Tokyo Women's Medical University
-
-
Fukuoka
-
Kitakyushu, Fukuoka, Japón, 802-8555
- Kokura Kinen Hospital
-
-
Mie
-
Ise, Mie, Japón, 516-0805
- Yamada Red Cross Hospital
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
- Patient eligible for 1-, 2- or 3-chamber ICD implantation according to current guidelines or patient already implanted with a Lumax 540 device or successor
- New York Heart Association (NYHA)-class II to IV
- Left ventricular ejection fraction (LVEF) lower or equal to 35%
- Patient accepts Home Monitoring concept and has sufficient Global System for Mobile Communications (GSM) /General Packet Radio Service (GPRS) net coverage
- Increased risk for HF-related hospitalization according to pre-defined criteria
- Patient information
- Informed consent
Exclusion Criteria:
- Age < 18 years
- Contraindication for ICD implantation
- Post heart transplant (HTX) or actively listed for HTX
- Cardiac surgery within the previous 3 months or planned at time of inclusion
- Acute coronary syndrome within the previous 3 months
- Chronic renal dialysis
- Pregnant or breast-feeding women
- Limited contractual capability
- Participation in another study
- Anticipated non-compliance with the follow-up scheme
- Life expectancy not longer than 1.5 years due to a non-cardiac disease
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Cuidados de apoyo
- Asignación: N / A
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
HF Events
Periodo de tiempo: 21 months follow-up
|
Long-term impedance trends in patients with clinically relevant HF events, to support the development of impedance based detection algorithms.
A posterior assessment of sensitivity and false alarm rate of the detection algorithms.
|
21 months follow-up
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: Shoda Morio, DR, Tokyo Women's University Hospital
Publicaciones y enlaces útiles
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio
Finalización primaria (Actual)
Finalización del estudio (Actual)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Estimar)
Actualizaciones de registros de estudio
Última actualización publicada (Estimar)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- JHC2
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .