Use of AlignRT System During Stereotactic Radiosurgery
Aperçu de l'étude
Statut
Statut
Les conditions
Les conditions
Intervention / Traitement
Intervention / Traitement
Description détaillée
This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.
Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.
Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.
A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.
Type d'étude
Type d'étude
Inscription (Réel)
Inscription
Phase
Phase
- N'est pas applicable
Contacts et emplacements
Lieux d'étude
-
-
-
Ljubljana, Slovénie, 1000
- Institute of Oncology Ljubljana
-
-
Critères de participation
Critère d'éligibilité
Critère d'éligibilité
Âges éligibles pour étudier
- Adulte
- Adulte plus âgé
Accepte les volontaires sains
La description
Inclusion Criteria:
- Adults aged 18 years or older
- Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
- Treatment performed using the TrueBeam STx linear accelerator
- Able and willing to provide written informed consent.
Exclusion Criteria:
- Patients younger than 18 years
- Patients receiving fractionated stereotactic radiotherapy
- Treatment performed using a linear accelerator other than TrueBeam STx.
Plan d'étude
Comment l'étude est-elle conçue ?
Détails de conception
- Objectif principal: Autre
- Répartition: N / A
- Modèle interventionnel: Affectation à un seul groupe
- Masquage: Aucun (étiquette ouverte)
Nombre de bras
Armes et Interventions
Groupe de participants / BrasGroupe de participants / Bras |
Intervention / TraitementIntervention / Traitement |
|---|---|
|
Expérimental: Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery.
A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
|
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
|
Que mesure l'étude ?
Principaux critères de jugement
Principaux critères de jugement
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
Délai: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
Délai: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Mesures de résultats secondaires
Mesures de résultats secondaires
Mesure des résultats |
Description de la mesure |
Délai |
|---|---|---|
|
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
Délai: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning.
Deviations were expressed in centimeters.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
Délai: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning.
Deviations were expressed in degrees.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Collaborateurs et enquêteurs
Parrainer
Parrainer
Les enquêteurs
Les enquêteurs
- Chercheur principal: Klemen Salmič, MSc, Institute of Oncology Ljubljana
Dates d'enregistrement des études
Dates principales de l'étude
Début de l'étude (Réel)
Début de l'étude
Achèvement primaire (Réel)
Achèvement primaire
Achèvement de l'étude (Réel)
Achèvement de l'étude
Dates d'inscription aux études
Première soumission
Première soumission
Première soumission répondant aux critères de contrôle qualité
Première soumission répondant aux critères de contrôle qualité
Première publication (Réel)
Première publication
Mises à jour des dossiers d'étude
Dernière mise à jour publiée (Réel)
Dernière mise à jour publiée
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière mise à jour soumise répondant aux critères de contrôle qualité
Dernière vérification
Dernière vérification
Plus d'information
Termes liés à cette étude
Termes MeSH pertinents supplémentaires
Autres numéros d'identification d'étude
Autres numéros d'identification d'étude
- ERIDNPVO-0017/2025
Plan pour les données individuelles des participants (IPD)
Prévoyez-vous de partager les données individuelles des participants (DPI) ?
Informations sur les médicaments et les dispositifs, documents d'étude
Étudie un produit pharmaceutique réglementé par la FDA américaine
Étudie un produit d'appareil réglementé par la FDA américaine
Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .