Effects of Supplemental Oxygen Delivered by a Portable Oxygen Concentrator Compared to a Liquid Oxygen Device in COPD

14 febbraio 2018 aggiornato da: Klaus Kenn, Schön Klinik Berchtesgadener Land

Effects of Supplemental Oxygen Delivered by a Portable Oxygen Concentrator Compared to a Liquid Oxygen Device in Hypoxemic COPD Patients at Rest

The aim of this study is to investigate the effects of supplemental oxygen on blood oxygenation at rest in Patients with severe to very severe COPD comparing the portable oxygen concentrator (Activox™ 4L) to a liquid continuous oxygen device (Companion®).

Panoramica dello studio

Descrizione dettagliata

It has been proven that supplemental oxygen increases oxygen saturation and reduces dyspnea in hypoxemic COPD Patients. There are different oxygen delivery systems available, some of them operating with liquid oxygen and others by concentrating the ambient air (concentrators). An advantage of the concentrator system is, that there is no refill required, but only a power plug and plug socket, enabling the patient for higher mobility and to be more autonomous. There are only few publications about whether portable oxygen concentrators have a comparable efficiency on oxygen saturation to liquid oxygen devices with continuous flow (LOD) or not. Until now there are no official recommendations about how to convert the oxygen flow rate for LOD (liter per minute) into the corresponding level of the POC.

Therefore, we investigate the comparability of the portable oxygen concentrator Activox™ 4L (POC) to a liquid oxygen device (Companion®) in 30 hypoxemic COPD patients at rest.

As a baseline assessment, the patients will receive a bodyplethysmography, blood gas analysis without using supplemental oxygen and an evaluation of the diffusion capacity of the lung for CO.

The study will be conducted on 2 consecutive days. 15 patients will be randomized into two different groups: First group will start with POC and will continue the following day with LOD; The second group will start with LOD and continue with POC (cross-over design). All patients will use the same oxygen devices during the study assessments (LOD: Companion 1000 (CE 0050), Chart Industries Inc., Garfiel Heights, OH, USA; POC: Activox™ 4L, Inovalabs Inc., Texas, USA). On both days of the study, patients will be connected to one of the two systems via nasal cannula for a total time period of 40 minutes, while the patient remains in a sitting position without talking. The oxygen flow rate starts at 1liter/min (LOD) or Level 1 (POC) and will be increased every 10 minutes to the next higher level until the maximum of 4l/min (LOD) or level 4 (POC) is reached. Blood gases were taken and breathing frequency will be recorded at the end of each Oxygen Level.

Tipo di studio

Interventistico

Iscrizione (Effettivo)

20

Fase

  • Non applicabile

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Luoghi di studio

      • Schoenau Am Koenigssee, Germania
        • Schoen Klinik Berchtesgadener Land

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

Da 40 anni a 80 anni (Adulto, Adulto più anziano)

Accetta volontari sani

No

Sessi ammissibili allo studio

Tutto

Descrizione

Inclusion Criteria:

  • COPD patients with hypoxemia at rest (paO2 < 55 or 60 mmHg, according to the recent supplemental oxygen guidelines [Hadringe, M., et al., British Thoracic Society guidelines for home oxygen use in adults. Thorax, 2015. 70 Suppl 1: p. i1-43.]
  • Participation in an inpatient pulmonary rehabilitation program (Schön Klinik BGL)
  • Written informed consent

Exclusion Criteria:

  • The Need of more than 4 Liter/min Oxygen at rest to achieve PaO2 > 55 or 60mmHg
  • Signs of acute exacerbation
  • General exclusion criteria for exercise tests, e.g. acute coronary syndrome, acute myo- or pericarditis, acute lung embolism, pulmonary infarction, acute uncontrolled heart insufficiency

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Scopo principale: Trattamento
  • Assegnazione: Randomizzato
  • Modello interventistico: Assegnazione incrociata
  • Mascheramento: Doppio

Armi e interventi

Gruppo di partecipanti / Arm
Intervento / Trattamento
Sperimentale: COPD patients with delivery order 1, 2
Patients will use two different Oxygen devices at rest (1.liquid oxygen device (Companion R), 2. portable Oxygen concentrator (Activox TM 4L))
This oxygen Supplementation is used in special order
Altri nomi:
  • Companion 1000 Chart Industries, Inc. Garfield Heights, USA
This oxygen Supplementation is used in special order
Altri nomi:
  • Activox™ 4L, Inovalabs Inc., Texas, USA
Sperimentale: COPD patients with delivery order 2,1
Patients will use two different Oxygen devices at rest (1. portable Oxygen concentrator (Activox TM 4L), 2. liquid oxygen device (Companion R))
This oxygen Supplementation is used in special order
Altri nomi:
  • Companion 1000 Chart Industries, Inc. Garfield Heights, USA
This oxygen Supplementation is used in special order
Altri nomi:
  • Activox™ 4L, Inovalabs Inc., Texas, USA

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Difference between LOD and POC in Partial pressure of oxygen levels
Lasso di tempo: after 10 minutes of each LOD step compared to 10 minutes of each POC step
PO2 values between LOD and POC at comparable Levels of oxygen supply (1liter/minute LOD will be compared to Level 1 of POC until 4liters/Minute LOD to Level 4 of POC)
after 10 minutes of each LOD step compared to 10 minutes of each POC step

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Difference between LOD and POC in Partial pressure of carbon dioxide levels
Lasso di tempo: after 10 minutes of each LOD step compared to 10 minutes of each POC step
PCO2 values between LOD and POC at comparable Levels of oxygen supply (1liter/minute LOD will be compared to Level 1 of POC until 4liters/Minute LOD to Level 4 of POC)
after 10 minutes of each LOD step compared to 10 minutes of each POC step
Change of partial pressure of Oxygen from one POC step to the next step
Lasso di tempo: after 10 minutes of one POC step compared to after 10 minutes of the next POC step
Change in PO2 values from e.g. POC step 1 to POC step 2, step 2 to 3 etc.
after 10 minutes of one POC step compared to after 10 minutes of the next POC step
Change of partial pressure of carbon dioxide from one POC step to the next step
Lasso di tempo: after 10 minutes of one POC step compared to after 10 minutes of the next POC step
Change in PCO2 values between e.g. POC step 1 to POC step 2, step 2 to 3 etc.
after 10 minutes of one POC step compared to after 10 minutes of the next POC step

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

2 maggio 2017

Completamento primario (Effettivo)

15 settembre 2017

Completamento dello studio (Effettivo)

15 settembre 2017

Date di iscrizione allo studio

Primo inviato

30 maggio 2017

Primo inviato che soddisfa i criteri di controllo qualità

30 maggio 2017

Primo Inserito (Effettivo)

2 giugno 2017

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

15 febbraio 2018

Ultimo aggiornamento inviato che soddisfa i criteri QC

14 febbraio 2018

Ultimo verificato

1 febbraio 2018

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • ActivoxTitration

Piano per i dati dei singoli partecipanti (IPD)

Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?

INDECISO

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .