Denne siden ble automatisk oversatt og nøyaktigheten av oversettelsen er ikke garantert. Vennligst referer til engelsk versjon for en kildetekst.

Effects of Supplemental Oxygen Delivered by a Portable Oxygen Concentrator Compared to a Liquid Oxygen Device in COPD

14. februar 2018 oppdatert av: Klaus Kenn, Schön Klinik Berchtesgadener Land

Effects of Supplemental Oxygen Delivered by a Portable Oxygen Concentrator Compared to a Liquid Oxygen Device in Hypoxemic COPD Patients at Rest

The aim of this study is to investigate the effects of supplemental oxygen on blood oxygenation at rest in Patients with severe to very severe COPD comparing the portable oxygen concentrator (Activox™ 4L) to a liquid continuous oxygen device (Companion®).

Studieoversikt

Detaljert beskrivelse

It has been proven that supplemental oxygen increases oxygen saturation and reduces dyspnea in hypoxemic COPD Patients. There are different oxygen delivery systems available, some of them operating with liquid oxygen and others by concentrating the ambient air (concentrators). An advantage of the concentrator system is, that there is no refill required, but only a power plug and plug socket, enabling the patient for higher mobility and to be more autonomous. There are only few publications about whether portable oxygen concentrators have a comparable efficiency on oxygen saturation to liquid oxygen devices with continuous flow (LOD) or not. Until now there are no official recommendations about how to convert the oxygen flow rate for LOD (liter per minute) into the corresponding level of the POC.

Therefore, we investigate the comparability of the portable oxygen concentrator Activox™ 4L (POC) to a liquid oxygen device (Companion®) in 30 hypoxemic COPD patients at rest.

As a baseline assessment, the patients will receive a bodyplethysmography, blood gas analysis without using supplemental oxygen and an evaluation of the diffusion capacity of the lung for CO.

The study will be conducted on 2 consecutive days. 15 patients will be randomized into two different groups: First group will start with POC and will continue the following day with LOD; The second group will start with LOD and continue with POC (cross-over design). All patients will use the same oxygen devices during the study assessments (LOD: Companion 1000 (CE 0050), Chart Industries Inc., Garfiel Heights, OH, USA; POC: Activox™ 4L, Inovalabs Inc., Texas, USA). On both days of the study, patients will be connected to one of the two systems via nasal cannula for a total time period of 40 minutes, while the patient remains in a sitting position without talking. The oxygen flow rate starts at 1liter/min (LOD) or Level 1 (POC) and will be increased every 10 minutes to the next higher level until the maximum of 4l/min (LOD) or level 4 (POC) is reached. Blood gases were taken and breathing frequency will be recorded at the end of each Oxygen Level.

Studietype

Intervensjonell

Registrering (Faktiske)

20

Fase

  • Ikke aktuelt

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiesteder

      • Schoenau Am Koenigssee, Tyskland
        • Schoen Klinik Berchtesgadener Land

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

40 år til 80 år (Voksen, Eldre voksen)

Tar imot friske frivillige

Nei

Kjønn som er kvalifisert for studier

Alle

Beskrivelse

Inclusion Criteria:

  • COPD patients with hypoxemia at rest (paO2 < 55 or 60 mmHg, according to the recent supplemental oxygen guidelines [Hadringe, M., et al., British Thoracic Society guidelines for home oxygen use in adults. Thorax, 2015. 70 Suppl 1: p. i1-43.]
  • Participation in an inpatient pulmonary rehabilitation program (Schön Klinik BGL)
  • Written informed consent

Exclusion Criteria:

  • The Need of more than 4 Liter/min Oxygen at rest to achieve PaO2 > 55 or 60mmHg
  • Signs of acute exacerbation
  • General exclusion criteria for exercise tests, e.g. acute coronary syndrome, acute myo- or pericarditis, acute lung embolism, pulmonary infarction, acute uncontrolled heart insufficiency

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

  • Primært formål: Behandling
  • Tildeling: Randomisert
  • Intervensjonsmodell: Crossover-oppdrag
  • Masking: Dobbelt

Våpen og intervensjoner

Deltakergruppe / Arm
Intervensjon / Behandling
Eksperimentell: COPD patients with delivery order 1, 2
Patients will use two different Oxygen devices at rest (1.liquid oxygen device (Companion R), 2. portable Oxygen concentrator (Activox TM 4L))
This oxygen Supplementation is used in special order
Andre navn:
  • Companion 1000 Chart Industries, Inc. Garfield Heights, USA
This oxygen Supplementation is used in special order
Andre navn:
  • Activox™ 4L, Inovalabs Inc., Texas, USA
Eksperimentell: COPD patients with delivery order 2,1
Patients will use two different Oxygen devices at rest (1. portable Oxygen concentrator (Activox TM 4L), 2. liquid oxygen device (Companion R))
This oxygen Supplementation is used in special order
Andre navn:
  • Companion 1000 Chart Industries, Inc. Garfield Heights, USA
This oxygen Supplementation is used in special order
Andre navn:
  • Activox™ 4L, Inovalabs Inc., Texas, USA

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Difference between LOD and POC in Partial pressure of oxygen levels
Tidsramme: after 10 minutes of each LOD step compared to 10 minutes of each POC step
PO2 values between LOD and POC at comparable Levels of oxygen supply (1liter/minute LOD will be compared to Level 1 of POC until 4liters/Minute LOD to Level 4 of POC)
after 10 minutes of each LOD step compared to 10 minutes of each POC step

Sekundære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Difference between LOD and POC in Partial pressure of carbon dioxide levels
Tidsramme: after 10 minutes of each LOD step compared to 10 minutes of each POC step
PCO2 values between LOD and POC at comparable Levels of oxygen supply (1liter/minute LOD will be compared to Level 1 of POC until 4liters/Minute LOD to Level 4 of POC)
after 10 minutes of each LOD step compared to 10 minutes of each POC step
Change of partial pressure of Oxygen from one POC step to the next step
Tidsramme: after 10 minutes of one POC step compared to after 10 minutes of the next POC step
Change in PO2 values from e.g. POC step 1 to POC step 2, step 2 to 3 etc.
after 10 minutes of one POC step compared to after 10 minutes of the next POC step
Change of partial pressure of carbon dioxide from one POC step to the next step
Tidsramme: after 10 minutes of one POC step compared to after 10 minutes of the next POC step
Change in PCO2 values between e.g. POC step 1 to POC step 2, step 2 to 3 etc.
after 10 minutes of one POC step compared to after 10 minutes of the next POC step

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart (Faktiske)

2. mai 2017

Primær fullføring (Faktiske)

15. september 2017

Studiet fullført (Faktiske)

15. september 2017

Datoer for studieregistrering

Først innsendt

30. mai 2017

Først innsendt som oppfylte QC-kriteriene

30. mai 2017

Først lagt ut (Faktiske)

2. juni 2017

Oppdateringer av studieposter

Sist oppdatering lagt ut (Faktiske)

15. februar 2018

Siste oppdatering sendt inn som oppfylte QC-kriteriene

14. februar 2018

Sist bekreftet

1. februar 2018

Mer informasjon

Begreper knyttet til denne studien

Andre studie-ID-numre

  • ActivoxTitration

Plan for individuelle deltakerdata (IPD)

Planlegger du å dele individuelle deltakerdata (IPD)?

UBESLUTTE

Legemiddel- og utstyrsinformasjon, studiedokumenter

Studerer et amerikansk FDA-regulert medikamentprodukt

Nei

Studerer et amerikansk FDA-regulert enhetsprodukt

Nei

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .

3
Abonnere