EMG-Based Hand-Wrist Control: Study B Mirrored (PSICON-B)
EMG-Based Hand-Wrist Control: Study B: Comparison of Contralateral (Mirrored) EMG-
調査の概要
状態
状態
条件
条件
介入・治療
介入・治療
詳細な説明
研究の種類
研究の種類
入学 (実際)
入学
段階
段階
- 初期フェーズ 1
連絡先と場所
研究場所
-
-
Massachusetts
-
Holliston、Massachusetts、アメリカ、01746
- Liberating Technologies, Inc
-
Worcester、Massachusetts、アメリカ、01609
- Worcester Polytechnic Institute
-
-
参加基準
適格基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Limb-absent subjects: unilateral trans-radial limb absence or amputation
- Be capable of completing the requested contractions on the affected dominant side
Exclusion Criteria:
- Past injuries to the upper limbs that would limit their ability to complete the requested contractions
- Scars that would impede the use of surface electrodes
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:基礎科学
- 割り当て:非ランダム化
- 介入モデル:順次割り当て
- マスキング:なし(オープンラベル)
アーム数
武器と介入
参加者グループ / アーム参加者グループ / アーム |
介入・治療介入・治療 |
|---|---|
|
実験的:Study B: Limb-Absent Subject
Ten unilateral transradial limb-absent subjects will each participate in one, half-day experimental session.
Subjects will be seated and prepared in test apparatus seat (16 electrodes on the affected side, hand-wrist on the able side secured to load cells).
Subjects will complete the 1-DoF and 2-DoF "dynamic" (force-varying) contractions (40-s duration, 0.75 Hz bandlimited, uniform random target).
With 2 DoF contraction trials, hand Opn-Cls will always be one of the dimensions.
The subject will then be released from the cuff and their able side not further involved in the experiment.
The force feedback triangle cursor on the computer screen will be deleted such that only the target remains.
Subjects will then repeat 1-DoF and 2-DoF trials in which the affected side attempts to produce hand-wrist effort that mimic movement of the target (with no feedback provided).
|
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist [hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)].
sEMG system used 16 custom encased bipolar electrodes.
A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC.
A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement.
A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces.
The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target.
The subject controlled a second triangular cursor.
Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
他の名前:
|
|
実験的:Study B: Able-bodied Subjects
Ten able-bodied subjects, the electrodes will be mounted on the dominant arm for the EMG-force and EMG-target tracking trials.
In addition, the non-dominant arm will also be constrained and measured in a second load cell.
This load cell will not be used for feedback during the experiment, but will compare (RMS error, off-line) the dominant vs. non-dominant forces.
For EMG-target tracking, the dominant hand will remain in the wrist cuff (to prevent flailing during contractions), with the screen feedback disabled.
These subjects will repeat the trials with the electrodes moved to the non-dominant side.
EMG-force tracking will be repeated using mirrored contractions.
The three training methods (EMG-force ipsilateral, EMG-force contralateral mirrored, EMG-target on the dominant side) will be contrasted to help understand the source of errors when training with limb-absent subjects.
|
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist [hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)].
sEMG system used 16 custom encased bipolar electrodes.
A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC.
A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement.
A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces.
The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target.
The subject controlled a second triangular cursor.
Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
他の名前:
|
この研究は何を測定していますか?
主要な結果の測定
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
EMG-Force Models
時間枠:1 day visit
|
Electromyogram waveform measurements (signal amplitude, frequency) will be related to force measurements (magnitude and moment) as related to the target tracking tasks. These signal-force data will be use to generate the models. The specific outcome for each subject will be the RMS error measured in the tracking tasks as the difference between the generated control signal from the user versus the prompted target signal, as measured by an EMG array. |
1 day visit
|
協力者と研究者
協力者
協力者
研究記録日
主要日程の研究
研究開始 (実際)
研究開始
一次修了 (実際)
一次修了
研究の完了 (実際)
研究の完了
試験登録日
最初に提出
最初に提出
QC基準を満たした最初の提出物
QC基準を満たした最初の提出物
最初の投稿 (実際)
最初の投稿
学習記録の更新
投稿された最後の更新 (実際)
投稿された最後の更新
QC基準を満たした最後の更新が送信されました
QC基準を満たした最後の更新が送信されました
最終確認日
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
その他の研究ID番号
- 124039
- IRB-17-155 (その他の識別子:Worcester Polytechnic Institute)
- 5R42HD076519-03 (米国 NIH グラント/契約)
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
米国で製造され、米国から輸出された製品。
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。