Dozee Validation for Contactless Heart Rate and Respiration Monitoring in Clinical Settings

2026年7月29日 更新者:Turtle Shell Technologies Pvt. Ltd.

Dozee System Validation for Contactless Heart Rate and Respiration Monitoring in Clinical Settings

The Dozee VS system (K221555) is a contactless device that tracks heart rate and breathing rate using a sensor sheet placed under the mattress. It captures body vibrations and converts them into vital signs, which can be monitored remotely in real time. The system also supports add-ons like a blood pressure cuff, pulse oximeter, and thermometer, offering a complete monitoring solution. This study aims to validate the updated AI-powered heart rate and improved breathing rate algorithms across different sensor sheet types to confirm clinical accuracy

調査の概要

状態

募集

条件

介入・治療

詳細な説明

The Dozee VS system (K221555), developed by Turtle Shell Technologies, is a non-invasive, contactless device that continuously monitors heart rate (HR) and breathing rate (RR). A thin sensor sheet placed under the mattress detects tiny body vibrations from the heart and lungs, which are processed by smart algorithms and displayed on a secure online dashboard for real-time monitoring.Unlike traditional methods such as ECGs, nasal tubes, or thermistors-which are invasive, require skilled placement, and are uncomfortable for long-term use-Dozee is simple, comfortable, and doesn't rely on patient compliance like wearables do. The system can also connect to add-on devices, including a blood pressure cuff, pulse oximeter, and thermometer, giving a full picture of a patient's vital signs and helping spot early signs of deterioration. Hardware upgrades now include dual-sensor sheets that improve accuracy across different mattresses and patient types. The new HR algorithm uses AI/ML for better accuracy, while the improved RR algorithm performs more reliably even with movement or weak signals.Study Goal: This study aims to validate the updated algorithms (Gen 2) for HR and RR across the three sensor sheet types (DS, S1, and S1-I) to ensure they meet clinical accuracy standards

研究の種類

観察的

入学 (推定)

20

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

研究場所

    • Texas
      • Lake Jackson、Texas、アメリカ、77566
        • 募集
        • Oak Village Healthcare
        • コンタクト:

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

  • 大人
  • 高齢者

健康ボランティアの受け入れ

はい

サンプリング方法

確率サンプル

調査対象母集団

The study will be conducted on atleast 20 subjects in the age group of 22 years and above, regardless of gender. The patients/volunteers be informed about the study and a written informed consent form will be obtained from them. Dozee will be placed under the mattress of the patient's bed and the data will be collected through the device whenever a patient is on the bed

説明

Inclusion Criteria:

  • Adults Male or female. (≥ 22 years of age)
  • Provide written informed consent.
  • Weight over 40kg/88.18lbs

Exclusion Criteria:

  • Are connected to a device that may interfere with the device monitoring in this study.
  • Are receiving any bedside care which may be incompatible with the study procedures.
  • A likely need to receive or undergo a procedure during the testing period.
  • Cannot accept a nasal cannula, or have an ECG leads placed on the chest.
  • Have a significant medical condition in the judgement of the investigator, which may compromise the study testing procedures.
  • Prisoners

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

コホートと介入

グループ/コホート
介入・治療
Single Arm Study (Cohort)
All participants will be on standard telemetry monitoring for Heart Rate and Respiration Rate using the gold standard (Electrocardiography and Capnography respectively). In addition, a contactless monitoring devices will be placed on the patient bed to measure data. This device is Dozee VS (Investigational Device). Data from all devices will be measured simultaneously.
All participants will be on standard telemetry monitoring for Heart Rate, Respiration Rate and movements using the gold standard (Electrocardiography and Capnography respectively)

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Mean Absolute Error (MAE) of Dozee's Heart Rate a from the Gold Standard ECG
時間枠:6 months
Heart Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Heart Rate will be calculated
6 months
Mean Absolute Error (MAE) of Dozee's Respiratory Rate from the Gold Standard Capnography
時間枠:6 months
Respiratory Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Respiratory Rate will be calculated.
6 months

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

スポンサー

捜査官

  • 主任研究者:Lindsay Darthard, LVN LNFA、Oak Village Healthcare

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (実際)

2026年7月15日

一次修了 (推定)

2026年12月15日

研究の完了 (推定)

2026年12月31日

試験登録日

最初に提出

2026年7月24日

QC基準を満たした最初の提出物

2026年7月29日

最初の投稿 (実際)

2026年7月30日

学習記録の更新

投稿された最後の更新 (実際)

2026年7月30日

QC基準を満たした最後の更新が送信されました

2026年7月29日

最終確認日

2026年7月1日

詳しくは

本研究に関する用語

キーワード

その他の研究ID番号

  • DZ_VS_VAL_ON

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

IPD プランの説明

Individual Participant Data (IPD) are protected under HIPPA Act of 1996

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

はい

米国で製造され、米国から輸出された製品。