Cette page a été traduite automatiquement et l'exactitude de la traduction n'est pas garantie. Veuillez vous référer au version anglaise pour un texte source.

Dozee Validation for Contactless Heart Rate and Respiration Monitoring in Clinical Settings

29 juillet 2026 mis à jour par: Turtle Shell Technologies Pvt. Ltd.

Dozee System Validation for Contactless Heart Rate and Respiration Monitoring in Clinical Settings

The Dozee VS system (K221555) is a contactless device that tracks heart rate and breathing rate using a sensor sheet placed under the mattress. It captures body vibrations and converts them into vital signs, which can be monitored remotely in real time. The system also supports add-ons like a blood pressure cuff, pulse oximeter, and thermometer, offering a complete monitoring solution. This study aims to validate the updated AI-powered heart rate and improved breathing rate algorithms across different sensor sheet types to confirm clinical accuracy

Aperçu de l'étude

Statut

Recrutement

Les conditions

Intervention / Traitement

Description détaillée

The Dozee VS system (K221555), developed by Turtle Shell Technologies, is a non-invasive, contactless device that continuously monitors heart rate (HR) and breathing rate (RR). A thin sensor sheet placed under the mattress detects tiny body vibrations from the heart and lungs, which are processed by smart algorithms and displayed on a secure online dashboard for real-time monitoring.Unlike traditional methods such as ECGs, nasal tubes, or thermistors-which are invasive, require skilled placement, and are uncomfortable for long-term use-Dozee is simple, comfortable, and doesn't rely on patient compliance like wearables do. The system can also connect to add-on devices, including a blood pressure cuff, pulse oximeter, and thermometer, giving a full picture of a patient's vital signs and helping spot early signs of deterioration. Hardware upgrades now include dual-sensor sheets that improve accuracy across different mattresses and patient types. The new HR algorithm uses AI/ML for better accuracy, while the improved RR algorithm performs more reliably even with movement or weak signals.Study Goal: This study aims to validate the updated algorithms (Gen 2) for HR and RR across the three sensor sheet types (DS, S1, and S1-I) to ensure they meet clinical accuracy standards

Type d'étude

Observationnel

Inscription (Estimé)

20

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Coordonnées de l'étude

Lieux d'étude

    • Texas
      • Lake Jackson, Texas, États-Unis, 77566
        • Recrutement
        • Oak Village Healthcare
        • Contact:

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

  • Adulte
  • Adulte plus âgé

Accepte les volontaires sains

Oui

Méthode d'échantillonnage

Échantillon de probabilité

Population étudiée

The study will be conducted on atleast 20 subjects in the age group of 22 years and above, regardless of gender. The patients/volunteers be informed about the study and a written informed consent form will be obtained from them. Dozee will be placed under the mattress of the patient's bed and the data will be collected through the device whenever a patient is on the bed

La description

Inclusion Criteria:

  • Adults Male or female. (≥ 22 years of age)
  • Provide written informed consent.
  • Weight over 40kg/88.18lbs

Exclusion Criteria:

  • Are connected to a device that may interfere with the device monitoring in this study.
  • Are receiving any bedside care which may be incompatible with the study procedures.
  • A likely need to receive or undergo a procedure during the testing period.
  • Cannot accept a nasal cannula, or have an ECG leads placed on the chest.
  • Have a significant medical condition in the judgement of the investigator, which may compromise the study testing procedures.
  • Prisoners

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

Cohortes et interventions

Groupe / Cohorte
Intervention / Traitement
Single Arm Study (Cohort)
All participants will be on standard telemetry monitoring for Heart Rate and Respiration Rate using the gold standard (Electrocardiography and Capnography respectively). In addition, a contactless monitoring devices will be placed on the patient bed to measure data. This device is Dozee VS (Investigational Device). Data from all devices will be measured simultaneously.
All participants will be on standard telemetry monitoring for Heart Rate, Respiration Rate and movements using the gold standard (Electrocardiography and Capnography respectively)

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Mean Absolute Error (MAE) of Dozee's Heart Rate a from the Gold Standard ECG
Délai: 6 months
Heart Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Heart Rate will be calculated
6 months
Mean Absolute Error (MAE) of Dozee's Respiratory Rate from the Gold Standard Capnography
Délai: 6 months
Respiratory Rate collected at 1 Min intervals from gold standard and Dozee. Mean Absolute Error (MAE) of the Dozee Respiratory Rate will be calculated.
6 months

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Parrainer

Les enquêteurs

  • Chercheur principal: Lindsay Darthard, LVN LNFA, Oak Village Healthcare

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude (Réel)

15 juillet 2026

Achèvement primaire (Estimé)

15 décembre 2026

Achèvement de l'étude (Estimé)

31 décembre 2026

Dates d'inscription aux études

Première soumission

24 juillet 2026

Première soumission répondant aux critères de contrôle qualité

29 juillet 2026

Première publication (Réel)

30 juillet 2026

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Réel)

30 juillet 2026

Dernière mise à jour soumise répondant aux critères de contrôle qualité

29 juillet 2026

Dernière vérification

1 juillet 2026

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • DZ_VS_VAL_ON

Plan pour les données individuelles des participants (IPD)

Prévoyez-vous de partager les données individuelles des participants (DPI) ?

NON

Description du régime IPD

Individual Participant Data (IPD) are protected under HIPPA Act of 1996

Informations sur les médicaments et les dispositifs, documents d'étude

Étudie un produit pharmaceutique réglementé par la FDA américaine

Non

Étudie un produit d'appareil réglementé par la FDA américaine

Oui

produit fabriqué et exporté des États-Unis.

Non

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .